Teledyne API T802 Paramagnetic O2 Analyzer Operation Manual

59
3.4.3.3. SET O2 SPAN GAS CONCENTRATION

Set the expected O2 span gas concentration. This should be 80% of concentration range for which the

analyzer’s analog output range is set.

O2 M-P CAL O2 SPAN CONC:20.95 %
0 2 0 .9 5 ENTR EXIT
SAMPLE GAS TO CAL:O2
O2 CO2 ENTREXIT
SAMPLE RANGE=100.00 % O2=XXX.XX
<TST TST> CAL CALZ CALS SETUP
O2 M-P CAL RANGE=100.00 % O2=XXX.XX
<TST TST> ZERO SPANCONC EXIT
SAMPLE RANGE TO CAL:RNG1
RNG1 RNG2 ENTREXIT
The O2span concentration value is
automatically default to
20.95 %.
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the O2calibration
gas.
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the
CONCENTRATION
MENU.
Only appears if either
the optional CO2sensor
is installed.
Only appears if the
analyzer is set for
DUAL range mode

07275B DCN6418