Section 4. Calibration

Introduction

WARNING

Exercise caution when using and servicing power supplies. High energy levels can be stored at the output voltage terminals on all power supplies in normal operation. In addition, potentially lethal voltages exist in the power circuit and the output connector of power supplies which are rated at 40V and over. Filter capacitors store potentially dangerous energy for some time after power is removed.

You can calibrate the RS-232 interface by adjusting the signal levels on the interface card so that they correspond to the expected signal levels on the power supply's main assembly. You may need to recalibrate the interface if you replace parts either on the interface board or on the main power supply board, or if the unit falls out of specification due to component aging drifts.

You can calibrate the RS-232 Interface for:

Voltage program

Voltage readback

Current program

Current readback

Overvoltage protection

The following equipment will be required to accurately calibrate your unit:

Digital Voltmeter, 5 1/2 digit, 0.1% accuracy or better, with test leads

Current sensing shunt resistor, rated for 150% of maximum output current, 0.25% accuracy or better

Connection wires rated for the unit's maximum output current and voltage

An RS-232 equipped computer to send the calibration commands to the unit

Calibrate the unit according to the following procedures, referring to “Command Reference” for more information about the calibration commands used.

The calibration procedures in this section are designed to be performed at an ambient temperature of 25°C ± 5°C.