Installation and Configuration

Remote/Local Operation

the power supply into local mode, or use the REM OFF command to turn off LLO and return the power supply to local mode. When you toggle to remote mode from local mode, the power supply has the same conditions are were last set.

Remote You can use the Remote Enable command to toggle between local mode and Enable (REM) remote mode. The default setting at startup is REM ON. With the remote enable

Command command set at REM ON, you can return the power supply to remote mode whenever you send any command from the computer. Any time that you change from local to remote mode during a session, the output is programmed to turn off, since the remote mode settings may be different from the local mode settings. Send the OUT ON command to restore the output.

You can turn off Remote Enable and move power supply control to the front panel by sending the REM OFF command.To check whether the power supply is remote enabled, use the REM? query command.

Example:

REM ON

Enable the power supply to be sent into

 

remote mode

VSET 10;ISET 2

Put the unit into remote operation mode

 

(Use any command to do this.)

Release 1.2

25

Page 27
Image 27
Xantrex Technology RS232-XPD manual Example