Agilent Technologies 85032B/E 50 Environmental Requirements, Temperature-What to Watch Out For

Page 12

Specifications

Environmental Requirements

Environmental Requirements

Table 2-1 Environmental Requirements

Parameter

Limits

 

 

Operating temperaturea

+15 °C to +35 °C (+59 °F to +95 °F)

Error-corrected temperature rangeb

±1 °C of measurement calibration temperature

Storage temperature

−40 °C to +75 °C (−40 °F to +167 °F)

 

 

Altitude

 

Operation

< 4,500 meters (≈15,000 feet)

Storage

< 15,000 meters (≈50,000 feet)

 

 

Relative humidity

Always non-condensing

Operation

0 to 80% (26 °C maximum dry bulb)

Storage

0 to 95%

 

 

a.The temperature range over which the calibration standards maintain conformance to their specifications.

b.The allowable network analyzer ambient temperature drift during measurement calibration and during measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.

Temperature—What to Watch Out For

Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range specified in Table 2-1.

IMPORTANT Avoid unnecessary handling of the devices during calibration because your fingers are a heat source.

2-2

85032B/E

Image 12
Contents User’s and Service Guide Agilent Part Number Print Date JulyDocumentation Warranty Contents Troubleshooting General Information Kit Contents Calibration Kit OverviewInstallation of the Calibration Definitions Calibration DefinitionsAdapters OptionsIncoming Inspection Recording the Device Serial Numbers Serial Number Record for 85032BSerial Number Record for 85032E Clarifying the Sex of a Connector Preventive MaintenanceSpecifications Temperature-What to Watch Out For Environmental RequirementsEnvironmental Requirements Parameter LimitsMechanical Characteristics Pin DepthPin Depth Limit Electrical Specifications Supplemental Electrical CharacteristicsElectrical Specifications for 50Ω Type-N Devices Supplemental Electrical CharacteristicsElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeWhat Causes Connector Wear? Visual InspectionLook for Obvious Defects and Damage First Inspect the Mating Plane SurfacesCleaning Connectors Inspect Female ConnectorsUse Compressed Air or Nitrogen Clean the Mating Plane Surfaces ReinspectGaging Connectors Connector Gage AccuracyWhen to Gage Connectors Reading the Connector GageGaging Procedures Gaging Male Type-N ConnectorsGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors How to Make a Connection Final Connection Using a Torque WrenchConnections Preliminary ConnectionWrench Positions Connecting the Two-Piece Female Open How to Separate a Connection Handling and StorageHandling and Storage 85032B/E Performance Verification Introduction How Agilent Verifies the Devices in This KitWhere to Send a Kit for Recertification RecertificationLimited Recertification Option How Often to RecertifyRecertification 85032B/E Troubleshooting Troubleshooting Process Troubleshooting FlowchartReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Protective End Caps for Connectors Replaceable Parts for the 85032B Calibration KitAdapters included with Option 100b Calibration Kit Storage CaseReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit Connector Cleaning Supplies Replaceable Parts-Items Not Included in the Calibration KitWrenches ESD Protective DevicesIntroduction 85032B/E Standard Definitions Standard Class Assignments Table A-1 Standard Class AssignmentsClass Standard Class Label Blank Form Table A-2 Standard Class Assignments Blank FormCalibration Kit Label Class Standard ClassSetting the System Impedance Nominal Standard DefinitionsTable A-3 Standard Definitions System Z a = 50 Ω Calibration Kit LabelType Min MaxStandardb Offset Freq Number Type 10C0× 10C1× PsDelay Table A-4 Standard Definitions Blank FormSystem Z 0 a = 10C2× 10C3×Index Index-2 85032B/E Index-3