Agilent Technologies 85032B/E 50 manual Introduction, How Agilent Verifies the Devices in This Kit

Page 34

Performance Verification

Introduction

Introduction

The performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.

How Agilent Verifies the Devices in This Kit

Agilent verifies the specifications of these devices as follows:

1.The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to NIST (National Institute of Standards and Technology). The airline and short characteristics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements.

2.Each calibration device is electrically tested on this system. For the initial (before sale) testing of the calibration devices, Agilent includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified kits.

These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in this kit is traceable to NIST through Agilent Technologies.

4-2

85032B/E

Image 34
Contents User’s and Service Guide Agilent Part Number Print Date JulyDocumentation Warranty Contents Troubleshooting General Information Kit Contents Calibration Kit OverviewAdapters Installation of the Calibration DefinitionsCalibration Definitions OptionsIncoming Inspection Serial Number Record for 85032B Recording the Device Serial NumbersSerial Number Record for 85032E Clarifying the Sex of a Connector Preventive MaintenanceSpecifications Environmental Requirements Temperature-What to Watch Out ForEnvironmental Requirements Parameter LimitsMechanical Characteristics Pin DepthPin Depth Limit Electrical Specifications for 50Ω Type-N Devices Electrical SpecificationsSupplemental Electrical Characteristics Supplemental Electrical CharacteristicsElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeLook for Obvious Defects and Damage First What Causes Connector Wear?Visual Inspection Inspect the Mating Plane SurfacesInspect Female Connectors Cleaning ConnectorsUse Compressed Air or Nitrogen Clean the Mating Plane Surfaces ReinspectGaging Connectors Connector Gage AccuracyWhen to Gage Connectors Reading the Connector GageGaging Procedures Gaging Male Type-N ConnectorsGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Connections How to Make a ConnectionFinal Connection Using a Torque Wrench Preliminary ConnectionWrench Positions Connecting the Two-Piece Female Open How to Separate a Connection Handling and StorageHandling and Storage 85032B/E Performance Verification Introduction How Agilent Verifies the Devices in This KitLimited Recertification Option Where to Send a Kit for RecertificationRecertification How Often to RecertifyRecertification 85032B/E Troubleshooting Troubleshooting Process Troubleshooting FlowchartReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Adapters included with Option 100b Protective End Caps for ConnectorsReplaceable Parts for the 85032B Calibration Kit Calibration Kit Storage CaseReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit Wrenches Connector Cleaning SuppliesReplaceable Parts-Items Not Included in the Calibration Kit ESD Protective DevicesIntroduction 85032B/E Standard Definitions Table A-1 Standard Class Assignments Standard Class AssignmentsClass Standard Class Label Calibration Kit Label Class Blank FormTable A-2 Standard Class Assignments Blank Form Standard ClassSetting the System Impedance Nominal Standard DefinitionsType Table A-3 Standard DefinitionsSystem Z a = 50 Ω Calibration Kit Label Min MaxSystem Z 0 a = Standardb Offset Freq Number Type 10C0× 10C1× PsDelayTable A-4 Standard Definitions Blank Form 10C2× 10C3×Index Index-2 85032B/E Index-3