6. PARAMETER DESCRIPTION
Parameter setting sequence (for multiple ranges switching via external status input)
F10: SO1 Function | F11: SO2 Function | F12: SI1 Function | F13: SI2 Function |
Select a function | Select a function | Select a function | Select a function |
F14: SO1/2 Active Mode |
| F15: SI1/2 Active Mode |
|
Select whether SO1 and SO2 output is to be “Closed (on) Act” or | Select whether SI1 and SI2 input is to be “Short Active” | ||
“Open (off) Act.” |
| or “Open Active.” |
|
Function Selection |
| Function Selection |
|
No Function: Output is stopped (*: other functions may be selected) | No Function: No input function (*: other functions may be | ||
Fwd/Rev Ranges: Forward and reverse flow rate measurement | selected) |
| |
Ext 2 Answer: Answerback; | Ext 2 Ranges: | ||
Ext 3 Answer: Answerback; | Ext 3 Ranges: | ||
Ext 4 Answer: Answerback; | Ext 4 Ranges: |
Refer to Table 6.4.2 and set F10 and F11 or F12 and F13.
B23:Flow Span |
| F30:Forward |
|
| Span2 |
Forward No. 1 range |
| Forward No. 2 range |
|
|
|
|
|
|
F33:Reverse |
| F34:Reverse |
Span1 |
| Span2 |
Reverse No. 1 range |
| Reverse No. 2 range |
|
|
|
F31:Forward
Span3
Forward No. 3 range
F35:Reverse
Span3
Reverse No. 3 range
F32:Forward
Span4
Forward No. 4 range
F36:Reverse
Span4
Reverse No. 4 range
Set the spans for the ranges to be used.
No. 1 range No. 2 range No. 3 range No. 4 range
If “instantaneous flow rate % display (FR)” is selected for the Display Mode and A10: FLOW RATE(%) is selected for BRAIN communication, the instantaneous flow rate % and following symbols will be displayed only for multiple ranges and forward/reverse flow measurement.
Forward No. 1 range : [F1] | Reverse No. 1 range : [R1] |
Forward No. 2 range : [F2] | Reverse No. 2 range : [R2] |
Forward No. 3 range : [F3] | Reverse No. 3 range : [R3] |
Forward No. 4 range : [F4] | Reverse No. 4 range : [R4] |
F0609.EPS
[F40: Auto Range Hys] Setting of automatic range- switching hysteresis width
Automatic switching takes place for multiple ranges switching when 100% of the range is exceeded, and this parameter allows a hysteresis width to be set for this switching.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis Widths.
[F41: Bi Direction Hys] Setting of forward/reverse flow measurement hysteresis width
This parameter sets the hysteresis for forward/reverse flow rate measurement as a % value of the minimum flow span.
Refer to Figure 6.4.1: Multiple Ranges and Hysteresis Widths.
(6) Menu G: Alarm Setting items
(Refer to Section 6.5: Alarm Functions for more details.)
Menu G principally contains setting items relevant to alarms.
[G10: Low Alarm] Low alarm setting
This parameter sets the low limit (L) alarm value, and this is done using a % value of the maximum span.
A setting value of
[G11: High Alarm] High alarm setting
This parameter sets the high limit (H) alarm value, and this is done using a % value of the maximum span.
A setting value of 110% indicates that the alarm is disabled.
[G12: Low Low Alarm]
A setting value of
IM |