APC AXFA11G user manual Accessories

Page 113

11. OUTLINE

ACCESSORIES

Fuse (Time lag fuse): 1 pc.

Power supply code 1; T2.5 A, 250 V

Power supply code 2; T3.15A, 250 V Mounting bracket: 1 set

MODEL AND SUFFIX CODE

AXFA11 Magnetic Flowmeter Remote Converter:

Model

 

 

Suffix Code

 

Description

 

 

 

 

 

AXFA11

 

 

 

Magnetic Flowmeter Remote Converter

 

 

 

 

 

 

 

G · · · · · · · · ·

General-Purpose Use

Use

·For AXF Remote Flowtube of size

 

 

 

 

2.5 to 2600 mm (0.1 in. to 104 in.)

 

 

 

 

 

Output Signal

 

-D · · · · · · ·

·4 to 20 mA DC,

 

 

BRAIN Communication

and

 

-E · · · · · · ·

·4·to 20 mA DC,

Communication

 

 

 

 

HART Communication

 

 

 

1 · · · · · ·

100 V to 240 V AC or

Power Supply

· ·

 

 

2 · · · · · ·

100 to 120 V DC

 

 

 

· ·

 

(*1)

 

 

 

 

24 V AC/DC

 

Electrical

 

-0 · · · · ·JIS· G1/2 female

 

-2 · · · · ·ANSI·

1/2 NPT female

Connections

 

 

-4 · · · · ·ISO· M20  1.5 female

 

 

 

Indicator

 

1 · · · ·With Indicator

 

 

 

 

 

Option

 

/

Optional code (See the Table of

 

 

Optional Specifications)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T03.EPS

*1: In case of power supply code 2 (24 V AC/DC), optional code A (lighting protector) is mandatory.

Signal Cable:

Model

Suffix Code

 

Description

 

 

 

 

 

 

 

 

 

 

 

Magnetic Flowmeter

 

AXFC

·· · · ·· · · · ·

· ·

 

 

 

 

Dedicated Signal cable for the

 

 

 

 

 

ADMAG AXF series

 

 

 

 

 

 

 

-0 · · ········ ·

 

No Termination. A set of termination

 

Termination

 

 

 

parts for M4 screws is attached.

 

 

 

 

Terminated for the AXFA11/14

 

 

-4 · · · ·· ·· ·· ·

 

 

 

 

 

 

Converter.

 

 

 

 

 

Designate the cable length, unit: m

 

 

 

 

 

Following “L”, specify the cable in

 

 

 

 

 

three digits as multiple of 1 meter

 

 

 

 

 

(e.g., 001, 002, or 005)

 

Cable

-L

··· ·

 

for a length up to 5 m, or as

 

Length

 

a multiple of 5 meters (i.e., 005,

 

 

 

 

 

010, 015, or the like).

 

 

 

 

 

The maximum cable length:

 

 

 

 

 

200 m for combined use with AXFA11

 

 

 

 

 

100 m for combined use with AXFA14

 

 

 

 

 

 

 

 

 

/C

 

With termination parts sets.

 

Option

 

 

Following “C”, specify the q’ty of sets

 

 

 

 

 

of termination parts in one digits.

 

 

 

 

 

 

 

 

 

 

 

T04.EPS

Note: • The cable is constructed with double shielding over the two conductors, and uses heat-resistant vinyl as the outer covering material.

Finished diameter: 10.5 mm (0.413 in.)

Maximum temperature: +80˚C (+176˚F)

Unnecessary to order the above cable for submersible type flowtube or for the optional code DHC flowtube because the flowtube is wired with 30 m (98 ft) cable.

For excitation cable, prepare a two-core cable at the customer side.

11-5

IM 01E20C01-01E

Image 113
Contents Users ManualAXFA11G Contents Operation VIA Hart Communicator Revision Record Introduction Safety and Modification PrecautionsGuarantee will not apply in the following cases Using the Magnetic Flowmeter SafelyWarranty Combination Remote Flowtubes Checking Model and Specifications Handling PrecautionsAccessories Storage PrecautionsInstallation Location InstallationMounting Surface Mounting Wall MountingPanel Mounting Inch Pipe MountingInch pipe Mounting fixture Washer Clamp screw Unit mm approx. inch Panel cutout 203 Mounting fixture 172Wiring Precautions WiringDedicated Signal Cable Axfc CablesExcitation Cable/Power Cable/Output Cable Wiring PortsUnit mm approx. inch Crimp Terminal On the converter sideWiring using waterproof glands When there are no particular optional specificationsConduit Wiring Plastic glandRemoving Cover Wiring ConnectionsTerminal Configuration DC Power Connection Precautions for Wiring of Power Supply CablesAxfc Dedicated signal cable Excitation cable Wiring the Remote Flowtube with the AXFA11 ConverterGrounding Protective grounding terminalsTo 20 mA DC Current Output Connecting to External Instru- mentsStatus Input Pulse Output13 Status Output Connection Status Output / Alarm OutputData display area Operating Panel Configuration and FunctionsSetting switch operations Display items ContentReversed-character indication NoneDisplay Mode → Setting Mode Display Unit Setting MethodsSetting Mode To Parameter Search Mode Setting Example for Selection-Type Data Flow rate units Parameter Setting ProcedureSetting Mode Sub-item Parameter Search Mode Sub-item Current unit settingB21 Base Flow Unit Parameter Sub-itemSetting Mode Condition Setting Example for Numeric-Type Data Flow rate spanMajor Item Parameter Search Mode Setting mode B23 Flow SpanSetting Mode Condition Setting Example for Alphanumeric-Type Data Tag numberSub-item Parameter Search Mode Major item Incorrect settingModified value Search Mode Selection CSub-item Major item Parameter Description Item a Menu a Display items Item B Menu B Easy Setup itemsIndicated item Default value Units Description Display unitName Data range Name Data range Position Default value Units Display unit Item C Menu C Basic Setting itemsDefault value Units Display unit Item D Menu D Total Setting itemsName Data range Position Default value Display unit NameItem F Menu F Status Functions Setting items Item E Menu E Pulse Setting itemsItem G Menu G Alarm Setting items ALM HYS ALM-SETTINGALM OUT ACT ALM-SIG OverItem J Menu J Auxiliary Function Setting items Item H Menu H Display Setting itemsUnits Display unit Item K Menu K Diagnostic Function Setting itemsItem P Menu P Parameter Protection items Item N Menu N Loop Test Setting itemsData Range Menu B Easy Setup itemsUnit/P 10-9 FU u Unit/P 10-6 FU Pulse/s Example When 1000 m3/h is set for B23/C42 Flow SpanDescription B41/H11 Display Select2 and B42/H12 Display Menu C Basic Setting itemsMeter Factor Settings C47 User Span Select Selection of the use Menu D Total Setting itemsC49 Flow User Span Is set forB22/C41 Base Time UnitF13 TOTAL, and A32 DIF Total when BrainFlow Unit Is set forB22/C41 Base TimeE20 Pulse Active Mode Menu E Pulse Setting itemsSetting for E20 Pulse Active Mode Setting Function Description Menu F Status Functions Setting itemsSwitch Lower and D24 Ttl Switch Upper G33 Alm-HH/LL to YesVal Upper No Function No input functionMultiple Ranges Setting Automatic multiple ranges switching Multiple ranges settingStatus Output for Automatic Multiple Ranges Switching 1 Status Output for Automatic Multiple Ranges SwitchingFor automatic multiple ranges switching Parameter setting sequenceF14 SO1/2 Active Mode F10 SO1 Function F11 SO2 FunctionF15 2 Multiple Ranges Switching via External Status InputG11 High Alarm High alarm setting Menu G Alarm Setting itemsG10 Low Alarm Low alarm setting 100 High Alarm High High Alarm100 High Alarm Instantaneous flow rate Low Low Alarm Time t Output Example35% Low Alarm 100 85% 75% High Alarm100 25%Example Setting FunctionG22 4-20mA Burn Out Display of the current ClosedOn OpenOffAlarm Items G41 Alm Record1 Alarm record1G43 Alm Record2 Alarm record2 G45 Alm Record3 Alarm record3Setting range 0 to 10% Hysteresis Menu J Auxiliary Function Setting itemsDetermining rate limit value and dead time Function orExample 1 Step input Example 2 Slurry noiseMenu K Diagnostic Function Setting items J31 Power FrequencyK10 Adhesion Check Setting of adhesion Power Supply Code 100 to 240 V AC or 100 to 120 V DCMenu M Automatic Zero Adjustment Function Setting items Adhesion Diagnostic FunctionMenu N Loop Test Setting items SO1 Function or F11 SO2 Function, thenDefault setting Enable N11 Test Output Value Setting for test outputMenu P Parameter Protection items Write Protect functionAfter password setting Default settingOut Act Mode G20 Alm Out Act Mode Setting of the activeDisplay and output condition for process alarms Alarm description Selection Display and output condition for setting alarm occurrencesTotali Pulse Display Alarm Output UnitAlarms & Warning Messages K12, K13 and K14 are correct BT200 Basic Operations Operation VIA Brain Terminal BT200Key Descriptions Key Layout and DisplayKey-in Sequence Command FunctionFunction Keys 1 BT200 Connection AXFA11 Operation Using a BT200Data update and upload/ download function of BT200 3 BT200 Screens & Flow Rate Data Display Parameter Setting Using a BT200B21 Flow UNIT, press 1 BT200 Setting of Selection-Type Data Flow rate unitsMoved to B Easy Setup B23 Flow SPAN, press 2 BT200 Setting of Numeric-Type Data Flow rate spanMoved to C Basic Setup 3 BT200 Setting of Alphanumeric-Type Data Tag numberTo C Basic Setup Operation VIA Hart Communicator Conditions of Communication Line Specifications for Communication LineInterconnectwion between AXFA11 and Hart Communicator Communication Line RequirementsKeys and Functions Basic Operation of the Hart Communicator ModelKey Operation DisplayCalling Up Menu Addresses Parameter Configuration Entering, Setting and Sending DataParameters Data RenewingExample of Setting Selection-Type Data Base Flow Unit Checking for ProblemsSetting Parameters PV Span Example of Setting Numeric-Type Data PV SpanDisplay Example of Setting Alphanumeric-Type Data Tag NumberSelect MF Set Example 1 of Other Settings Meter FactorSelect Power Frequency Trim Example 3 of Other Settings Trim Analog OutputReturning fld dev to original output appears Select the Scaled D/A trim Returning fld dev to original output appears Setting of Burst Mode Example 4 of Other Settings Burst ModeExiting Burst Mode Example 5 of Other Settings Multidrop ModePolling Example 6 of Other Settings Write Protection Exiting Multidrop ModeOpen Wrt Protect Menu in the Hot key Menu Setting PasswordEnable Wrt 10min Changing PasswordOther Functions and Operations Joker password and Software SealOn next Menu TreeE12 J24 J30 P22 P23 Actual Operation Pre-operation Zero AdjustmentSelection D Major item Sub- item selection D Zero Adjustment Using Display Unit SwitchesZero Adjustment via External Status Input Ext Auto Zero Components Replacement MaintenanceDisplay Unit Replacement Fuse ReplacementAmplifier Replacement Setting of SwitchesSetting of Burnout Switch Output Setting Pins for Burnout4 Switch Configuration Setting of Write Protect SwitchNo Indication TroubleshootingStart Unstable ZeroDisagreement Between Indication and Actual Flow 65 10 Cf 10,000 OutlineFunctions How to Set ParametersStandard Performance Safety Requirement Standards Maximum Power ConsumptionPower Supply Power supply code Size 2200 mm 88 in. to 2600 mm 104Accessories Terminal Wiring Optional Specifications for AXFA11 Remote ConverterSurface mounting Remote Converter AXFA11GGeneral tolerance in the dimensional outline drawing Remote Converter AXFA11G /RK AM11 Replacement BracketAXFC-0-L Dedicated Signal Cable AXFC-4-LEdition Date Revised item Revision Record