B&K 1786B Frame format, Communication protocol, Description, Setting the remote control mode 20H

Page 20

4.3 Frame format

Frame length is 26 bytes, the format is as follows:

Start

Address

Command

4-25 bytes are information content

Check sum

Description:

1. Start bit is AAH, occupies a byte.

2. Address range is 0 to FE, occupies a byte.

3. Command occupies a byte.

a. 20H----Setting the remote control mode

b. 21H----Setting the output ON/OFF state

c. 22H----Setting the maximum output voltage

d. 23H----Setting the output voltage

e. 24H----Setting the output current

f. 25H----Setting the communication address

g. 26H----Reading the present current/voltage, maximum voltage, setup voltage/current and

operation states of the power supply. h. 27H----Enter the calibration mode

i. 28H----Reading the calibration mode state

j. 29H----Calibrate voltage value.

k. 2AH----Sending the actual output voltage to calibration program.

l. 2BH----Calibrate current value.

m.2CH----Sending the actual output current to calibration program.

n.2DH----Save the calibration data to EEPROM.

o.2EH----Setting calibration information.

p.2FH----Reading calibration information.

q.31H----Reading product’s model, series number and version information.

r.32H----Restoring the factory default calibration data.

s.37H----Enable the local key.

t.12H---- The return information of command operation in power supply.

Note: You must change the power supply to remote control mode firstly, then you can control the power supply output by computer. The command for remote control is 20H?

If you want to calibrate the power supply, set the calibration information or want to set the product serial number, you must set the calibration protection mode to OFF state firstly, the command for calibration protection is 27H.

When the power supply is in calibration mode, it is not allowed to change the output

4.4th to 25th bytes are information content

5.26th byte is check sum, the sum of the former 25 bytes.

4.4Communication protocol

1. Setting the remote control mode (20H)

1st byte

2nd byte

Start bit( AAH )

Address(0~0XFE)

20

Image 20
Contents Page Content About your safety Quick ReferenceGeneral information Protection from electric shockCertification Certification and WarrantyIntroduction Front Panel at a Glance Display annunciators Function keys descriptionMenu description Rear Panel at a Glance CalConnect the power cord and turn on the power supply Quick StartPreliminary Checkout Check the list of supplied itemsCurrent Output Checkout Output CheckoutVoltage Output Checkout If the Power Supply Does Not Turn On To Adjust the Carrying HandleVerify that there is AC power to the power supply Verify the power-line voltage settingTo Rack Mount the Instrument Bench-top viewing positions Carrying positionSpecifications SpecificationsSupplemental Characteristics Maximum input powerUnit mm Environmental ConditionsConstant Voltage Operation Front-panel OperationFront-panel Operation Overview Constant Current Operation Saving and Recalling OperationMenu Operation Set Maximum voltageMAX VoltSetting the Key SoundKEY Sound Setting the Baud RateBAUDRATEInitiating the Output stateINIT OUT Initiating the Output Voltage Init VoltSetting Address Address Setting password for function keysKEY LockIT-E131 Communication cable Remote Operation ModeCommunication setting Setting the remote control mode 20H Frame formatCommunication protocol DescriptionSetting the output voltage 23H Setting the output state ON/OFF 21HSetting the maximum output voltage 22H Setting the output current 24H Setting the communication address 25HEntering the calibration mode 27H Reading the calibration state 28HCalibrating the voltage value 29H Calibrate the current value 2BHReading calibration information 2FH Setting calibration information 2EHSave the calibration data to Eeprom 2DH Return information of command operation in power supply 12H Restore the factory default calibration data 32HEnable the local key 37H Calibration Configuration Voltage calibrationCurrent calibration Finish calibratingPV1785B1788 Software Reset to the default calibration settingsInstallation Functions of PV1785B1788 1Configure the system Status barSetting Voltage and current Quickly settingProgram setting 4. GO/NG Test FunctionSave and Open Present Voltage/Current ChartChart Description Page Limited One-Year Warranty Service Information USA