General Information

Calibration Kit Overview

Adapters

Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repeatable connections. The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized.

The adapters are designed so that their nominal electrical lengths are the same, which allows them to be used in calibration procedures for non-insertable devices.

Calibration Definitions

The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.

The calibration definitions can be:

resident within the analyzer

entered from the front panel

Installation of the Calibration Definitions

The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.

If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.

Options

The following options are available for the Agilent 85032B/E.

Option 100 (85032B only) Option 100 adds the four type-N to 7-mm adapters to the calibration kit.

Option 003 (85032B only) This option provides a limited calibration for the devices in the calibration kit to 3 GHz instead of 6 GHz. This calibration option can be requested only from an Agilent service center. It cannot be ordered from the factory.

Option UK6 This option adds a certificate of calibration and the corresponding calibration data for the devices in the calibration kit.

Equipment Required but Not Supplied

Gages, torque and open-end wrenches, and various connector cleaning supplies are not included in the calibration kit but are required to ensure successful operation of the calibration kit. Refer to Table 6-3 on page 6-5for ordering information

85032B/E

1-3

Image 7
Contents Agilent Part Number Print Date July User’s and Service GuideDocumentation Warranty Contents Troubleshooting General Information Calibration Kit Overview Kit ContentsOptions Installation of the Calibration DefinitionsCalibration Definitions AdaptersIncoming Inspection Serial Number Record for 85032B Recording the Device Serial NumbersSerial Number Record for 85032E Preventive Maintenance Clarifying the Sex of a ConnectorSpecifications Parameter Limits Temperature-What to Watch Out ForEnvironmental Requirements Environmental RequirementsPin Depth Mechanical CharacteristicsPin Depth Limit Supplemental Electrical Characteristics Electrical SpecificationsSupplemental Electrical Characteristics Electrical Specifications for 50Ω Type-N DevicesElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices Electrostatic Discharge ESD Protection SetupInspect the Mating Plane Surfaces What Causes Connector Wear?Visual Inspection Look for Obvious Defects and Damage FirstInspect Female Connectors Cleaning ConnectorsUse Compressed Air or Nitrogen Reinspect Clean the Mating Plane SurfacesConnector Gage Accuracy Gaging ConnectorsReading the Connector Gage When to Gage ConnectorsGaging Male Type-N Connectors Gaging ProceduresGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Preliminary Connection How to Make a ConnectionFinal Connection Using a Torque Wrench ConnectionsWrench Positions Connecting the Two-Piece Female Open Handling and Storage How to Separate a ConnectionHandling and Storage 85032B/E Performance Verification How Agilent Verifies the Devices in This Kit IntroductionHow Often to Recertify Where to Send a Kit for RecertificationRecertification Limited Recertification OptionRecertification 85032B/E Troubleshooting Troubleshooting Flowchart Troubleshooting ProcessContacting Agilent Returning a Kit or Device to AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Calibration Kit Storage Case Protective End Caps for ConnectorsReplaceable Parts for the 85032B Calibration Kit Adapters included with Option 100bReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit ESD Protective Devices Connector Cleaning SuppliesReplaceable Parts-Items Not Included in the Calibration Kit WrenchesIntroduction 85032B/E Standard Definitions Table A-1 Standard Class Assignments Standard Class AssignmentsClass Standard Class Label Standard Class Blank FormTable A-2 Standard Class Assignments Blank Form Calibration Kit Label ClassNominal Standard Definitions Setting the System ImpedanceMin Max Table A-3 Standard DefinitionsSystem Z a = 50 Ω Calibration Kit Label Type10C2× 10C3× Standardb Offset Freq Number Type 10C0× 10C1× PsDelayTable A-4 Standard Definitions Blank Form System Z 0 a =Index Index-2 85032B/E Index-3