Xantrex Technology RS232-XPD manual device-dependent commands available with this interface

Models: RS232-XPD

1 60
Download 60 pages 29.02 Kb
Page 26
Image 26
device-dependent commands available with this interface.

Installation and Configuration

Remote/Local Operation

Table 2.10 Remote Mode Power On Conditions

Condition

Default Settings

7.5-140 Model Example

 

 

 

Voltage

0 V

VSET 0

 

 

 

Current

0 A

ISET 0

 

 

 

Soft Voltage Limit

VMAX (see models)

VMAX 7.5

 

 

 

Soft Current Limit

IMAX (see models)

IMAX 140

 

 

 

OVP Trip Voltage

Model VMAX + 10%

OVSET 8.25

 

 

 

Delay

0.5 s

DLY 0.5S

 

 

 

Remote Enable

ON

REM ON

 

 

 

Foldback Protection

OFF

FOLD OFF

 

 

 

Output

ON

OUT ON

 

 

 

Hold

OFF

HOLD OFF

 

 

 

Unmask

NONE

UNMASK NONE

 

 

 

AUXA

OFF

AUXA OFF

 

 

 

AUXB

OFF

AUXB OFF

 

 

 

Remote Mode While in remote mode, use the interface commands to control the output of the Operation power supply from a computer. See Section 3 for a complete list of

device-dependent commands available with this interface.

From remote mode, change to local mode operation by sending either the GTL (Go to Local) command or the REM (Remote Enable) OFF command.

To return to remote mode, ensure that the REM command is set to ON and then send any valid device-dependent command.

Local Mode In local mode operation, you set the voltage and current output levels and the OVP Operation trip level with controls located on the front panel. Refer to the operating manual for

a description of the functions available at the front panel.

While in local mode operation, you can change power supply control to remote mode by ensuring that REM is ON, then sending any software command and OUT ON to restore the output. Return to local mode by using the GTL command to send

24

Operating Manual for RS-232 for XPD Series Power Supply

Page 26
Image 26
Xantrex Technology RS232-XPD manual device-dependent commands available with this interface