a/d/s/ iSeries Analog Output Retransmission, Enter Analog Output Menu, CURRENT/VOLTAGE Submenu

Page 43

3.2.10 Analog Output (Retransmission)

This unit is equipped with three physical outputs that can only be configured as follows:

a)Alarm 1, Alarm 2 & Analog Out

b)Output 1, Alarm 2 & Analog Out

Analog Output is available only if the Analog Output option board is factory installed.

It is required that you put the controller in the Standby Mode for any configuration change other than Set Points & Alarms.

Figure 3.9 Flow Chart for Analog Output (Retransmission)

ENTER ANALOG OUTPUT MENU:

Press a 1) Press a, if necessary, until CNFG prompt appears. Press d 2) Display advances to INPT Input Menu.

Press a 3) Press a, if necessary, until Display advances to ANLG Analog Output Menu.

Press d 4) Display advances to Analog Output CURR or VoLT Current/Voltage Submenu and flashes the previous selection.

CURRENT/VOLTAGE SUBMENU:

Press d 5) Display flashes CURR Current or VoLT Voltage.

Press b 6) Scroll through the available selection: Current or Voltage. (Example VoLT).

Press d 7) Display shows STRD stored message momentarily and then advances to RD1 Submenu only if it was changed, otherwise press a to advance to RD1 Reading 1 Submenu.

39

Image 43
Contents Newport Electronics,Inc Newport Electronics,Inc Table of Contents List of Tables Flow Chart for Alarm 1 Menu Flow Chart for Alarm 2 MenuInput Properties Factory Preset Values List of FiguresTIP Description Part IntroductionEMC Considerations Safety ConsiderationsBefore You Begin Optional Remote Programmer Front Panel Part SetupOutput PowerFront Panel Connections InputMain Power Connections Fuse Requirement see specifications Electrical Installation Power ConnectionsThermocouple Input Thermocouple Wiring Hookup TC Wire Color ChartOhm Wiring Hookup 3 Two/Three/Four-Wire RTDProcess Voltage Process CurrentIsolated Analog Output Wiring Hookup Wiring Outputs10 Snubber Circuits Wiring Hookup 13 Excitation Output Button Function in Configuration Mode Operation Configuration Mode IntroductionPart Flow Chart for ID and Set Points Menu ConfigurationEntering Your NON-DEFAULT SETPOINT/ID Security ID Number Entering Your NON-DEFAULT Full Security ID NumberID Number Setpoint Set PointsEnter Configuration Menu Configuration MenuInput Type Menu Thermocouple Types Input Type ThermocoupleEnter Input Type Menu Thermocouple SubmenuRTD Value Submenu Input Type RTDRTD Submenu RTD TypesProcess Types Reading Configuration MenuInput Type Process Process SubmenuFilter Constant Submenu Enter Reading Configuration MenuTemperature Unit Submenu Decimal Point SubmenuINPUT/READING Scale and Offset Submenu Reading Configuration If Process was selectedRange Conversion Number Conversion TableEnter Alarm 1 Menu AlarmAlarm 1 ABSOLUTE/DEVIATION Submenu Alarm 1 ENABLE/DISABLE SubmenuAlarm 1 LATCH/UNLATCH Submenu Active Submenu Contact Closure SubmenuAlarm 1 LOW Value Submenu Alarm ENABLE/DISABLE AT Power onAlarm 1 HI Value Submenu Enter Alarm 2 Menu Loop Break Time Value Submenu Loop Break ENABLE/DISABLE SubmenuLoop Break Time/Field Calibration Enter Loop Break Time MenuSetpoint Deviation ENABLE/DISABLE Submenu Reading Adjust SubmenuThermocouple Field Calibration Submenu Output Flow Chart for OutputSelf Submenu Enter Output 1 MenuMINIMUM/PERCENT LOW Submenu Control Type Output Relay, SSR, or Pulse MAXIMUM/PERCENT High SubmenuAction Type Submenu Anti Integral Submenu Auto PID SubmenuProportional Band Submenu Start Auto Tune PIDRate Setup Submenu Reset Setup SubmenuCycle Time Submenu Deadband Submenu Damping Factor SubmenuEnter Analog Output Menu Analog Output RetransmissionCURRENT/VOLTAGE Submenu OUT ReadingEnter Ramp and Soak Menu Ramp ENABLE/DISABLE SubmenuSoak ENABLE/DISABLE Submenu Ramp & SoakSoak Value Submenu Ramp Value SubmenuID Code Entering or Changing Your NON-DEFAULT ID CodeEnter ID Code Menu Full Security Level Submenu Entering or Changing Your Default ID CodeSETPOINT/ID Security Level Submenu Communication Option 12 Flow Chart for Communication OptionParity Submenu Enter Communication Option MenuCommunication Parameters Submenu Baud SubmenuLine Feed Submenu Stop BIT SubmenuBUS Format Submenu Modbus Protocol SubmenuCommunication Interface Standard Submenu Data Flow Mode SubmenuData Separation Character Submenu Peak Value Submenu Alarm Status SubmenuData Format Submenu Main Reading SubmenuTransmit Time Interval Value Submenu Address Setup SubmenuAddress Value Submenu Transmit Time Interval SubmenuNormal Color Display Submenu Alarm 1 Display Color SubmenuDisplay Color Selection Enter Display Color Selection MenuGreen RED Amber Alarm 2 Display Color SubmenuExample Display Colors change sequencesAmber RED Green Part Specifications Communications ControlControl Output SSRInsulation ExcitationGeneral Input Type Range Accuracy Input PropertiesFactory preset value Part 5 Factory Preset ValuesDisplay Color Colr Measurement Category Part CE Approvals InformationPage Page Page Warranty/Disclaimer M4165/N/0206