Installation and Configuration
Remote/Local Operation
Table 2.10 Remote Mode Power On Conditions
Condition | Default Settings |
|
|
|
|
Voltage | 0 V | VSET 0 |
|
|
|
Current | 0 A | ISET 0 |
|
|
|
Soft Voltage Limit | VMAX (see models) | VMAX 7.5 |
|
|
|
Soft Current Limit | IMAX (see models) | IMAX 140 |
|
|
|
OVP Trip Voltage | Model VMAX + 10% | OVSET 8.25 |
|
|
|
Delay | 0.5 s | DLY 0.5S |
|
|
|
Remote Enable | ON | REM ON |
|
|
|
Foldback Protection | OFF | FOLD OFF |
|
|
|
Output | ON | OUT ON |
|
|
|
Hold | OFF | HOLD OFF |
|
|
|
Unmask | NONE | UNMASK NONE |
|
|
|
AUXA | OFF | AUXA OFF |
|
|
|
AUXB | OFF | AUXB OFF |
|
|
|
Remote Mode While in remote mode, use the interface commands to control the output of the Operation power supply from a computer. See Section 3 for a complete list of
From remote mode, change to local mode operation by sending either the GTL (Go to Local) command or the REM (Remote Enable) OFF command.
To return to remote mode, ensure that the REM command is set to ON and then send any valid
Local Mode In local mode operation, you set the voltage and current output levels and the OVP Operation trip level with controls located on the front panel. Refer to the operating manual for
a description of the functions available at the front panel.
While in local mode operation, you can change power supply control to remote mode by ensuring that REM is ON, then sending any software command and OUT ON to restore the output. Return to local mode by using the GTL command to send
24 | Operating Manual for |