Step2. Press

V-Set

key.

 

Step3. Use the numeric keys 0 to

9 or ? and ? keys to change the voltage value.

Step4. Press

Enter

to confirm the value

3.3 Constant Current Operation

The constant current output range is from 0A to the maximum current value of each type. It is very easy for you to set the constant current output.

Step1. Power on the Power Supply

Step2. Press I-Setkey

Step3. Use the numeric keys 0 to 9 or use ? and ? keys to change the current value Step4. Press Enter key to confirm the value

3.4 Saving and Recalling Operation

You can store up to 16 different output states in storage register locations (1 to 16). Each output state includes Constant voltage value, Constant current value and Maximum output voltage value. When shipped from factory, storage locations “1” through “16” are empty. You can recall the saved settings by Recall function.

Step1. After you setting an output state (CV value, CC value and Maximum voltage), press Save key. Step2.Use the numeric keys 0 to 9 or ? and ? keys to select the memory location (the range

is 1 to 16) which you want to store in.

Step3. Press

Enter

to confirm the memory location.

Step4. Press

Recall

key.

Step5. Use numeric keys 0 to 9 or ? and ? keys to select the states which you want to recall.

Step6. Press Enter key to confirm. Then the saved settings will come on.

Note: 1. If the function keys were locked by password, you need to enter the correct password after you press function keys (V-set, I-set, Save, Recall and Menu), then you can do the settings.

2. If you want to cancel a function operation (V-set, I-set, Save, Recall or Menu), just

press Esc key to exit.

3.5 Menu Operation

nSet Maximum voltage(>MAX VOLT)

Please be well known that the Max voltage value should be in the range of each type of IT68xx Power supply.

Step1. Press Menu key.

16

Image 16
Contents Page Content About your safety Quick ReferenceGeneral information Protection from electric shockIntroduction Certification and WarrantyCertification Front Panel at a Glance Menu description Function keys descriptionDisplay annunciators Rear Panel at a Glance CalConnect the power cord and turn on the power supply Quick StartPreliminary Checkout Check the list of supplied itemsVoltage Output Checkout Output CheckoutCurrent Output Checkout If the Power Supply Does Not Turn On To Adjust the Carrying HandleVerify that there is AC power to the power supply Verify the power-line voltage settingTo Rack Mount the Instrument Bench-top viewing positions Carrying positionSpecifications SpecificationsSupplemental Characteristics Maximum input powerUnit mm Environmental ConditionsFront-panel Operation Overview Front-panel OperationConstant Voltage Operation Constant Current Operation Saving and Recalling OperationMenu Operation Set Maximum voltageMAX VoltSetting the Key SoundKEY Sound Setting the Baud RateBAUDRATEInitiating the Output stateINIT OUT Initiating the Output Voltage Init VoltSetting Address Address Setting password for function keysKEY LockCommunication setting Remote Operation ModeIT-E131 Communication cable Setting the remote control mode 20H Frame formatCommunication protocol DescriptionSetting the maximum output voltage 22H Setting the output state ON/OFF 21HSetting the output voltage 23H Setting the output current 24H Setting the communication address 25HEntering the calibration mode 27H Reading the calibration state 28HCalibrating the voltage value 29H Calibrate the current value 2BHSave the calibration data to Eeprom 2DH Setting calibration information 2EHReading calibration information 2FH Enable the local key 37H Restore the factory default calibration data 32HReturn information of command operation in power supply 12H Calibration Configuration Voltage calibrationCurrent calibration Finish calibratingInstallation Reset to the default calibration settingsPV1785B1788 Software Functions of PV1785B1788 1Configure the system Status barSetting Voltage and current Quickly settingProgram setting 4. GO/NG Test FunctionChart Description Present Voltage/Current ChartSave and Open Page Limited One-Year Warranty Service Information USA