Local Operation
3-14 975-0200-01-01

Shipped Configuration (Local Operation)

The power supply is configured for local operation at the factory. See
Table 3-3 for a summary of this configuration. For more information on
default settings, see Table 3-9 on page 3–35.
Turning the Output On/Off
Turning the output on will also turn off the ALARM LED if the ALARM
LED has been latched on due to an alarm triggering and automatically
clearing. See “Alarms and Errors” on page 3–17.
To turn on the output power:
uPress the OUTPUT ON/OFF button on the front panel or enter the
SCPI command.
The OUTPUT ON/OFF button will illuminate.
To turn off the output power:
uPress the OUTPUT ON/OFF button when the OUTPUT ON/OFF
button is illuminated.
The output will turn off and the LEDs will not be illuminated.
Quick Tip Remote
Operation uses
these commands.
See page 5–1.
The SCPI command (s) for these instructions are:
[:]OUTPut[<channel>][:POWer][:STATe]
Table 3-3
Shipped Configuration
Local Control Configuration Additional References
Use the front panel controls to
adjust the output voltage and
current set point settings.
See Chapter 3, “Local Operation” for front
panel operation. See Chapter 4, “Analog
Programming (APG) and Isolated Analog
Programming (ISOL)” for analog
programming procedures.
The OVP set point is adjusted at
the front panel to 105% above the
maximum rated output voltage.
See “Using Over Voltage Protection
(OVP)” on page 3–24 for the adjustment
procedure.
Important:
The main output will not turn on if the shutdown function is
activated. See “Using the External Shutdown Function” on page 3–31.