B&K 1786B, 1788, 1785B, 1787B Calibrating the voltage value 29H, Calibrate the current value 2BH

Page 24

1st byte

Start bit(AAH)

2nd byte

Address(0~0XFE)

3rd byte

Command(28H)

4th byte

Calibration protection state

5th byte

System reserve

26th byte

Check sum

10. Calibrating the voltage value (29H)

1st byte

Start bit(AAH)

2nd byte

Address(0~0XFE)

3rd byte

Command(29H)

4th byte

Calibrated voltage points(point 1-3)

5th to 25th byte

System reserve

26th byte

Check sum

 

 

Note: To calibrate the 3 points of voltage sequentially.

11. Sending the present output voltage to calibration program (2AH)

1st byte

Start bit (AAH)

 

 

2nd byte

Address(0~0XFE)

3rd byte

Command(2AH)

4th byte

The byte 0 of present voltage value

5th byte

The byte 1 of present voltage value

6th byte

The byte 2 of present voltage value

7th byte

The byte 3 of present voltage value

8th to 25th byte

System reserve

26th byte

Check sum

 

 

12. Calibrate the current value (2BH)

1st byte

Start bit(AAH)

2nd byte

Address(0-0XFE)

3rd byte

Command(2BH)

4th byte

Calibrated current points( point 1-2)

5th to 25th byte

System reserve

26th byte

Check sum

 

 

24

Image 24
Contents Page Content About your safety Quick ReferenceGeneral information Protection from electric shockCertification and Warranty IntroductionCertification Front Panel at a Glance Function keys description Menu descriptionDisplay annunciators Rear Panel at a Glance CalConnect the power cord and turn on the power supply Quick StartPreliminary Checkout Check the list of supplied itemsOutput Checkout Voltage Output CheckoutCurrent Output Checkout If the Power Supply Does Not Turn On To Adjust the Carrying HandleVerify that there is AC power to the power supply Verify the power-line voltage settingTo Rack Mount the Instrument Bench-top viewing positions Carrying positionSpecifications SpecificationsSupplemental Characteristics Maximum input powerUnit mm Environmental ConditionsFront-panel Operation Front-panel Operation OverviewConstant Voltage Operation Constant Current Operation Saving and Recalling OperationMenu Operation Set Maximum voltageMAX VoltSetting the Key SoundKEY Sound Setting the Baud RateBAUDRATEInitiating the Output stateINIT OUT Initiating the Output Voltage Init VoltSetting Address Address Setting password for function keysKEY LockRemote Operation Mode Communication settingIT-E131 Communication cable Setting the remote control mode 20H Frame formatCommunication protocol DescriptionSetting the output state ON/OFF 21H Setting the maximum output voltage 22HSetting the output voltage 23H Setting the output current 24H Setting the communication address 25HEntering the calibration mode 27H Reading the calibration state 28HCalibrating the voltage value 29H Calibrate the current value 2BHSetting calibration information 2EH Save the calibration data to Eeprom 2DHReading calibration information 2FH Restore the factory default calibration data 32H Enable the local key 37HReturn information of command operation in power supply 12H Calibration Configuration Voltage calibrationCurrent calibration Finish calibratingReset to the default calibration settings InstallationPV1785B1788 Software Functions of PV1785B1788 1Configure the system Status barSetting Voltage and current Quickly settingProgram setting 4. GO/NG Test FunctionPresent Voltage/Current Chart Chart DescriptionSave and Open Page Limited One-Year Warranty Service Information USA