with air as the oxidant in the SCD and oxygen as the oxidant in the NCD.
For the SCD, detector gases must be sulfur free (<1ppb) for proper Detector operation. In general, bottled air is preferred to “house” air from a compressor because compressors tend to generate lower quality air. In addition, pressure fluctuations induced by the air generator may be detrimental to Detector performance.
For the NCD and GC, use gases that are low in nitrogen and total hydrocarbons. Use gases with an "instrument" or "chromatographic" purity rating. Agilent recommends gases with a purity range of 99.995% to 99.9995%. Do not use
Use of appropriate traps on carrier, hydrogen and oxidant sources to improve support gas quality, such as a sulfur trap gas purifier, is recommended. The oxidant used to supply the ozone generator should be dry to prevent internal corrosion. Use of a moisture trap or dryer is recommended.
Two-stage pressure regulators
The combustion gases must be supplied to the Controller at a pressure of
25 psig or less. Use a
If you use gases with a
Particle Filtration
WARNING | Ozone is a hazardous gas and a strong oxidant. Minimize exposure to ozone by using the | |
instrument in a | ||
| ||
| fume hood. Turn off the ozone generator when the instrument is not in use. | |
|
| |
| Supply Tubing for Combustion Gases | |
| Use preconditioned and cleaned copper or stainless steel tubing to supply | |
| gases to the 355 SCD and 255 NCD. Do not use ordinary copper tubing since it | |
| contains oils and other contaminants. Plastic tubing is not recommended since | |
| it is permeable to oxygen and other contaminants that can damage columns | |
| and detectors or cause elevated background. Teflon tubing may be acceptable | |
| in some clean environments. |
Operation and Maintenance Manual | 43 |