0.25Ω (6023A)
2.0Ω (6028A)
load resistor that is capable of safely dissipating 250 watts. Proceed as follows:
a.Connect the test equipment as shown in Figure
b.Switch the unit's power on and turn the output voltage all the way up.
c.Turn up output current to: 29.0Vdc (6023A)
10Vdc (6028A)
Check that the unit's CC LED remains lighted.
d.Check that the rms noise current measured by the current probe and rms voltmeter is no more than: 15mA rms (6023A).
5mA rms (6028A)
Figure 2-8. CC PARD Test Setup
22