Agilent Technologies 85032B/E 50 manual Where to Send a Kit for Recertification

Page 35

Performance Verification

Recertification

Recertification

The following will be provided with a recertified kit:

a new calibration sticker affixed to the case

a certificate of calibration

a calibration report for each device in the kit listing measured values, specifications, and uncertainties

NOTE

A list of NIST traceable numbers may be purchased upon request to be

 

included in the calibration report.

 

 

Agilent Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Agilent Technologies. See Table 5-1 on page 5-3.

Limited Recertification (Option 003)

The 50Ω loads in the calibration kit are specified for use up to 6 GHz. For many applications, the performance above 3 GHz is not utilized.

For a standard recertification, the devices in the calibration kit are tested and calibrated up to 6 GHz. However, a limited recertification can be requested. For this limited recertification, the devices are tested and calibrated up to 3 GHz.

To request a limited recertification, make sure the following is clearly written on the order: Limited Calibration DC–3 GHz (Option 003).

All loads that receive a limited calibration are supplied with a limited calibration label applied to the device.

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.

NOTE

The recertification interval should begin on the date the kit is first used after

 

the recertification date.

 

 

 

 

 

 

 

 

 

 

 

 

Where to Send a Kit for Recertification

Contact Agilent Technologies for information on where to send your kit for recertification. See Table 5-1 on page 5-3.

When you return the kit, complete and attach a service tag. Refer to “Returning a Kit or Device to Agilent” on page 5-3for details.

85032B/E

4-3

Image 35
Contents Agilent Part Number Print Date July User’s and Service GuideDocumentation Warranty Contents Troubleshooting General Information Calibration Kit Overview Kit ContentsOptions Installation of the Calibration DefinitionsCalibration Definitions AdaptersIncoming Inspection Serial Number Record for 85032E Recording the Device Serial NumbersSerial Number Record for 85032B Preventive Maintenance Clarifying the Sex of a ConnectorSpecifications Parameter Limits Temperature-What to Watch Out ForEnvironmental Requirements Environmental RequirementsPin Depth Mechanical CharacteristicsPin Depth Limit Supplemental Electrical Characteristics Electrical SpecificationsSupplemental Electrical Characteristics Electrical Specifications for 50Ω Type-N DevicesElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices Electrostatic Discharge ESD Protection SetupInspect the Mating Plane Surfaces What Causes Connector Wear?Visual Inspection Look for Obvious Defects and Damage FirstUse Compressed Air or Nitrogen Cleaning ConnectorsInspect Female Connectors Reinspect Clean the Mating Plane SurfacesConnector Gage Accuracy Gaging ConnectorsReading the Connector Gage When to Gage ConnectorsGaging Male Type-N Connectors Gaging ProceduresGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Preliminary Connection How to Make a ConnectionFinal Connection Using a Torque Wrench ConnectionsWrench Positions Connecting the Two-Piece Female Open Handling and Storage How to Separate a ConnectionHandling and Storage 85032B/E Performance Verification How Agilent Verifies the Devices in This Kit IntroductionHow Often to Recertify Where to Send a Kit for RecertificationRecertification Limited Recertification OptionRecertification 85032B/E Troubleshooting Troubleshooting Flowchart Troubleshooting ProcessContacting Agilent Returning a Kit or Device to AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Calibration Kit Storage Case Protective End Caps for ConnectorsReplaceable Parts for the 85032B Calibration Kit Adapters included with Option 100bReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit ESD Protective Devices Connector Cleaning SuppliesReplaceable Parts-Items Not Included in the Calibration Kit WrenchesIntroduction 85032B/E Standard Definitions Class Standard Class Label Standard Class AssignmentsTable A-1 Standard Class Assignments Standard Class Blank FormTable A-2 Standard Class Assignments Blank Form Calibration Kit Label ClassNominal Standard Definitions Setting the System ImpedanceMin Max Table A-3 Standard DefinitionsSystem Z a = 50 Ω Calibration Kit Label Type10C2× 10C3× Standardb Offset Freq Number Type 10C0× 10C1× PsDelayTable A-4 Standard Definitions Blank Form System Z 0 a =Index Index-2 85032B/E Index-3