a/d/s/ iSeries manual Communication Interface Standard Submenu, Data Flow Mode Submenu

Page 52

Press d 25) Display flashes previous selection for “Echo”. Press b 26) Scroll through the available selections: NO, YES.

Press d 27) Display flashes STRD stored message momentarily and then advances to STND only if it was changed, otherwise press a to advance to STND Communication Standard Submenu.

COMMUNICATION INTERFACE STANDARD SUBMENU:

Determines whether device should be connected to an RS-232C serial port (as is commonly used on IBM PC-compatible computers) or via an RS-485 bus connected through appropriate RS-232/485 converter. When used in RS-485 Mode, the device must be accessed with an appropriate Address Value as selected in the Address Submenu described later.

Press d 28) Display flashes previous selection for “Standard”. Press b 29) Scroll through the available selections: 232C, 485.

Press d 30) Display shows STRD stored message momentarily and then advances to MoDE only, if it was changed, otherwise press a to advance to MoDE Data Flow Mode Submenu.

DATA FLOW MODE SUBMENU:

Determines whether the instrument will wait for commands and data requests from the Serial Bus or whether the instrument will send data automatically and continuously to the Serial Bus. Devices configured for the RS-485 Communications Standard operate properly only under Command Mode.

Press d 31) Display flashes previous selection for “Mode”.

Press b 32) Scroll through the available selections: CMD_ “Command”, CoNT “Continuous”.

Press d 33) Display shows STRD stored message momentarily and then advances to SEPR only, if it was changed, otherwise press a to advance to SEPR Data Separation Submenu.

DATA SEPARATION CHARACTER SUBMENU:

Determines whether data sent from the device in Continuous Data Flow Mode will be separated by spaces or by Carriage Returns.

Press d 34) Display flashes previous selection for “Separation” Submenu. Press b 35) Scroll through the available selections: SPCE “Space” or

_çR_ “Carriage Return”.

Press d 36) Display shows STRD stored message momentarily and then advances to DAT.F only, if it was changed, otherwise press a to advance to DAT.F Data Format Submenu.

48

Image 52
Contents Newport Electronics,Inc Newport Electronics,Inc Table of Contents Flow Chart for Alarm 1 Menu Flow Chart for Alarm 2 Menu Input Properties Factory Preset ValuesList of Figures List of TablesTIP Part Introduction DescriptionSafety Considerations EMC ConsiderationsBefore You Begin Part Setup Optional Remote Programmer Front PanelPower Front Panel ConnectionsInput OutputElectrical Installation Power Connections Main Power Connections Fuse Requirement see specificationsThermocouple Wiring Hookup TC Wire Color Chart Thermocouple Input3 Two/Three/Four-Wire RTD Ohm Wiring HookupProcess Current Process VoltageWiring Outputs Isolated Analog Output Wiring Hookup10 Snubber Circuits Wiring Hookup 13 Excitation Output Button Function in Configuration Mode Operation Configuration Mode IntroductionPart Menu Configuration Flow Chart for ID and Set PointsEntering Your NON-DEFAULT SETPOINT/ID Security ID Number Entering Your NON-DEFAULT Full Security ID NumberID Number Set Points SetpointEnter Configuration Menu Configuration MenuInput Type Menu Input Type Thermocouple Enter Input Type MenuThermocouple Submenu Thermocouple TypesInput Type RTD RTD SubmenuRTD Types RTD Value SubmenuReading Configuration Menu Input Type ProcessProcess Submenu Process TypesEnter Reading Configuration Menu Temperature Unit SubmenuDecimal Point Submenu Filter Constant SubmenuReading Configuration If Process was selected INPUT/READING Scale and Offset SubmenuConversion Table Range Conversion NumberAlarm Enter Alarm 1 MenuAlarm 1 ABSOLUTE/DEVIATION Submenu Alarm 1 ENABLE/DISABLE SubmenuAlarm 1 LATCH/UNLATCH Submenu Contact Closure Submenu Active SubmenuAlarm 1 LOW Value Submenu Alarm ENABLE/DISABLE AT Power onAlarm 1 HI Value Submenu Enter Alarm 2 Menu Loop Break ENABLE/DISABLE Submenu Loop Break Time/Field CalibrationEnter Loop Break Time Menu Loop Break Time Value SubmenuSetpoint Deviation ENABLE/DISABLE Submenu Reading Adjust SubmenuThermocouple Field Calibration Submenu Flow Chart for Output OutputSelf Submenu Enter Output 1 MenuMINIMUM/PERCENT LOW Submenu Control Type Output Relay, SSR, or Pulse MAXIMUM/PERCENT High SubmenuAction Type Submenu Auto PID Submenu Anti Integral SubmenuStart Auto Tune PID Proportional Band SubmenuRate Setup Submenu Reset Setup SubmenuCycle Time Submenu Damping Factor Submenu Deadband SubmenuEnter Analog Output Menu Analog Output RetransmissionCURRENT/VOLTAGE Submenu Reading OUTRamp ENABLE/DISABLE Submenu Soak ENABLE/DISABLE SubmenuRamp & Soak Enter Ramp and Soak MenuRamp Value Submenu Soak Value SubmenuID Code Entering or Changing Your NON-DEFAULT ID CodeEnter ID Code Menu Full Security Level Submenu Entering or Changing Your Default ID CodeSETPOINT/ID Security Level Submenu 12 Flow Chart for Communication Option Communication OptionEnter Communication Option Menu Communication Parameters SubmenuBaud Submenu Parity SubmenuStop BIT Submenu BUS Format SubmenuModbus Protocol Submenu Line Feed SubmenuCommunication Interface Standard Submenu Data Flow Mode SubmenuData Separation Character Submenu Alarm Status Submenu Data Format SubmenuMain Reading Submenu Peak Value SubmenuAddress Setup Submenu Address Value SubmenuTransmit Time Interval Submenu Transmit Time Interval Value SubmenuAlarm 1 Display Color Submenu Display Color SelectionEnter Display Color Selection Menu Normal Color Display SubmenuAlarm 2 Display Color Submenu ExampleDisplay Colors change sequences Green RED AmberAmber RED Green Part Specifications Control Control OutputSSR CommunicationsInsulation ExcitationGeneral Input Properties Input Type Range AccuracyPart 5 Factory Preset Values Factory preset valueDisplay Color Colr Part CE Approvals Information Measurement CategoryPage Page Page Warranty/Disclaimer M4165/N/0206