Agilent Technologies AGILENT MODELS 6023A and 6028A service manual Calibration Procedure, Mon

Page 12

Table 2-2. Calibration Procedure

TEST

TESTED

TEST POINTS

 

TEST SEQUENCE AND ADJUSTMENTS

 

EXPECTED

 

VARIABLE

 

 

 

 

RESULTS

Meter F/S

Meter Ref.

A2J3 pin 6 ( + )

a.

Connect DVM across test points and turn on

0.5V ± 50∝V

Adjust.

Voltage

M ( - )

 

ac power.

 

 

 

 

 

 

 

 

 

b.

Adjust A2R24 to obtain the voltage range

 

 

 

 

 

 

 

 

 

 

specified in the results.

 

 

 

 

 

 

Resistance

Prog. Voltage

VP ( + )

a.

Connect a 2KΩ 0.01%, ¼W programming

2.5V ±4mV

Programming

 

P ( - )

 

resistor and DVM between test points.

 

 

 

 

 

 

F/S

 

 

b.

Set MODE switch as in Figure 2-1 and turn on

 

 

 

 

 

 

Adjust.

 

 

 

ac power.

 

 

 

 

 

 

 

 

 

c.

Adjust A2R23 to obtain the voltage range

 

 

 

 

 

 

 

 

 

 

specified in the results.

 

 

 

 

 

 

V-MON

V-MON

VM ( + )

a.

Set voltage and current controls to minimum

0 ± 20∝V

Zero

 

M ( - )

 

settings.

 

 

 

 

 

 

Adjust.

 

 

b.

Disable power supply as in Initial Setup step i.

 

 

 

 

 

 

 

 

 

c.

Short circuit output terminals and connect the

 

 

 

 

 

 

 

 

 

 

DVM between test points. Turn on power

 

 

 

 

 

 

 

 

 

 

supply.

 

 

 

 

 

 

 

 

 

d.

Adjust V-MON Zero trim pot A2R22 to

 

 

 

 

 

 

 

 

 

 

voltage range specified in the results.

 

 

 

 

 

 

Common

Residual

VM ( + )

a.

Set voltage and current controls to minimum.

IR ± 20∝V

Mode

Output

M ( - )

b.

Disable power supply as Initial Setup step i.

 

 

 

 

 

 

Adjust.

Voltage

 

c.

Turn on ac power and record the initial

 

 

 

 

 

 

 

VM( + )

 

 

voltage (IR) with DVM across test points.

 

 

 

 

 

 

 

 

 

d.

Remove the - S( + ) and – OUT( - ) and

 

 

 

 

 

 

 

 

 

 

connect a 1Vdc power supply between - S( + )

 

 

 

 

 

 

 

 

 

 

and – OUT( - ). See Figure 2-1.

 

 

 

 

 

 

 

 

 

e.

Adjust A2R21 to the voltage range specified.

 

 

 

 

 

 

 

 

 

f.

Remove the 1V supply and replace jumpers.

 

 

 

 

 

 

I-MON

I-MON

IM ( + )

a.

Set voltage and current controls to minimum.

0 ± 100∝V

Zero

 

M (-)

b.

Turn on ac power.

(6023A)

 

Adjust.

 

 

c.

Connect DVM across test points and adjust

0

±

 

 

 

 

 

 

 

I-MON Zero trim pot A2R8 as shown in

 

25 V

 

 

 

 

(6028A)

 

 

 

 

 

results.

 

 

 

 

 

 

I-MON

I-MON

IM ( + )

a.

Perform I-MON Zero Adjust before

IR*

 

 

 

 

F / S

 

M ( - )

 

proceeding .

 

 

 

 

 

 

Adjust.

 

 

b.

Connect a 0.001Ω 0.05% (6023A), 0.100Ω

 

 

 

 

 

 

 

 

 

 

0.05% (6028A) current monitoring resistor

 

 

 

 

 

 

 

 

 

 

Rm across the output terminals.

 

 

 

 

 

 

 

 

 

c.

Turn on ac power and using the “Display

 

 

 

 

 

 

 

 

 

 

Setting”, set current control to 30A (6023A),

 

 

 

 

 

 

 

 

 

 

10A (6028A), and voltage control to 5V.

 

 

 

 

 

 

 

 

 

d.

Connect DVM across test points and take an

 

 

 

 

 

 

 

 

 

 

initial reading (IR).

 

 

 

 

 

 

 

 

Rm ( + )

e.

Connect DVM across Rm monitoring

0.006 IR*

 

 

Rm ( - )

 

terminals and adjust A2R9 as shown in the

+40∝V (6023A),

 

 

 

 

results.

0.200

±

1

 

 

 

 

 

 

V

 

 

 

 

 

(6028A)

 

*IR = Initial Reading

 

 

 

 

 

 

 

 

 

12

Image 12
Contents Autoranging Certification Safety Summary Safety Symbol Definitions Printing HistoryTable of Contents Replaceable Parts Troubleshooting Circuit DiagramsSafety Considerations ScopeManual Revisions Item DescriptionCalibration Procedure IntroductionTest Equipment Required Operation Verification TestsType Required Characteristics USE Recommended Model Test Equipment RequiredInitial Setup 1KVACalibration Procedure MONVout IoutPerformance Tests Measurement TechniquesCR7 Current-Monitoring Resistor Setup Constant Voltage CV TestsBasic Test Setup RMS Measurement Test Setup, CV Pard Test Peak-To-Peak Measurement Test Setup, CV Pard Test 6023A 6028A KHz Noise, CV Peak-to-Peak PardLoad Transient Recovery Waveform Constant Current CC Tests CC Pard Test Setup Troubleshooting Initial Troubleshooting ProceduresControl Board Test Connector, A2J7 PIN no Signal NameVdc WAVEFORM/CONDITIONS SourceRepair and Replacement Electrostatic ProtectionA2 Control Board Removal A4 Power Mesh Board Removal Overall Troubleshooting ProcedureA3 Front-Panel Board Removal A1 Main Board RemovalUsing the Tables SetupNode MeasurementMain Troubleshooting Setup Main Troubleshooting SetupModified Mains Cord Set For Troubleshooting Troubleshooting No-Output Failures Front-Panel TroubleshootingPWM-ON PWM-OFF Defective Check Functional Circuits A2J7-26 A2J7-25Board Troubleshooting Bias Supplies Symptoms Defective Circuit Check ComponentsSymptoms Defective Check Functional Circuits Board Node + Measurement SourceNode + Measurement SourceNode + N0DE Power Section Blocks Troubleshooting AC-Turn-on CircuitsSetup Measurement Source Setup MeasurementTroubleshooting PWM & Clock Relay EnablePower Limit Troubleshooting DC-To-DC Converter + OUTPWM-ON PWM-OFFTroubleshooting CV Circuit Troubleshooting CC CircuitTroubleshooting Down Programmer Troubleshooting OVP CircuitA4U4OUT Waveforms SET Voltage Vdc Node + Page Autoranging Power Block Diagram OverviewBlock Diagram Simplified Schematic Simplified Schematic DC-to-DC Converter Down ProgrammerPfet Control Signals Timing Diagram Constant-Voltage CV CircuitPower-Limit Comparator Constant-Current CC CircuitOvervoltage Protection OVP Circuit Control-Voltage ComparatorInitial-Ramp Circuit Pulse-Width Modulator PWMBias Voltage Detector AC-Surge Dropout DetectorSecond-Delay Circuit Display CircuitsPage Replaceable Parts Reference DesignatorsOrdering Information Description AbbreviationsElectrical Parts CR1CR2 CR3XA2P1 XA2P2Mechanical Parts Control Board Assembly CR3 Page Page Page VR3 VR4VR5 CR5 DS1DS5 IC A/D Cmos 3-1/2 DGT VR1VR2 NPN SI PfetTS1 Chassis ElectricalA1TB1 A2 Board Mechanical PartsA3 Board Mechanical Parts A4 Board Mechanical PartsOption 220 220V Operation Option 240 240V OperationComponent Location and Circuit Diagrams Schematic Diagram Notes Pin locations for other semi-conductors are shown below Top View, Top Covers Removed Main Board A1 Component Location Control Board A2 Component Location Front-Panel Board A3 Component Location Power Mesh Board A4 Component Location Page Page Specifications General InformationOption 002 Hardware Table A-1. Specifications, Option Remote Programming Input Compliance Voltage ± Current Programming EnableOutput Impedance 10.2 k ohm ± 5% Temperature Coefficient Status IndicatorsRemote Trip and Remote Reset Timing Power-on PresetMaximum Output Voltage logic high + On State logic lowTable A-1. Specifications, Option Pulse Timing Bias Supplies DC Output Ratings 25C ±Short Circuit Output Current Pard TypicalInstallation Connector Assembly ProcedureFigure A-1. Mating Connector Assembly OperationLocal/Remote Programming Resistance Voltage or Current Remote Resistance Programming Figure A-4. Calculating Value of Series Dropping ResistorFigure A-5. Remote Resistance Programming Remote Monitoring Status Indicators Remote ControlRemote Reset OvervoltagePower-On Preset Figure A-8. Remote ControlAC Dropout Buffer Circuit Multiple Supply System ShutdownMaintenance Bias SuppliesTroubleshooting Resistance and Voltage Programming Troubleshooting Current ProgrammingFigure A-11. Troubleshooting Current Programming of CV Mode Figure A-13. Troubleshooting Status Indicators Page Table A-3. Replacement Parts REF. Desig Model no Description100 VR9 101Indicator and Qualifier Symbols OLD Symbol NEW SymbolLogic Symbols and Definitions 102 DefinitionsSchematic Notes Schematic Diagram Notes 103Schematic Notes Figure A-15. Option 002 Board, Component Location 104Figure A-16. Option 002 Board, Schematic Diagram 105 Page Model 6023A Model 6028A107 From 108