Ingersoll-Rand ETA5, ETA2 manual Incor-Tx

Page 60

7529_eng.fm Page 59 Wednesday, November 12, 2003 8:57 AM

Code

Message

Explanation

123

AZ DISCR

An auto-zero calibration on the transducer has

 

 

detected an excessive discrepancy between the

 

 

Tracking and Peak inputs.

124

Tx FAULT

Adc saturation due to transducer fault, no

 

 

transducer fitted or over torque.

200

INCOR-TX

The correct transducer is not connected.

300

IN ROUND

This characteristic has already been included in

 

 

the round.

301

CHARACTS

A minimum of two characteristics must be defined

 

 

before entering Edit Round.

306

RS232

Invalid RS232 data on the serial input.

400

INVALID

Incorrect time or date format.

600

ONLY 1

Only one round programmed so not possible to

 

 

generate master round.

601

NO RND

No rounds programmed so not possible to gener-

 

 

ate master round.

602

IN M RND

This round has already been entered into the

 

 

master round. It is not possible to enter a round

 

 

twice.

701

RND COMP

The round is complete, no skipping is allowed.

702

ALL SKIP

All other characteristics have already been

 

 

skipped.

800

MEMORY

Message on power up if memory error found.

 

 

(See note below)

801

MENU ERR

Invalid menu entered due to fault.

802

PROM CKS

The EPROM has an error.

59

Image 60
Contents Operators Manual Operators Manual for Model ETA2 and ETA5 Analyzers Placing Analyzer in Service 7529eng.fm Page 3 Wednesday, November 12, 2003 857 AM Save These INSTRUCTIONS. do not Destroy Table of Contents 7529eng.fm Page 6 Wednesday, November 12, 2003 857 AM How to use this Manual Following method is adopted throughout this manualPacking List Battery Charging Care and StorageEdit Transducer File For ETA2For ETA5 If you have chosen I/S If you have selected H/O Barcode read input for characteristic Names and comments If you have selected KEYMeasure Mode Up to 200 readings stored Type in amount of Dip press Measure Mode Circle-Store Option Printing of stored readings from measure modeTo select this option From main menu Press Direct Direct Measure Setup 7529eng.fm Page 18 Wednesday, November 12, 2003 857 AM To set up a Characteristic ETA5 only Select Frequency Response using Then press Enter Direct Characteristic Mode ETA5 only Through the available characteristics and press7529eng.fm Page 22 Wednesday, November 12, 2003 857 AM To Store Data single characteristic ETA5 only 7529eng.fm Page 24 Wednesday, November 12, 2003 857 AM To set up a Round ETA5 only Screen will show Characteristic No 2 of ?To Store Data by Round ETA5 only To set up a Master Round ETA5 only Set-upNext job, it will switch back when you press To Store Data by Master RoundTo return to the main menu, press Menu Menu If you haveTo Display Recorded Data on ETA2 To Display Recorded Data on ETA5To View Cp and Cpk To Erase Readings Stored in Measure Mode French Cp and CpkTo Erase Stored Data To set this functionTo Erase Stored Data and Setups on ETA5 To Erase Characteristic SetupTo Erase Characteristic Data To Erase Round Setup Information To Erase Round DataTo print details of Setup or Analysis of Data from ETA5 PrintingTo print details of readings by round To print details of setup and analysis of readings by round To print details of all characteristic setupsTo print details of a particular round setup To print readings taken by characteristicThen press Print Measure Printing Data to PC To Set Date and TimeTo Set Power Off and Backlight Time Delays To Configure Printer PortTo Set Password Protection To Display Software Version NumberDisplay window Pressing Will remove the protection To Clear Password Protection Screen will say Confirm passwordTo Perform Software Reset To Communicate with a PC using PC Comms Software Cm min samplesExample Printout of Readings taken in Measure Mode Printout of Setup and Readings taken by Characteristic Printout of Analysis by Characteristic Printout of details of all Characteristic setups Printout of Round setup Printout of Readings taken by Characteristic Rotary and Stationary Joint Kits Nominal Joint Stiffnesses Deg/lbft LbftDwg. TPA1919 Parts List NotPictured Callout Description Joint Kit NumberCharacteristic Glossary of TermsCpK High Output Transducer H/O Horizontal SamplingIndustry Standard Transducer I/S Max Torque ValueThreshold Torque Value Units of MeasureVertical Plus Prompt Error Messages Code Message ExplanationINCOR-TX OP Code ERR External Connections Dwg. TPA18227529eng.fm Page 62 Wednesday, November 12, 2003 857 AM 7529eng.fm Page 63 Wednesday, November 12, 2003 857 AM 7529eng.fm Page 64 Wednesday, November 12, 2003 857 AM 0102