Troubleshooting
Maintenance described herein is performed with power supplied to the instrument, and protective covers removed. Such maintenance should be performed only by
Introduction
Before attempting to troubleshoot this instrument, ensure that the fault is with the instrument itself and not with an associated circuit. The performance test enables this to be determined without having to remove the covers from the supply.
The most important aspect of troubleshooting is the formulation of a logical approach to locating the source of trouble. A good understanding of the principles of operation is particularly helpful, and it is recommended that Chapter 4 of this manual be reviewed before attempting to troubleshoot the unit. Often the user will then be able to isolate a problem simply by using the operating controls and indicators. Once the principles of operation are understood, refer to the following paragraphs.
Table
If a component is found to be defective, replace it and
Initial Troubleshooting Procedures
If a problem occurs, follow the steps below in sequence:
a.Check that input power is available, and check the power cord and
b.Check that the settings of mode switch A2S1 are correct for the desired mode of operation. (See Operating Manual).
c.Check that all connections to the power supply are secure and that circuits between the supply and external devices are not interrupted.
d.If the power supply fails
Some circuits on the power mesh are connected directly to the ac power line. Exercise extreme caution when working on energized circuits. Energize the supply through an isolation transformer to avoid shorting ac energized circuits through the test instrument's input leads. The isolation transformer must have a power rating of at least 1KVA. During work on energized circuits, the safest practice is to disconnect power, make or change the test connections, and then
Make certain that the supply's ground terminal (┴) is securely connected to an earth ground before applying power. Failure to do so will cause a potential shock hazard that could result in personal injury.