Agilent Technologies 85032B/E 50 manual Electrical Specifications, Certification

Page 15

Specifications

Electrical Specifications

Electrical Specifications

The electrical specifications in Table 2-3apply to the devices in your calibration kit when connected with an Agilent precision interface.

Table 2-3 Electrical Specifications for 50Ω Type-N Devices

Device

Specification

Frequency (GHz)

 

 

 

Loads

Return loss ≥ 49 dB (ρ ≤ 0.00355)

DC to ≤ 2

 

Return loss ≥ 46 dB (ρ ≤ 0.00501)

> 2 to ≤ 3

 

Return loss ≥ 40 dB (ρ ≤ 0.01000)

> 3 to ≤ 6

 

 

 

Female opena

±0.501 ° ±0.484 °/GHz deviation from nominal

DC to ≤ 6

a

±0.490 ° ±0.385 °/GHz deviation from nominal

DC to ≤ 6

Female short

 

 

Male opena

±0.501 ° ±0.234 °/GHz deviation from nominal

DC to ≤ 6

Male shorta

±0.441 ° ±0.444 °/GHz deviation from nominal

DC to ≤ 6

Adapters (type-N to 7-mm)

Return loss ≥ 30 dΒ (ρ ≤ 0.03162)

DC to ≤ 6

 

 

 

a.The specifications for the opens and shorts are given as allowed deviation from the nominal model as defined in the standard definitions. See Table A-3 on page A-5.

Certification

Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Agilent Verifies the Devices in This Kit” on page 4-2for more information.

Supplemental Electrical Characteristics

Supplemental electrical characteristics are values which are typically met by a majority of the calibration kit devices tested. These supplemental characteristics are intended to provide information in calibration kit applications by giving typical, but non-warranted, performance parameters. Table 2-4lists the typical electrical characteristics of the 50Ω loads and adapters in the 85032B/E calibration kit.

Table 2-4 Supplemental Electrical Characteristics

Device

Specification

Frequency (GHz)

 

 

 

Loads

Return loss ≥ 23 dΒ (ρ ≤ 0.07079)

> 6 to ≤ 18

 

 

 

Adapters (type-N to 7-mm)

Return loss ≥ 34 dΒ (ρ ≤ 0.01995)

DC to ≤ 8

 

Return loss ≥ 28 dΒ (ρ ≤ 0.03981)

> 8 to ≤ 18

 

 

 

85032B/E

2-5

Image 15
Contents Agilent Part Number Print Date July User’s and Service GuideDocumentation Warranty Contents Troubleshooting General Information Calibration Kit Overview Kit ContentsOptions Installation of the Calibration DefinitionsCalibration Definitions AdaptersIncoming Inspection Recording the Device Serial Numbers Serial Number Record for 85032BSerial Number Record for 85032E Preventive Maintenance Clarifying the Sex of a ConnectorSpecifications Parameter Limits Temperature-What to Watch Out ForEnvironmental Requirements Environmental RequirementsPin Depth Mechanical CharacteristicsPin Depth Limit Supplemental Electrical Characteristics Electrical SpecificationsSupplemental Electrical Characteristics Electrical Specifications for 50Ω Type-N DevicesElectrical Specifications 85032B/E Use, Maintenance, and Care of the Devices Electrostatic Discharge ESD Protection SetupInspect the Mating Plane Surfaces What Causes Connector Wear?Visual Inspection Look for Obvious Defects and Damage FirstCleaning Connectors Inspect Female ConnectorsUse Compressed Air or Nitrogen Reinspect Clean the Mating Plane SurfacesConnector Gage Accuracy Gaging ConnectorsReading the Connector Gage When to Gage ConnectorsGaging Male Type-N Connectors Gaging ProceduresGaging Male Type-N Connectors Gaging Female Type-N Connectors Gaging Female Type-N Connectors Preliminary Connection How to Make a ConnectionFinal Connection Using a Torque Wrench ConnectionsWrench Positions Connecting the Two-Piece Female Open Handling and Storage How to Separate a ConnectionHandling and Storage 85032B/E Performance Verification How Agilent Verifies the Devices in This Kit IntroductionHow Often to Recertify Where to Send a Kit for RecertificationRecertification Limited Recertification OptionRecertification 85032B/E Troubleshooting Troubleshooting Flowchart Troubleshooting ProcessContacting Agilent Returning a Kit or Device to AgilentReturning a Kit or Device to Agilent 85032B/E Replaceable Parts Calibration Kit Storage Case Protective End Caps for ConnectorsReplaceable Parts for the 85032B Calibration Kit Adapters included with Option 100bReplaceable Parts for the 85032B Calibration Kit Replaceable Parts for the 85032E Calibration Kit ESD Protective Devices Connector Cleaning SuppliesReplaceable Parts-Items Not Included in the Calibration Kit WrenchesIntroduction 85032B/E Standard Definitions Standard Class Assignments Table A-1 Standard Class AssignmentsClass Standard Class Label Standard Class Blank FormTable A-2 Standard Class Assignments Blank Form Calibration Kit Label ClassNominal Standard Definitions Setting the System ImpedanceMin Max Table A-3 Standard DefinitionsSystem Z a = 50 Ω Calibration Kit Label Type10C2× 10C3× Standardb Offset Freq Number Type 10C0× 10C1× PsDelayTable A-4 Standard Definitions Blank Form System Z 0 a =Index Index-2 85032B/E Index-3