Agilent Technologies 85032B/E 50 manual Calibration Kit Overview, Kit Contents

Page 6

General Information

Calibration Kit Overview

Calibration Kit Overview

The Agilent 85032B and 85032E type-N calibration kits are used to calibrate Agilent network analyzers up to 6 GHz for measurements of components with 50Ω type-N connectors.

Kit Contents

The 85032B calibration kit contains the following:

one male and one female open termination

one male and one female short termination

one male and one female 50Ω load

two type-N-male to 7-mm adapters (included with Option 100)

two type-N-female to 7-mm adapters (included with Option 100)

Refer to Table 6-1and Figure 6-1for a complete list of kit contents and their associated part numbers.

The 85032E calibration kit contains the following:

one male combination open/short termination

one male 50Ω load

Refer to Table 6-2and Figure 6-2for a complete list of kit contents and their associated part numbers.

Broadband Loads

The broadband loads are instrument-grade, 50Ω terminations that have been optimized for performance up to 6 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.

Opens and Shorts

The opens and shorts are built from parts that are machined to the current state-of the-art precision machining.

The short’s inner conductors have a one-piece construction, common with the shorting plane. This construction provides for extremely repeatable connections.

The female open has a separate-piece inner conductor that is made from a low-dielectric-constant plastic to minimize compensation values.

Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The lengths of these offsets are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.

1-2

85032B/E

Image 6
Contents User’s and Service Guide Agilent Part Number Print Date JulyDocumentation Warranty Contents Troubleshooting General Information Kit Contents Calibration Kit OverviewAdapters Installation of the Calibration DefinitionsCalibration Definitions OptionsIncoming Inspection Recording the Device Serial Numbers Serial Number Record for 85032BSerial Number Record for 85032E Clarifying the Sex of a Connector Preventive MaintenanceSpecifications Environmental Requirements Temperature-What to Watch Out ForEnvironmental Requirements Parameter LimitsMechanical Characteristics Pin DepthPin Depth Limit Electrical Specifications for 50Ω Type-N Devices Electrical SpecificationsSupplemental Electrical Characteristics Supplemental Electrical CharacteristicsElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeLook for Obvious Defects and Damage First What Causes Connector Wear?Visual Inspection Inspect the Mating Plane SurfacesCleaning Connectors Inspect Female ConnectorsUse Compressed Air or Nitrogen Clean the Mating Plane Surfaces ReinspectGaging Connectors Connector Gage AccuracyWhen to Gage Connectors Reading the Connector GageGaging Procedures Gaging Male Type-N ConnectorsGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Connections How to Make a ConnectionFinal Connection Using a Torque Wrench Preliminary ConnectionWrench Positions Connecting the Two-Piece Female Open How to Separate a Connection Handling and StorageHandling and Storage 85032B/E Performance Verification Introduction How Agilent Verifies the Devices in This KitLimited Recertification Option Where to Send a Kit for RecertificationRecertification How Often to RecertifyRecertification 85032B/E Troubleshooting Troubleshooting Process Troubleshooting FlowchartReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Adapters included with Option 100b Protective End Caps for ConnectorsReplaceable Parts for the 85032B Calibration Kit Calibration Kit Storage CaseReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit Wrenches Connector Cleaning SuppliesReplaceable Parts-Items Not Included in the Calibration Kit ESD Protective DevicesIntroduction 85032B/E Standard Definitions Standard Class Assignments Table A-1 Standard Class AssignmentsClass Standard Class Label Calibration Kit Label Class Blank FormTable A-2 Standard Class Assignments Blank Form Standard ClassSetting the System Impedance Nominal Standard DefinitionsType Table A-3 Standard DefinitionsSystem Z a = 50 Ω Calibration Kit Label Min MaxSystem Z 0 a = Standardb Offset Freq Number Type 10C0× 10C1× PsDelayTable A-4 Standard Definitions Blank Form 10C2× 10C3×Index Index-2 85032B/E Index-3