Agilent Technologies G6600-90006 manual Two-stage pressure regulators

Page 43

with air as the oxidant in the SCD and oxygen as the oxidant in the NCD.

For the SCD, detector gases must be sulfur free (<1ppb) for proper Detector operation. In general, bottled air is preferred to “house” air from a compressor because compressors tend to generate lower quality air. In addition, pressure fluctuations induced by the air generator may be detrimental to Detector performance.

For the NCD and GC, use gases that are low in nitrogen and total hydrocarbons. Use gases with an "instrument" or "chromatographic" purity rating. Agilent recommends gases with a purity range of 99.995% to 99.9995%. Do not use oil-pumped supplied air since it contains large concentrations of hydrocarbons.

Use of appropriate traps on carrier, hydrogen and oxidant sources to improve support gas quality, such as a sulfur trap gas purifier, is recommended. The oxidant used to supply the ozone generator should be dry to prevent internal corrosion. Use of a moisture trap or dryer is recommended.

Two-stage pressure regulators

The combustion gases must be supplied to the Controller at a pressure of

25 psig or less. Use a two-stage pressure regulator rather than a single-stage regulator to eliminate pressure surges. High-quality, stainless-steel diaphragm-type regulators are recommended. On/Off valves at the regulators are useful but are not essential. Mount the valves at the outlet fitting of the pressure regulators.

If you use gases with a two-stage regulator up-stream, it will be sufficient to use single-stage regulators at the Detector.

Particle Filtration

Up-stream, inline particle filtration of 10 micron or better is required to prevent damage to Controller and Burner components.

WARNING

Ozone is a hazardous gas and a strong oxidant. Minimize exposure to ozone by using the

instrument in a well-ventilated area and venting the exhaust of the vacuum pump to a

 

 

fume hood. Turn off the ozone generator when the instrument is not in use.

 

 

 

Supply Tubing for Combustion Gases

 

Use preconditioned and cleaned copper or stainless steel tubing to supply

 

gases to the 355 SCD and 255 NCD. Do not use ordinary copper tubing since it

 

contains oils and other contaminants. Plastic tubing is not recommended since

 

it is permeable to oxygen and other contaminants that can damage columns

 

and detectors or cause elevated background. Teflon tubing may be acceptable

 

in some clean environments.

Operation and Maintenance Manual

43

Image 43
Contents Operation and Maintenance Manual Manual Part Number WarrantySafety Notices AcknowledgementsWA R N I N G EnglishIs not in use Español Français Deutsch Italiano Veiligheidsvoorzieningen niet meer DutchFollowing symbols are used on the equipment Contents Overview InstallationOperation Front Panel Controls and Initial StartupPump Maintenance MaintenanceIndex TroubleshootingOperation and Maintenance Manual Introduction Operation and Maintenance Manual System Description Technical Information 355 SCD SpecificationsTechnical Information 255 NCD Power requirements Physical SpecificationsVoltage Sulfur Chemiluminescence Detector Theory of OperationNitrogen Chemiluminescence Detector Dual Plasma Burner with the 355 SCD Dual Plasma ControllerDual Plasma Burner Description of Major ComponentsCross-Section of the Dual Plasma Burner for the 355 SCD Cross-Section of the Dual Plasma Burner for the 255 NCD Chemiluminescence Reaction Cell and Photomultiplier Tube Ozone Generator355 SCD Left Side 355 SCD Right Side 255 NCD Left Side View Vacuum Pump Pressure TransducerNCD Reaction Cell Chemical TrapOil Coalescing Filter FID Adapter OptionalNCD Photomultiplier Tube and Cooler Schematic for 355 SCD Schematic for 255 NCD, in Nitrogen Mode Schematic for 255 NCD, in Nitrosamine Mode Operation and Maintenance Manual Installation Overview Selecting a Location Power RequirementsCombustion Gas Requirements Environmental ConsiderationsTwo-stage pressure regulators Venting Gases Unpack and Inspect the Instrument Required Installation ToolsInstalling the Edwards RV5 Pump Oil-Sealed Vacuum Pump Set Up the Vacuum PumpWA R N I N G Ard RV5 Oil-Sealed Vacuum Pump and Associated Traps Front Side RV5 Oil- Sealed Vacuum Pump and Associated Traps Top Setting the Gas Ballast Position RV5 Installing the Welch Dry Piston Vacuum Pump Operation Notes Welch Pump Welch Dry Piston Pump Operation and Maintenance Manual SCD 230 V Unit Rear Panel Diagram Connect the Power CordInstall the Dry Compressed Air or O2 Supply Install the Signal Output Cables Additional information Install the Dual Plasma BurnerDual Plasma Controller Rear Panel Install the Dual Plasma ControllerCapillary Columns Install Column ConnectionsPacked Columns and Columns with an Outside Diameter 0.8 mm Install the Transfer Line Front Panel Controls and Initial Startup Detector Front Panel Controls Power ControlsDisplay Output Controls Signal ControlsDual Plasma Controller Front Panel Dual Plasma Controller ControlsVacuum Test Initial StartupRecorder Test Initial Checkout Detector Interface SetupIgnition Operation Start-Up Procedure Coking Detector OperationDetector Stability and Response Column BleedFluctuating Pressures Contaminated GasesTypical Operating Conditions Typical Operating ConditionsExpected Detection Limits for Chromatographic Conditions Detection LimitsDaily Shutdown Instrument Shut-DownComplete Shutdown Special Operating Modes Using the 255 NCD in Nitrosamine ModeOperation and Maintenance Manual Operation and Maintenance Manual Maintenance Operating Life of Components for Edwards RV5 Vacuum Pump Pump MaintenanceCleaning the Detector Changing the Oil Mist Filter RV5 Equipment Reaction Cell CleaningReaction Cell, PMT Housing and PMT Socket WA R N I N G Flow Sensor Calibration MDL Detector SensitivitySCD Proper Ferrule Orientation to the Large Ceramic Tube Orientation of the Double Taper Ferrule Tightening the Heater Swivel Nut Operation and Maintenance Manual NCD Proper Ferrule Orientation to the Large Quartz Tube Burner Assembly Detail Tube Replacement for the SCD Positioning the Upper Tube in the Union Fitting 100 Operation and Maintenance Manual Operation and Maintenance Manual 101 Tube Replacement for the NCD Operation and Maintenance Manual 103 NCD Tube Replacement Detail Troubleshooting Solving Detector Problems Dual Plasma Controller Fuses Power ProblemsDetector Fuse Vacuum Pump FuseFuse Positions on the Power Supply Board High Voltage Transformer Ozone Generation ProblemsPlugged Restrictor Lines Response Problems Temperature Reading Problems Troubleshooting Detector Issues Diagnosing General ProblemsDetector gases Troubleshooting Pump Issues Troubleshooting Burner Issues Index Agilent Technologies, Inc