Table 7 Troubleshooting Burner Issues
Problem | Possible Cause | Diagnosis | Corrective Action |
Fuses Blow on Startup | Shorted heater element. | Inspect for exposed wires or |
|
| broken insulation around |
|
| wire leads. Measure |
|
| resistance leads, should not |
|
| be 0. |
Take Burner top off and reseat insulation around heater leads.
Low Sensitivity | Leaks at ferrules or fittings. | For SCD, use a microliter | Inspect fittings for scoring |
|
| syringe containing a small | and replace if necessary. |
|
| amount of sulfur compound | Replace ferrules if worn. |
|
| e.g., CS2 to “snoop” for leaks | Hand tighten all fittings |
|
| while watching mV output. | mounted onto vespel |
|
| Leaks are very evident by a | ferrules, and wrench tighten |
|
| large increase in signal | all connections with Valco |
|
| displayed on the LED. | ferrules. |
|
| For NCD, a high background |
|
|
| could be observed from |
|
|
| atmospheric nitrogen. |
|
|
|
|
|
Low Sensitivity | Improper gas flows. | Measure gas flows. | Reset gas flows. |
|
|
|
|
Low Sensitivity | Broken ceramic tubes or | Remove and inspect tubes | Replace tubes or ferrules. |
| ferrules | and ferrules for chips and/or |
|
High Pressure |
| breaks. |
|
|
|
|
|
Low Sensitivity | Improper column | Inspect for charred Detector | After cutting off charred end |
| positioning. | end of column. | of column, measure column |
High Background |
|
| length |
|
|
| ferrule (or |
|
|
| bottom of nut) and |
|
|
| column. |
|
|
|
|
116 | Operation and Maintenance Manual |