B&K 1788, 1786B, 1785B Setting the output state ON/OFF 21H, Setting the maximum output voltage 22H

Page 21

3rd byte

Command ( 20H)

4th byte

Operation mode(0 represent front panel operation mode, 1

 

represent remote operation mode)

5th to 25th byte

System reserve

26th byte

Check sum

 

 

 

 

Note: You can not control the power supply from the front panel when the power supply is in calibration mode.

2. Setting the output state ON/OFF (21H)

1st byte

Start bit (AAH )

2nd byte

Address(0~0XFE)

3rd byte

Command (21H)

4th byte

Output state(0 is OFF, 1 is ON)

5th to 25th byte

System reserve

26th byte

Check sum

3. Setting the maximum output voltage (22H)

1st byte

Start bit (AAH )

2nd byte

Address(0~0XFE)

3rd byte

Command (21H)

4th byte

The lowest byte of voltage upper limit

5th byte

The lower byte of voltage upper limit

6th byte

The higher byte of voltage upper limit

7th byte

The highest byte of voltage upper limit

8th to 25th byte

System reserve

26th byte

Check sum

Note: We use 4 bytes of Hex number to represent a maximum voltage value. For example the maximum voltage is 16.000V, the hex code of 16.000 is 0X00003EB0, so the 4th byte is 0XB0, 5th bye is 0X3E, 6th byte is 0X00, 7th byte is 0X00.

4. Setting the output voltage (23H)

1st byte

Start bit ( AAH )

2nd byte

Address(0~0XFE)

3rd byte

Command( 23H)

4th byte

The byte 0 of output voltage value

5th byte

The byte 1 of output voltage value

6th byte

The higher byte of output voltage value

21

Image 21
Contents Page Content Quick Reference About your safetyGeneral information Protection from electric shockCertification and Warranty IntroductionCertification Front Panel at a Glance Function keys description Menu descriptionDisplay annunciators Cal Rear Panel at a GlanceQuick Start Connect the power cord and turn on the power supplyPreliminary Checkout Check the list of supplied itemsOutput Checkout Voltage Output CheckoutCurrent Output Checkout To Adjust the Carrying Handle If the Power Supply Does Not Turn OnVerify that there is AC power to the power supply Verify the power-line voltage settingBench-top viewing positions Carrying position To Rack Mount the InstrumentSpecifications SpecificationsMaximum input power Supplemental CharacteristicsEnvironmental Conditions Unit mmFront-panel Operation Front-panel Operation OverviewConstant Voltage Operation Saving and Recalling Operation Constant Current OperationMenu Operation Set Maximum voltageMAX VoltSetting the Baud RateBAUDRATE Setting the Key SoundKEY SoundInitiating the Output stateINIT OUT Initiating the Output Voltage Init VoltSetting password for function keysKEY Lock Setting Address AddressRemote Operation Mode Communication settingIT-E131 Communication cable Frame format Setting the remote control mode 20HCommunication protocol DescriptionSetting the output state ON/OFF 21H Setting the maximum output voltage 22HSetting the output voltage 23H Setting the communication address 25H Setting the output current 24HReading the calibration state 28H Entering the calibration mode 27HCalibrate the current value 2BH Calibrating the voltage value 29HSetting calibration information 2EH Save the calibration data to Eeprom 2DHReading calibration information 2FH Restore the factory default calibration data 32H Enable the local key 37HReturn information of command operation in power supply 12H Calibration Voltage calibration ConfigurationCurrent calibration Finish calibratingReset to the default calibration settings InstallationPV1785B1788 Software Functions of PV1785B1788 Status bar 1Configure the systemQuickly setting Setting Voltage and current4. GO/NG Test Function Program settingPresent Voltage/Current Chart Chart DescriptionSave and Open Page Limited One-Year Warranty Service Information USA