5.1 Configuration

Start the PC1785B_1788 System, the windows is as follows:

Voltage calibration

Current calibration

The diagram of connecting the multi- meter to the calibrated power supply

The diagram of connecting the multi-

meter to the calibrated power supply

Before the calibration, please click “Config” button to set the COM interface, address and baud rate. The default Com port is COM, and the default baud rate is 9600. Please make sure that the baud rate and address in the software and in power supply are the same.

5.2 Voltage calibration

Step 1: Prepare to use multi-meter to measure the output voltage, and connect the voltage meter to the power supply according to the diagram in the above windows.

Step 2: Click “Linked” button. It will go into Calibration mode, the Rmt and Cal indicators will light up.

Step 3: Fill the first voltage value you measured from the multi-meter to the blank of “0.000V”, and click “OK” button to confirm the value.

Step 4: Fill the second voltage value you measured from the multi-meter to the blank of “0.000V”, and click “OK” button to confirm the value.

Step 5: Fill the third voltage value you measured from the multi-meter to the blank of “0.000V”, and click “OK” button to confirm the value.

5.3 Current calibration

Step 1: Prepare to use multi-meter to measure the short circuit current, and connect the current meter to the power supply according to the diagram in the above windows.

Step 2: Click “Linked” button, it will go into Calibration mode, the Rmt and Cal indicators will light up.

Step 3: Fill the first current value you measured from the multi-meter to the blank of “0.000A”, and click “OK” button to confirm the value.

Step 4: Fill the second current value you measured from the multi-meter to the blank of “0.000A”, and click “OK” button to confirm the value.

5.4 Finish calibrating

28

Image 28
Contents Page Content About your safety Quick ReferenceGeneral information Protection from electric shockIntroduction Certification and WarrantyCertification Front Panel at a Glance Menu description Function keys descriptionDisplay annunciators Rear Panel at a Glance CalConnect the power cord and turn on the power supply Quick StartPreliminary Checkout Check the list of supplied itemsVoltage Output Checkout Output CheckoutCurrent Output Checkout If the Power Supply Does Not Turn On To Adjust the Carrying HandleVerify that there is AC power to the power supply Verify the power-line voltage settingTo Rack Mount the Instrument Bench-top viewing positions Carrying positionSpecifications SpecificationsSupplemental Characteristics Maximum input powerUnit mm Environmental ConditionsFront-panel Operation Overview Front-panel OperationConstant Voltage Operation Constant Current Operation Saving and Recalling OperationMenu Operation Set Maximum voltageMAX VoltSetting the Key SoundKEY Sound Setting the Baud RateBAUDRATEInitiating the Output stateINIT OUT Initiating the Output Voltage Init VoltSetting Address Address Setting password for function keysKEY LockCommunication setting Remote Operation ModeIT-E131 Communication cable Setting the remote control mode 20H Frame formatCommunication protocol DescriptionSetting the maximum output voltage 22H Setting the output state ON/OFF 21HSetting the output voltage 23H Setting the output current 24H Setting the communication address 25HEntering the calibration mode 27H Reading the calibration state 28HCalibrating the voltage value 29H Calibrate the current value 2BHSave the calibration data to Eeprom 2DH Setting calibration information 2EHReading calibration information 2FH Enable the local key 37H Restore the factory default calibration data 32HReturn information of command operation in power supply 12H Calibration Configuration Voltage calibrationCurrent calibration Finish calibratingInstallation Reset to the default calibration settingsPV1785B1788 Software Functions of PV1785B1788 1Configure the system Status barSetting Voltage and current Quickly settingProgram setting 4. GO/NG Test FunctionChart Description Present Voltage/Current ChartSave and Open Page Limited One-Year Warranty Service Information USA