the unit's output voltage. To reduce
a.Connect the test equipment as shown in Figure
b.Turn the unit's power on, and turn up current setting to full output.
c.Turn up output voltage to: 7.0Vdc (6023A)
20Vdc (6028A)
d.Reduce the resistance of the load to draw an output current of: 29.0Adc (6023A)
10Adc (6028A)
Check that the unit's CV LED remains lighted.
e.Set the oscilloscope's input impedance to 50Ω and bandwidth to 20MHz. Adjust the controls to show the 20KHz and higher frequency
f.Check that the
Figure 2-5. Peak-To-Peak Measurement Test Setup, CV PARD Test