8Clean any deposits on the optical filter using a soft cloth or Kimwipe dampened with methanol or deionized (DI) water. Do not leave fingerprints or fibers on the cleaned filter. Deposits inside the reaction cell can be cleaned in the same manner, however, care must be taken to avoid bending the ozone inlet tube that extends into the cell.
9Reseat the pass filter into the PMT housing. Confirm that the two
10To
11Reconnect the PMT high voltage cable to the MHV connector (longer connector) and the PMT signal cable to the BNC connector (shorter connector) at the back of the socket assembly (see Figure 23).
WA R N I N G | The high voltage and signal cables must be attached to the proper connectors on the pmt | |
socket; damage to the pmt will occur if the cables are not properly connected. | ||
| ||
|
|
12While the side panel is still removed, reconnect the AC power cord to the Detector and switch the front panel power to ON. To ensure that the
13If no change in the baseline is observed, check the system for vacuum leaks by turning on the vacuum pump and monitoring the pressure as described in Section 7. If the pressure readings are comparable to
Operation and Maintenance Manual | 87 |