Ingersoll-Rand ETA2, ETA5 manual If you have selected H/O

Page 13

7529_eng.fm Page 12 Wednesday, November 12, 2003 8:57 AM

If you have selected H/O

Select units of measure- inlb, ftlb, etc. using

Note: High Output devices usually produce an output voltage of approximately 1-2 Volts. This analogue voltage signal will not be exact and may not be linear. If you are using a device that has a nominal 1Volt output, we suggest that you multiply the span of your transducer by 2.5 to determine the new span setting.

Type in your new span; press

Enter

Type in Serial Number; press

Enter

Screen will indicate ‘End of Set-up’; press MENU to return to main menu

Connect your high output device and select measure mode

(refer to instruction sheet for measure mode only)

Use an independent measuring device to apply a torque to the transducer and check the read- ing on the independent unit against the reading shown on the ETA. It will be necessary to return to the ‘Edit Tx’ menu option and adjust the span setting until the two readings agree. You may have to repeat the above procedure several times until your high output transducer is correctly calibrated.

If you have selected KPD

Select location of decimal point using

then press Enter

 

Select Parity- Odd, Even or Disabled using

then press

Enter

To set-up another transducer press MENU

1

 

Select another memory (B to I) using the

keys and press

Enter

To return to main menu press MENU

 

 

12

Image 13
Contents Operators Manual Operators Manual for Model ETA2 and ETA5 Analyzers Placing Analyzer in Service 7529eng.fm Page 3 Wednesday, November 12, 2003 857 AM Save These INSTRUCTIONS. do not Destroy Table of Contents 7529eng.fm Page 6 Wednesday, November 12, 2003 857 AM Following method is adopted throughout this manual How to use this ManualPacking List Care and Storage Battery ChargingFor ETA2 Edit Transducer FileFor ETA5 If you have chosen I/S If you have selected H/O If you have selected KEY Barcode read input for characteristic Names and commentsMeasure Mode Up to 200 readings stored Type in amount of Dip press Printing of stored readings from measure mode Measure Mode Circle-Store OptionTo select this option From main menu Press Direct Direct Measure Setup 7529eng.fm Page 18 Wednesday, November 12, 2003 857 AM To set up a Characteristic ETA5 only Select Frequency Response using Then press Enter Through the available characteristics and press Direct Characteristic Mode ETA5 only7529eng.fm Page 22 Wednesday, November 12, 2003 857 AM To Store Data single characteristic ETA5 only 7529eng.fm Page 24 Wednesday, November 12, 2003 857 AM Screen will show Characteristic No 2 of ? To set up a Round ETA5 onlyTo Store Data by Round ETA5 only Set-up To set up a Master Round ETA5 onlyTo Store Data by Master Round Next job, it will switch back when you pressTo return to the main menu, press Menu Menu If you haveTo Display Recorded Data on ETA5 To Display Recorded Data on ETA2To View Cp and Cpk French Cp and Cpk To Erase Readings Stored in Measure ModeTo Erase Stored Data To set this functionTo Erase Characteristic Setup To Erase Stored Data and Setups on ETA5To Erase Characteristic Data To Erase Round Data To Erase Round Setup InformationPrinting To print details of Setup or Analysis of Data from ETA5To print details of readings by round To print details of all characteristic setups To print details of setup and analysis of readings by roundTo print readings taken by characteristic To print details of a particular round setupThen press Print Measure To Set Date and Time Printing Data to PCTo Configure Printer Port To Set Power Off and Backlight Time DelaysTo Display Software Version Number To Set Password ProtectionDisplay window Pressing Will remove the protection Screen will say Confirm password To Clear Password ProtectionTo Perform Software Reset Cm min samples To Communicate with a PC using PC Comms SoftwareExample Printout of Readings taken in Measure Mode Printout of Setup and Readings taken by Characteristic Printout of Analysis by Characteristic Printout of details of all Characteristic setups Printout of Round setup Printout of Readings taken by Characteristic Rotary and Stationary Joint Kits Deg/lbft Lbft Nominal Joint StiffnessesDwg. TPA1919 Not Parts ListPictured Callout Description Joint Kit NumberGlossary of Terms CharacteristicCpK Horizontal Sampling High Output Transducer H/OIndustry Standard Transducer I/S Max Torque ValueUnits of Measure Threshold Torque ValueVertical Plus Prompt Code Message Explanation Error MessagesINCOR-TX OP Code ERR Dwg. TPA1822 External Connections7529eng.fm Page 62 Wednesday, November 12, 2003 857 AM 7529eng.fm Page 63 Wednesday, November 12, 2003 857 AM 7529eng.fm Page 64 Wednesday, November 12, 2003 857 AM 0102