Installation and Configuration
User Options and Settings
User Options and Settings
CAUTION
If you remove the unit's cover, use proper static control techniques to avoid damage to
You can customize remote operation settings for OVP (over voltage protection) control and TTL shutdown by changing jumper positions on the
OVP Over voltage protection (OVP) on the
control the OVP trip level using the OVSET software command. If you return the power supply to local operation by using the REM software command, control of the OVP trip level changes from software control to the front panel OVP potentiometer. The default OVP trip level is set as 110% of the power supply's rated output voltage. See Table 2.10, on page 24 for a complete list of remote power ON default settings.
You can isolate the location of OVP control by changing the positions of the Local OVP Control jumper J2 and the Remote OVP Control jumper J20, both on the
Table 2.12 OVP Control Mode Selection
PCB Jumper J2 | PCB Jumper | OVP Programming Selection |
Position | J20 Position |
|
Closed (default) | Closed (default) | Software or Front Panel OVP control |
|
| (dependent on the power supply operating state) |
Closed | Open | Software OVP control only |
Open | Closed | Front Panel OVP control only |
Open | Open | Front Panel OVP control only |
28 | Operating Manual for |