a/d/s/ iSeries manual Ramp Value Submenu, Soak Value Submenu

Page 46

Ramp & Soak provides users with the flexibility to slowly bring the Process Variable (PV) to the desired setpoint. Ramp & Soak values are specified in HH.MM format. The Ramp value indicates the time specified to bring the process variable to Setpoint 1 (SP1). Once the set point is reached, the PID takes over and the Process Variable will be controlled at the desired set point indefinitely. If Soak is enabled, PID will control the Process Variable at the specified Setpoint for the duration of Soak time and then will turn off Output 1. To start a new Ramp/Soak cycle, reset the instrument by pressing a and then c button.

An active Ramp/Soak will change SP1 one degree above the PV and will cause the most significant digit to blink. The SP1 will be incremented by one degree until it reaches the original SP1. The minimum Ramp time must be at least twice the time that it will take the PV to reach the Setpoint Value (SV) with OUT 1 fully ON.

RAMP VALUE SUBMENU:

Press d 10) Display flashes 1st digit of previous stored “Ramp Value”. Press b & c 11) Press b and c buttons to enter a new “Ramp Value”. Press d 12) Display shows STRD stored message momentarily and then

advances to “Soak Value” Submenu.

SOAK VALUE SUBMENU:

Press d 13) Display flashes 1st digit of previous stored “Soak Value”. Press b & c 14) Press b and c buttons to enter a new “Soak Value”. Press d 15) Display shows STRD stored message and advances to the

ID ID Code Menu.

The Ramp and Soak time is 00:00 to 99:59 i.e. HH.MM. (from zero to 99 hours and 59 minutes) During Ramp & Soak do not perform any operations or settings before first stopping it. Any alarms or other output are disabled during this time. To stop Ramp & Soak first put instrument into Standby Mode, then go to Ramp & Soak Menu and disable it.

42

Image 46
Contents Newport Electronics,Inc Newport Electronics,Inc Table of Contents List of Figures Flow Chart for Alarm 1 Menu Flow Chart for Alarm 2 MenuInput Properties Factory Preset Values List of TablesTIP Part Introduction DescriptionSafety Considerations EMC ConsiderationsBefore You Begin Part Setup Optional Remote Programmer Front PanelInput PowerFront Panel Connections OutputElectrical Installation Power Connections Main Power Connections Fuse Requirement see specificationsThermocouple Wiring Hookup TC Wire Color Chart Thermocouple Input3 Two/Three/Four-Wire RTD Ohm Wiring HookupProcess Current Process VoltageWiring Outputs Isolated Analog Output Wiring Hookup10 Snubber Circuits Wiring Hookup 13 Excitation Output Button Function in Configuration Mode Operation Configuration Mode IntroductionPart Menu Configuration Flow Chart for ID and Set PointsEntering Your NON-DEFAULT SETPOINT/ID Security ID Number Entering Your NON-DEFAULT Full Security ID NumberID Number Set Points SetpointEnter Configuration Menu Configuration MenuInput Type Menu Thermocouple Submenu Input Type ThermocoupleEnter Input Type Menu Thermocouple TypesRTD Types Input Type RTDRTD Submenu RTD Value SubmenuProcess Submenu Reading Configuration MenuInput Type Process Process TypesDecimal Point Submenu Enter Reading Configuration MenuTemperature Unit Submenu Filter Constant SubmenuReading Configuration If Process was selected INPUT/READING Scale and Offset SubmenuConversion Table Range Conversion NumberAlarm Enter Alarm 1 MenuAlarm 1 ABSOLUTE/DEVIATION Submenu Alarm 1 ENABLE/DISABLE SubmenuAlarm 1 LATCH/UNLATCH Submenu Contact Closure Submenu Active SubmenuAlarm 1 LOW Value Submenu Alarm ENABLE/DISABLE AT Power onAlarm 1 HI Value Submenu Enter Alarm 2 Menu Enter Loop Break Time Menu Loop Break ENABLE/DISABLE SubmenuLoop Break Time/Field Calibration Loop Break Time Value SubmenuSetpoint Deviation ENABLE/DISABLE Submenu Reading Adjust SubmenuThermocouple Field Calibration Submenu Flow Chart for Output OutputSelf Submenu Enter Output 1 MenuMINIMUM/PERCENT LOW Submenu Control Type Output Relay, SSR, or Pulse MAXIMUM/PERCENT High SubmenuAction Type Submenu Auto PID Submenu Anti Integral SubmenuStart Auto Tune PID Proportional Band SubmenuRate Setup Submenu Reset Setup SubmenuCycle Time Submenu Damping Factor Submenu Deadband SubmenuEnter Analog Output Menu Analog Output RetransmissionCURRENT/VOLTAGE Submenu Reading OUTRamp & Soak Ramp ENABLE/DISABLE SubmenuSoak ENABLE/DISABLE Submenu Enter Ramp and Soak MenuRamp Value Submenu Soak Value SubmenuID Code Entering or Changing Your NON-DEFAULT ID CodeEnter ID Code Menu Full Security Level Submenu Entering or Changing Your Default ID CodeSETPOINT/ID Security Level Submenu 12 Flow Chart for Communication Option Communication OptionBaud Submenu Enter Communication Option MenuCommunication Parameters Submenu Parity SubmenuModbus Protocol Submenu Stop BIT SubmenuBUS Format Submenu Line Feed SubmenuCommunication Interface Standard Submenu Data Flow Mode SubmenuData Separation Character Submenu Main Reading Submenu Alarm Status SubmenuData Format Submenu Peak Value SubmenuTransmit Time Interval Submenu Address Setup SubmenuAddress Value Submenu Transmit Time Interval Value SubmenuEnter Display Color Selection Menu Alarm 1 Display Color SubmenuDisplay Color Selection Normal Color Display SubmenuDisplay Colors change sequences Alarm 2 Display Color SubmenuExample Green RED AmberAmber RED Green Part Specifications SSR ControlControl Output CommunicationsInsulation ExcitationGeneral Input Properties Input Type Range AccuracyPart 5 Factory Preset Values Factory preset valueDisplay Color Colr Part CE Approvals Information Measurement CategoryPage Page Page Warranty/Disclaimer M4165/N/0206