Agilent Technologies AGILENT MODELS 6023A and 6028A Introduction, Test Equipment Required

Page 9

2

Calibration and Verification

Introduction

This section provides test and calibration procedures. The operation-verification tests comprise a short procedure to verify that the unit is performing properly, without testing all specified parameters. After troubleshooting and repair of a defective power supply you can usually verify proper operation with the turn-on checkout procedure in the Operating Manual.

Repairs to the A1 main board and the A2 control board can involve circuits which, although functional, may prevent the unit from performing within specified limits. So, after A1 or A2 board repair, decide if recalibration and operation verification tests are needed according to the faults you discover. Use the calibration procedure both to check repairs and for regular maintenance.

When verifying the performance of this instrument as described in this chapter, check only those specifications for which a performance test procedure is included.

Test Equipment Required

Table 2-1 lists the equipment required to perform the tests of this section. You can separately identify the equipment for performance tests, calibration and troubleshooting using the USE column of the table.

Operation Verification Tests

To assure that the unit is performing properly, without testing all specified parameters, first perform the turn-on checkout procedure in the Operating Manual. Then perform the following performance tests, in this section.

CV Load Effect

CC Load Effect

Calibration Procedure

Calibrate the unit twice per year and when required during repair. The following calibration procedures which follow should be performed in the sequence given. Table 2-2 describes in detail these calibration procedures and lists the expected results to which each adjustment must be made.

Note: Some of the calibration procedures for this instrument can be performed independently, and some procedures must be performed together and/or in a prescribed order. If a procedure contains no references to other procedures, you may assume that it can be performed independently.

To return a serviced unit to specifications as quickly as possible with minimal calibration, the technician need only perform calibration procedures that affect the repaired circuit. Table 2-3 lists various power supply circuits with calibration procedures that should be performed after those circuits are serviced.

9

Image 9
Contents Autoranging Certification Safety Summary Printing History Safety Symbol DefinitionsTable of Contents Replaceable Parts Circuit Diagrams TroubleshootingSafety Considerations ScopeItem Description Manual RevisionsIntroduction Calibration ProcedureTest Equipment Required Operation Verification TestsTest Equipment Required Type Required Characteristics USE Recommended Model1KVA Initial SetupMON Calibration ProcedureIout VoutPerformance Tests Measurement TechniquesCR7 Constant Voltage CV Tests Current-Monitoring Resistor SetupBasic Test Setup RMS Measurement Test Setup, CV Pard Test Peak-To-Peak Measurement Test Setup, CV Pard Test KHz Noise, CV Peak-to-Peak Pard 6023A 6028ALoad Transient Recovery Waveform Constant Current CC Tests CC Pard Test Setup Initial Troubleshooting Procedures TroubleshootingPIN no Signal Name Control Board Test Connector, A2J7Vdc WAVEFORM/CONDITIONS SourceElectrostatic Protection Repair and ReplacementA2 Control Board Removal Overall Troubleshooting Procedure A4 Power Mesh Board RemovalA3 Front-Panel Board Removal A1 Main Board RemovalSetup Using the TablesNode MeasurementMain Troubleshooting Setup Main Troubleshooting SetupModified Mains Cord Set For Troubleshooting Front-Panel Troubleshooting Troubleshooting No-Output FailuresPWM-ON PWM-OFF Defective Check Functional Circuits A2J7-26 A2J7-25Board Symptoms Defective Circuit Check Components Troubleshooting Bias SuppliesSymptoms Defective Check Functional Circuits Board Node + Measurement SourceNode + Measurement SourceNode + N0DE Troubleshooting AC-Turn-on Circuits Power Section BlocksSetup Measurement Source Setup MeasurementTroubleshooting PWM & Clock Relay EnablePower Limit + OUT Troubleshooting DC-To-DC ConverterPWM-ON PWM-OFFTroubleshooting CC Circuit Troubleshooting CV CircuitTroubleshooting Down Programmer Troubleshooting OVP CircuitA4U4OUT Waveforms SET Voltage Vdc Node + Page Block Diagram Overview Autoranging PowerBlock Diagram Simplified Schematic Simplified Schematic Down Programmer DC-to-DC ConverterConstant-Voltage CV Circuit Pfet Control Signals Timing DiagramConstant-Current CC Circuit Power-Limit ComparatorOvervoltage Protection OVP Circuit Control-Voltage ComparatorPulse-Width Modulator PWM Initial-Ramp CircuitBias Voltage Detector AC-Surge Dropout DetectorDisplay Circuits Second-Delay CircuitPage Reference Designators Replaceable PartsDescription Abbreviations Ordering InformationCR1 Electrical PartsCR2 CR3XA2P1 XA2P2Mechanical Parts Control Board Assembly CR3 Page Page Page VR3 VR4VR5 CR5 DS1DS5 IC A/D Cmos 3-1/2 DGT VR1VR2 Pfet NPN SIChassis Electrical TS1A2 Board Mechanical Parts A1TB1A3 Board Mechanical Parts A4 Board Mechanical PartsOption 240 240V Operation Option 220 220V OperationComponent Location and Circuit Diagrams Schematic Diagram Notes Pin locations for other semi-conductors are shown below Top View, Top Covers Removed Main Board A1 Component Location Control Board A2 Component Location Front-Panel Board A3 Component Location Power Mesh Board A4 Component Location Page Page Specifications General InformationOption 002 Hardware Input Compliance Voltage ± Current Programming Enable Table A-1. Specifications, Option Remote ProgrammingOutput Impedance 10.2 k ohm ± 5% Temperature Coefficient Status IndicatorsPower-on Preset Remote Trip and Remote Reset TimingMaximum Output Voltage logic high + On State logic lowBias Supplies DC Output Ratings 25C ± Table A-1. Specifications, Option Pulse TimingShort Circuit Output Current Pard TypicalConnector Assembly Procedure InstallationOperation Figure A-1. Mating Connector AssemblyLocal/Remote Programming Resistance Voltage or Current Figure A-4. Calculating Value of Series Dropping Resistor Remote Resistance ProgrammingFigure A-5. Remote Resistance Programming Remote Monitoring Remote Control Status IndicatorsOvervoltage Remote ResetFigure A-8. Remote Control Power-On PresetMultiple Supply System Shutdown AC Dropout Buffer CircuitBias Supplies MaintenanceTroubleshooting Current Programming Troubleshooting Resistance and Voltage ProgrammingFigure A-11. Troubleshooting Current Programming of CV Mode Figure A-13. Troubleshooting Status Indicators Page REF. Desig Model no Description Table A-3. Replacement Parts100 101 VR9OLD Symbol NEW Symbol Indicator and Qualifier SymbolsLogic Symbols and Definitions 102 DefinitionsSchematic Diagram Notes 103 Schematic NotesFigure A-15. Option 002 Board, Component Location 104 Schematic NotesFigure A-16. Option 002 Board, Schematic Diagram 105 Page Model 6023A Model 6028A107 108 From