a/d/s/ iSeries manual Address Setup Submenu, Address Value Submenu, Transmit Time Interval Submenu

Page 54

TEMPERATURE UNIT SUBMENU:

Includes a byte in the data string to indicate whether reading is in Celsius or Fahrenheit.

Press d 50) Display flashes previous selection for UNIT. Press b 51) Scroll through the available selections: NO, YES.

Press d 52) Display shows STRD stored message momentarily and then advances to ADDR only, if it was changed, otherwise press a to advance to ADDR Address Setup Submenu.

ADDRESS SETUP SUBMENU:

This menu is applicable to the RS-485 Option only.

Press d 53) Display advances to “Address Value” (0000 to 0199) Submenu.

ADDRESS VALUE SUBMENU:

Press d 54) Display flashes 1st digit of previously stored Address Value. Press b & c 55) Press b and c to enter new “Address Value”.

Press d 56) Display shows STRD stored message momentarily and then advances to TR.TM only, if it was changed, otherwise press a to advance to TR.TM Transmit Time Interval Submenu.

TRANSMIT TIME INTERVAL SUBMENU:

This menu is applicable if “Continuous” Mode was selected in the “Data Flow Mode” Submenu and the device is configured as an RS-232C Standard device. Also, one or more options under the Data Format Submenu must be enabled.

Press d 57) Display advances to “Transmit Time Value” Submenu.

TRANSMIT TIME INTERVAL VALUE SUBMENU:

Determines the interval at which data will be emitted to the RS-232 Serial Bus when the instrument is in Continuous Data Flow Mode.

Press d 58) Display flashes 1st digit of previous “Transmit Time Value” in seconds.

Press b & c 59) Press b and c to enter new “Transmit Time Value”, e.g. 0030 will send the data every 30 seconds in Continuous Mode.

Press d 60) Display shows STRD stored message momentarily and then advances to COLR only, if it was changed, otherwise press a to advance to COLR Color Display Selection Menu.

For more details, refer to the Communication Manual available at the website listed on the cover page of this manual or on the CD-ROM enclosed with your shipment.

50

Image 54
Contents Newport Electronics,Inc Newport Electronics,Inc Table of Contents List of Figures Flow Chart for Alarm 1 Menu Flow Chart for Alarm 2 MenuInput Properties Factory Preset Values List of TablesTIP Part Introduction DescriptionSafety Considerations EMC ConsiderationsBefore You Begin Part Setup Optional Remote Programmer Front PanelInput PowerFront Panel Connections OutputElectrical Installation Power Connections Main Power Connections Fuse Requirement see specificationsThermocouple Wiring Hookup TC Wire Color Chart Thermocouple Input3 Two/Three/Four-Wire RTD Ohm Wiring HookupProcess Current Process VoltageWiring Outputs Isolated Analog Output Wiring Hookup10 Snubber Circuits Wiring Hookup 13 Excitation Output Operation Configuration Mode Introduction Button Function in Configuration ModePart Menu Configuration Flow Chart for ID and Set PointsEntering Your NON-DEFAULT Full Security ID Number Entering Your NON-DEFAULT SETPOINT/ID Security ID NumberID Number Set Points SetpointConfiguration Menu Enter Configuration MenuInput Type Menu Thermocouple Submenu Input Type ThermocoupleEnter Input Type Menu Thermocouple TypesRTD Types Input Type RTDRTD Submenu RTD Value SubmenuProcess Submenu Reading Configuration MenuInput Type Process Process TypesDecimal Point Submenu Enter Reading Configuration MenuTemperature Unit Submenu Filter Constant SubmenuReading Configuration If Process was selected INPUT/READING Scale and Offset SubmenuConversion Table Range Conversion NumberAlarm Enter Alarm 1 MenuAlarm 1 ENABLE/DISABLE Submenu Alarm 1 ABSOLUTE/DEVIATION SubmenuAlarm 1 LATCH/UNLATCH Submenu Contact Closure Submenu Active SubmenuAlarm ENABLE/DISABLE AT Power on Alarm 1 LOW Value SubmenuAlarm 1 HI Value Submenu Enter Alarm 2 Menu Enter Loop Break Time Menu Loop Break ENABLE/DISABLE SubmenuLoop Break Time/Field Calibration Loop Break Time Value SubmenuReading Adjust Submenu Setpoint Deviation ENABLE/DISABLE SubmenuThermocouple Field Calibration Submenu Flow Chart for Output OutputEnter Output 1 Menu Self SubmenuMINIMUM/PERCENT LOW Submenu MAXIMUM/PERCENT High Submenu Control Type Output Relay, SSR, or PulseAction Type Submenu Auto PID Submenu Anti Integral SubmenuStart Auto Tune PID Proportional Band SubmenuReset Setup Submenu Rate Setup SubmenuCycle Time Submenu Damping Factor Submenu Deadband SubmenuAnalog Output Retransmission Enter Analog Output MenuCURRENT/VOLTAGE Submenu Reading OUTRamp & Soak Ramp ENABLE/DISABLE SubmenuSoak ENABLE/DISABLE Submenu Enter Ramp and Soak MenuRamp Value Submenu Soak Value SubmenuEntering or Changing Your NON-DEFAULT ID Code ID CodeEnter ID Code Menu Entering or Changing Your Default ID Code Full Security Level SubmenuSETPOINT/ID Security Level Submenu 12 Flow Chart for Communication Option Communication OptionBaud Submenu Enter Communication Option MenuCommunication Parameters Submenu Parity SubmenuModbus Protocol Submenu Stop BIT SubmenuBUS Format Submenu Line Feed SubmenuData Flow Mode Submenu Communication Interface Standard SubmenuData Separation Character Submenu Main Reading Submenu Alarm Status SubmenuData Format Submenu Peak Value SubmenuTransmit Time Interval Submenu Address Setup SubmenuAddress Value Submenu Transmit Time Interval Value SubmenuEnter Display Color Selection Menu Alarm 1 Display Color SubmenuDisplay Color Selection Normal Color Display SubmenuDisplay Colors change sequences Alarm 2 Display Color SubmenuExample Green RED AmberAmber RED Green Part Specifications SSR ControlControl Output CommunicationsExcitation InsulationGeneral Input Properties Input Type Range AccuracyPart 5 Factory Preset Values Factory preset valueDisplay Color Colr Part CE Approvals Information Measurement CategoryPage Page Page Warranty/Disclaimer M4165/N/0206