Programming | Voltage | <0.05% + 10mV | |
|
| ||
Accuracy, 12months, |
|
| |
(@ 25 °C ± 5 °C) |
|
| |
±(% of output+offset) | Current | <0.2% +10mA | |
| |||
|
|
| |
Readback Accuracy | Voltage | <0.05%+15mV(<20V), | |
<0.05%+120mV(20V)= | |||
| |||
12months, |
|
| |
(25 °C ± 5 °C) |
|
| |
|
| ||
±(% of output+offset) | Current | <0.1% +15mA | |
| |||
|
|
| |
Ripple | Voltage | 1mVrms/3mVp=-p | |
|
| ||
(20Hz ~20MHz) | Current | 5mArms | |
| |||
|
|
| |
Temperature | Voltage | <0.02%+5mV | |
|
| ||
Coefficient, |
|
| |
(0 °C ~ 40 °C) |
|
| |
±(% of output+offset) | Current | <0.1% +5mA | |
| |||
|
|
| |
Readback | Voltage | <0.02%+15mV(<20V), | |
<0.02%+120mV(20V)= | |||
| |||
Yemperature, |
|
| |
Coefficient, |
|
| |
|
| ||
±(% of output+offset) | Current | <0.1% +15mA | |
| |||
|
|
|
2.2 Supplemental Characteristics
State Storage Memory
Sixteen (16)
Recommended Calibration Interval
1 year
AC Input Ratings (selectable via switch on the rear panel)
Option OP1: 220VAC ± 10%, 47 to 63 Hz
Option OP2: 110 VAC ± 10%, 47 to 63 Hz
Maximum input power
350VA
Cooling
Fan cooled
Operating Temperature
32 to 104 °F (0 to 40 °C) for full rated output
Storage Temperature
13