B&K 1788, 1786B, 1785B, 1787B instruction manual Voltage Output Checkout, Current Output Checkout

Page 9
Out on/off
Out on/off

1.2 Output Checkout

The following procedures check to ensure that the power supply develops its rated outputs and properly responds to operation from the front panel.

1.2.1 Voltage Output Checkout

The following steps verify basic voltage functions without load.

1.Turn on the power supply.

2.Enable the outputs.

Press Out on/off key to let the ON annunciator and the CV annunciator turn on to light.

Notice: if the voltage value flash, then the power supply is in Set mode, ‘‘Set mode’’ means that the VFD display shows the setting output voltage and current. Or the power supply is in Meter mode, ‘Meter mode” means that the VFD display shows the actual output voltage and current.

3. Check that the front-panel voltmeter properly responds to number keys

Set some different voltage values, then wait till the Meter mode to check if the VFD displayed voltage value is the same as the set voltage value, and to check if the VFD displayed current value is nearly zero.

4. Ensure that the voltage can be adjusted from zero to the full rated value.

1.2.2 Current Output Checkout

The following steps check basic current functions with a short across the power supply’s output.

1.Turn on the power supply.

2.Disable the output

Press key to ensure that the output is disabled. The ON annunciator is turned off.

3.Connect a short across (+) and (-) output terminals with an insulated test lead. Use a wire size sufficient to handle the maximum current.

Warning: To satisfy safety requirements, load wires must be heavy enough not to overheat when carrying the maximum short-circuit output current of the power supply. If there is more than one load, then any pair of load wires must be capable of safety carrying the full-rated current of the power supply.

4. Enable the output.

Press key to ensure that the output is enabled. The ON annunciator is turned on.

5.Adjust the voltage value to 1.0 volt.

Adjust the voltage to 1.0 volt to ensure the power supply is in CC operation mode. The CC annunciator will turn on.

6.Adjust the current value.

Set some different voltage values, then wait till the Meter mode to check if the VFD displayed current value is the same as the set voltage value, and to check if the VFD displayed voltage value is nearly

9

Image 9
Contents Page Content Quick Reference About your safetyGeneral information Protection from electric shockCertification and Warranty IntroductionCertification Front Panel at a Glance Function keys description Menu descriptionDisplay annunciators Cal Rear Panel at a GlanceQuick Start Connect the power cord and turn on the power supplyPreliminary Checkout Check the list of supplied itemsOutput Checkout Voltage Output CheckoutCurrent Output Checkout To Adjust the Carrying Handle If the Power Supply Does Not Turn OnVerify that there is AC power to the power supply Verify the power-line voltage settingBench-top viewing positions Carrying position To Rack Mount the InstrumentSpecifications SpecificationsMaximum input power Supplemental CharacteristicsEnvironmental Conditions Unit mmFront-panel Operation Front-panel Operation OverviewConstant Voltage Operation Saving and Recalling Operation Constant Current OperationMenu Operation Set Maximum voltageMAX VoltSetting the Baud RateBAUDRATE Setting the Key SoundKEY SoundInitiating the Output stateINIT OUT Initiating the Output Voltage Init VoltSetting password for function keysKEY Lock Setting Address AddressRemote Operation Mode Communication settingIT-E131 Communication cable Frame format Setting the remote control mode 20HCommunication protocol DescriptionSetting the output state ON/OFF 21H Setting the maximum output voltage 22HSetting the output voltage 23H Setting the communication address 25H Setting the output current 24HReading the calibration state 28H Entering the calibration mode 27HCalibrate the current value 2BH Calibrating the voltage value 29HSetting calibration information 2EH Save the calibration data to Eeprom 2DHReading calibration information 2FH Restore the factory default calibration data 32H Enable the local key 37HReturn information of command operation in power supply 12H Calibration Voltage calibration ConfigurationCurrent calibration Finish calibratingReset to the default calibration settings InstallationPV1785B1788 Software Functions of PV1785B1788 Status bar 1Configure the systemQuickly setting Setting Voltage and current4. GO/NG Test Function Program settingPresent Voltage/Current Chart Chart DescriptionSave and Open Page Limited One-Year Warranty Service Information USA