CAUTION: This device should only be operated with the type of power source indicated on the instrument nameplate. If you are not sure of the type of power service available, please consult your local power company.
4.1.1. AC Powered Units
Plug the supplied power cable into the receptacle on the CMS back panel and then connect the power cable to a grounded AC outlet. The AC powered version of the CMS features a self adjusting power supply that automatically adapts to power supplies between 90 and 250 VAC.
4.1.2. DC Powered Units
The DC terminal block features two bus inputs as shown in Figure 4.1 below. This allows the user to connect only one DC power source, or connect two DC power sources where the second source functions as a backup.
When connecting the CMS to your DC power source, first remove the clear protective cover from the DC terminal block, attach the wires from the
Figure 4.1: Terminal Block Assembly (DC Units Only)
4.2. Configure Option Switches
When the CMS is shipped from the factory, the Option Switches are configured for 9600 baud, 8
The Option Switches should be configured to match the parameters your control device will use