Agilent Technologies AGILENT MODELS 6023A and 6028A Accuracy, Relay Resistance 500Ω ± 10%

Page 80
@25°C
CV: 0.38% + 16mV (6023A)
CC: 0.43% + 115mA (6023A)

Table A-1. Specifications, Option 002

Remote Programming

Resistance Programming: 0 to 4K ohm provides 0 to maximum rated voltage or current output.

Accuracy:

@25°C

 

 

 

CV: 0.5% + 12mV (6023A)

0.5%

+ 70mV (6028A)

 

CC: 1.0% + 110mA (6023A)

1.0%

+ 500mA (6028A)

Voltage Programming: 0 to 5V provides 0 to maximum rated voltage or current output.

Accuracy:

@25°C

 

 

CV: 0.25% + 12mV (6023A)

0.33% + 70mV (6028A)

 

CC: 0.30% + 110mA (6023A)

0.36% ± 500mA (6028A)

Current Programming: 0 to 2mA current sink provides 0 to maximum rated voltage or current output.

Accuracy:

0.43% ± 71mV (6028A)

0.50% + 500mV (6028A)

Input Compliance Voltage: ± 1V

Current Programming Enable:

Relays K2 (CV) and K1 (CC) are biased from the Control Isolator Bias input (See Remote Shutdown and OVP Clear).

Relay Bias Voltage: +4V minimum + 7V maximum

Relay Resistance: 500Ω ± 10%

Note

For Control Isolator Bias voltages greater than 7V, a series resistor must be used to maintain the relay bias voltage within specified limits.

Enabling either relay is accomplished by bringing CV or CC enable line to Control Isolator Bias common via a suitable driver; maximum driver off-state leakage =5mA.

Output Voltage and Current Monitor: 0 to 5V output indicates 0 to maximum rated output voltage or current.

Accuracy:

@25°C

 

 

TC: 10ppm +0.2mV/°C

 

CV: 0.25% + 2mV (6023A)

0.39% + 15mV (6028A)

 

CC: 0.30% + 15mA (6023A)

0.36% + 20mA (6028A)

TC: 70ppm +1.5mA/°C

Output Impedance: 10.2 k ohm ± 5%

 

 

 

Temperature Coefficient:

 

 

 

 

CV: 70 ppm/°C +600∝V/°C (6023A)

12.5 ppm/°C +810∝V/°C (6028A)

 

CC: 100 ppm/°C + 2.0mA/°C (6023A)

47 ppm/°C + 1.6mA/°C (6028A)

Status Indicators:

Status Isolator Bias input (referred to Status Isolator Common).

80

Image 80
Contents Autoranging Certification Safety Summary Safety Symbol Definitions Printing HistoryTable of Contents Replaceable Parts Troubleshooting Circuit DiagramsSafety Considerations ScopeManual Revisions Item DescriptionCalibration Procedure IntroductionTest Equipment Required Operation Verification TestsType Required Characteristics USE Recommended Model Test Equipment RequiredInitial Setup 1KVACalibration Procedure MONVout IoutCR7 Performance TestsMeasurement Techniques Current-Monitoring Resistor Setup Constant Voltage CV TestsBasic Test Setup RMS Measurement Test Setup, CV Pard Test Peak-To-Peak Measurement Test Setup, CV Pard Test 6023A 6028A KHz Noise, CV Peak-to-Peak PardLoad Transient Recovery Waveform Constant Current CC Tests CC Pard Test Setup Troubleshooting Initial Troubleshooting ProceduresControl Board Test Connector, A2J7 PIN no Signal NameVdc WAVEFORM/CONDITIONS SourceRepair and Replacement Electrostatic ProtectionA2 Control Board Removal A4 Power Mesh Board Removal Overall Troubleshooting ProcedureA3 Front-Panel Board Removal A1 Main Board RemovalUsing the Tables SetupNode MeasurementMain Troubleshooting Setup Main Troubleshooting SetupModified Mains Cord Set For Troubleshooting Troubleshooting No-Output Failures Front-Panel TroubleshootingBoard PWM-ON PWM-OFF Defective Check Functional CircuitsA2J7-26 A2J7-25 Troubleshooting Bias Supplies Symptoms Defective Circuit Check ComponentsSymptoms Defective Check Functional Circuits Board Node + Measurement SourceNode + N0DE Node +Measurement Source Power Section Blocks Troubleshooting AC-Turn-on CircuitsSetup Measurement Source Setup MeasurementPower Limit Troubleshooting PWM & ClockRelay Enable Troubleshooting DC-To-DC Converter + OUTPWM-ON PWM-OFFTroubleshooting CV Circuit Troubleshooting CC CircuitA4U4OUT Troubleshooting Down ProgrammerTroubleshooting OVP Circuit Waveforms SET Voltage Vdc Node + Page Autoranging Power Block Diagram OverviewBlock Diagram Simplified Schematic Simplified Schematic DC-to-DC Converter Down ProgrammerPfet Control Signals Timing Diagram Constant-Voltage CV CircuitPower-Limit Comparator Constant-Current CC CircuitOvervoltage Protection OVP Circuit Control-Voltage ComparatorInitial-Ramp Circuit Pulse-Width Modulator PWMBias Voltage Detector AC-Surge Dropout DetectorSecond-Delay Circuit Display CircuitsPage Replaceable Parts Reference DesignatorsOrdering Information Description AbbreviationsElectrical Parts CR1CR2 CR3Mechanical Parts XA2P1XA2P2 Control Board Assembly CR3 Page Page Page VR5 VR3VR4 DS5 CR5DS1 VR2 IC A/D Cmos 3-1/2 DGTVR1 NPN SI PfetTS1 Chassis ElectricalA1TB1 A2 Board Mechanical PartsA3 Board Mechanical Parts A4 Board Mechanical PartsOption 220 220V Operation Option 240 240V OperationComponent Location and Circuit Diagrams Schematic Diagram Notes Pin locations for other semi-conductors are shown below Top View, Top Covers Removed Main Board A1 Component Location Control Board A2 Component Location Front-Panel Board A3 Component Location Power Mesh Board A4 Component Location Page Page Option 002 Hardware SpecificationsGeneral Information Table A-1. Specifications, Option Remote Programming Input Compliance Voltage ± Current Programming EnableOutput Impedance 10.2 k ohm ± 5% Temperature Coefficient Status IndicatorsRemote Trip and Remote Reset Timing Power-on PresetMaximum Output Voltage logic high + On State logic lowTable A-1. Specifications, Option Pulse Timing Bias Supplies DC Output Ratings 25C ±Short Circuit Output Current Pard TypicalInstallation Connector Assembly ProcedureFigure A-1. Mating Connector Assembly OperationLocal/Remote Programming Resistance Voltage or Current Remote Resistance Programming Figure A-4. Calculating Value of Series Dropping ResistorFigure A-5. Remote Resistance Programming Remote Monitoring Status Indicators Remote ControlRemote Reset OvervoltagePower-On Preset Figure A-8. Remote ControlAC Dropout Buffer Circuit Multiple Supply System ShutdownMaintenance Bias SuppliesTroubleshooting Resistance and Voltage Programming Troubleshooting Current ProgrammingFigure A-11. Troubleshooting Current Programming of CV Mode Figure A-13. Troubleshooting Status Indicators Page Table A-3. Replacement Parts REF. Desig Model no Description100 VR9 101Indicator and Qualifier Symbols OLD Symbol NEW SymbolLogic Symbols and Definitions 102 DefinitionsSchematic Notes Schematic Diagram Notes 103Schematic Notes Figure A-15. Option 002 Board, Component Location 104Figure A-16. Option 002 Board, Schematic Diagram 105 Page 107 Model 6023AModel 6028A From 108