Xantrex Technology RS232-XPD manual Voltage Readback Calibration Procedure

Models: RS232-XPD

1 60
Download 60 pages 29.02 Kb
Page 55
Image 55

Voltage

Readback

Calibration

Procedure

Calibration

Voltage Mode Calibration

1.Set the power supply for calibration as in Figure 4.1.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command VRLO; IRLO to the power supply. Wait for the supply to settle. Measure and record the output shown on the external voltmeter. Send VRLO again.

4.Send VRHI; IRHI to the supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send VRHI again.

5.Send the command VRDAT <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter after the VRLO and VRHI commands were sent. The processor calculates the offset value required to calibrate the power supply.

When the power supply is calibrated, the low to high voltage readback calibration values (offsets) are stored as constants.

6.Use commands VSET and VOUT? commands to confirm that the calibration was successful and that linearity is observed. Refer to the voltage readback accuracy specification in Section 1.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

Release 1.2

53

Page 55
Image 55
Xantrex Technology RS232-XPD manual Voltage Readback Calibration Procedure