Agilent Technologies 85032B/E 50 manual Pin Depth Limit

Page 14

Specifications

Mechanical Characteristics

NOTE

The gages for measuring type-N connectors compensate for the designed

 

offset of 5.26 mm (0.207 inch), therefore, protrusion and recession readings

 

are in relation to a zero reference plane (as if the inner and outer conductor

 

planes were intended to be flush). Gage readings can be directly compared

 

with the observed values listed in Table 2-2.

 

 

The pin depth value of each calibration device in this kit is not specified, but is an important mechanical parameter. The electrical performance of the device depends, to some extent, on its pin depth. The electrical specifications for each device in this kit take into account the effect of pin depth on the device’s performance. Table 2-2lists the typical pin depths and measurement uncertainties, and provides observed pin depth limits for the devices in the kit. If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to Figure 2-1for an illustration of pin depth in type-N connectors.

Table 2-2 Pin Depth Limit

Device

Typical Pin Depth

Measurement Uncertaintya

Observed Pin Depth Limitsb

 

 

 

 

Opens

0 to −0.0127 mm

+0.0038 to −0.0038 mm

+0.0038 to −0.0165 mm

 

0 to −0.0005 in

+0.00015 to −0.00015 in

+0.00015 to −0.00065 in

 

 

 

 

Shorts

0 to −0.0127 mm

+0.0038 to −0.0038 mm

+0.0038 to −0.0165 mm

 

0 to −0.0005 in

+0.00015 to −0.00015 in

+0.00015 to −0.00065 in

 

 

 

 

Fixed Loads

0 to −0.0508 mm

+0.0038 to −0.0038 mm

+0.0038 to −0.0546 mm

 

0 to −0.0020 in

+0.00015 to −0.00015 in

+0.00015 to −0.00215 in

 

 

 

 

a.Approximately +2 sigma to −2 sigma of gage uncertainty based on studies done at the factory according to recommended procedures.

b.Observed pin depth limits are the range of observation limits seen on the gage reading due to measurement uncertainty. The depth could still be within specifications.

2-4

85032B/E

Image 14
Contents User’s and Service Guide Agilent Part Number Print Date JulyDocumentation Warranty Contents Troubleshooting General Information Kit Contents Calibration Kit OverviewAdapters Installation of the Calibration DefinitionsCalibration Definitions OptionsIncoming Inspection Serial Number Record for 85032E Recording the Device Serial NumbersSerial Number Record for 85032B Clarifying the Sex of a Connector Preventive MaintenanceSpecifications Environmental Requirements Temperature-What to Watch Out ForEnvironmental Requirements Parameter LimitsMechanical Characteristics Pin DepthPin Depth Limit Electrical Specifications for 50Ω Type-N Devices Electrical SpecificationsSupplemental Electrical Characteristics Supplemental Electrical CharacteristicsElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices ESD Protection Setup Electrostatic DischargeLook for Obvious Defects and Damage First What Causes Connector Wear?Visual Inspection Inspect the Mating Plane SurfacesUse Compressed Air or Nitrogen Cleaning ConnectorsInspect Female Connectors Clean the Mating Plane Surfaces ReinspectGaging Connectors Connector Gage AccuracyWhen to Gage Connectors Reading the Connector GageGaging Procedures Gaging Male Type-N ConnectorsGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Connections How to Make a ConnectionFinal Connection Using a Torque Wrench Preliminary ConnectionWrench Positions Connecting the Two-Piece Female Open How to Separate a Connection Handling and StorageHandling and Storage 85032B/E Performance Verification Introduction How Agilent Verifies the Devices in This KitLimited Recertification Option Where to Send a Kit for RecertificationRecertification How Often to RecertifyRecertification 85032B/E Troubleshooting Troubleshooting Process Troubleshooting FlowchartReturning a Kit or Device to Agilent Contacting AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Adapters included with Option 100b Protective End Caps for ConnectorsReplaceable Parts for the 85032B Calibration Kit Calibration Kit Storage CaseReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit Wrenches Connector Cleaning SuppliesReplaceable Parts-Items Not Included in the Calibration Kit ESD Protective DevicesIntroduction 85032B/E Standard Definitions Class Standard Class Label Standard Class AssignmentsTable A-1 Standard Class Assignments Calibration Kit Label Class Blank FormTable A-2 Standard Class Assignments Blank Form Standard ClassSetting the System Impedance Nominal Standard DefinitionsType Table A-3 Standard DefinitionsSystem Z a = 50 Ω Calibration Kit Label Min MaxSystem Z 0 a = Standardb Offset Freq Number Type 10C0× 10C1× PsDelayTable A-4 Standard Definitions Blank Form 10C2× 10C3×Index Index-2 85032B/E Index-3