| 6. PARAMETER DESCRIPTION | |
|
|
|
|
|
|
|
|
|
| Name |
| Data range |
| Position | Default value |
|
|
|
|
|
| Units |
|
| ||
Item | Display unit | R/W | Display unit | of decimal | Description |
| ||
(*): Indicated item |
| |||||||
|
| |||||||
| (BRAIN) |
| /BRAIN |
| point |
|
|
|
|
|
|
|
|
|
|
|
|
J30 | Power Synch | W | No |
|
| Yes | Selects whether or not the internal |
|
| (POWER SYNCH) |
| Yes |
|
|
| frequency is to be synchronized |
|
|
|
|
|
|
|
| with the power supply frequency. |
|
|
|
|
|
|
|
|
|
|
J31 | Power Frequency | R/W | 47.00 to 63.00 | Hz | 2 | 50.00 | Displays the |
|
| (POWER FREQ) |
|
|
|
|
| frequency (for Power Synch = |
|
|
|
|
|
|
|
| “Yes”), or sets the |
|
|
|
|
|
|
|
| frequency (for Power Synch=“No”). |
|
|
|
|
|
|
|
|
|
|
J40 | Memo 1 | W | ASCII 16 characters |
|
|
| Memo field |
|
| (MEMO 1) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
J41 | Memo 2 | W | ASCII 16 characters |
|
|
| Memo field |
|
| (MEMO 2) |
|
|
|
|
|
|
|
J42 | Memo 3 | W | ASCII 16 characters |
|
|
| Memo field |
|
| (MEMO 3) |
|
|
|
|
|
|
|
J50 | Software Rev No | R | — |
|
|
| Software revision number |
|
| (SOFTWARE REV) |
|
|
|
|
|
|
|
J60 | — | R | Good |
|
|
| See “6.5 Alarm Functions”. |
|
| (SELF CHECK) |
| Error |
|
|
|
|
|
(10) Item K (Menu K): Diagnostic Function Setting items
Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.
| Name |
| Data range |
| Position | Default value |
| |
|
|
|
| Units |
| |||
Item | Display unit | R/W | Display unit | of decimal | Description | |||
(*): Indicated item | ||||||||
| (BRAIN) |
| /BRAIN |
| point |
|
| |
K00 | Diagnosis |
|
|
|
|
|
| |
| (DIAGNOSIS) |
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |
K10 | Adhesion Check | W | No |
|
| No | Selects whether or not to perform diagnosis | |
| (ADHESION CHK) |
| Yes |
|
|
| of adhesion to the electrode. | |
K11 | Adhesion Level1 | W | 0.00 to 100.00 | M ohm | 2 | 0.10 | Sets the resistance value for adhesion | |
| (ADH LEVEL1) |
|
|
|
|
| Level 1 to the electorode. | |
|
|
|
|
|
|
|
| |
K12 | Adhesion Level2 | W | 0.00 to 100.00 | M ohm | 2 | 0.50 | Sets the resistance value for adhesion | |
| (ADH LEVEL2) |
|
|
|
|
| Level 2 to the electorode. | |
K13 | Adhesion Level3 | W | 0.00 to 100.00 | M ohm | 2 | 1.00 | Sets the resistance value for adhesion | |
| (ADH LEVEL3) |
|
|
|
|
| Level 3 to the electorode. | |
K14 | Adhesion Level4 | W | 0.00 to 100.00 | M ohm | 2 | 3.00 | Sets the resistance value for adhesion | |
| (ADH LEVEL4) |
|
|
|
|
| Level 4 to the electorode. | |
|
|
|
|
|
|
|
| |
K15 | Adh Measure Value | R | — | M ohm | 2 |
| Displays the resistance value for | |
| (ADH MEAS VAL) |
|
|
|
|
| adhesion to the electrode. | |
K60 | — | R | Good |
|
|
| See “6.5 Alarm Functions”. | |
| (SELF CHECK) |
| Error |
|
|
|
|
T0611.EPS
(11) Item M (Menu M): Automatic
Menu M contains items that are relevant to automatic zero adjustment.
| Name |
| Data range |
| Position | Default value |
| |
|
|
|
| Units |
| |||
Item | Display unit | R/W | Display unit | of decimal | Description | |||
(*): Indicated item | ||||||||
| (BRAIN) |
| /BRAIN |
| point |
|
| |
M00 | Adjustment |
|
|
|
|
|
| |
| (ADJUSTMENT) |
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |
M10 | Auto Zero Exe | W | No Execution |
|
| No Execution | Selects whether or not automatic | |
| (AUTOZERO EXE) |
| Execution |
|
|
| zero adjustment is carried out. | |
|
|
|
|
|
|
| Linked with B50. | |
M11 | Magflow Zero | R/W |
| 3 | 0.000 | Displays the result of the automatic | ||
| (MAGFLOW ZERO) |
|
|
|
|
| zero adjustment, or sets the zero point. | |
|
|
|
|
|
|
|
| |
M60 | — | R | Good |
|
|
| See “6.5 Alarm Functions”. | |
| (SELF CHECK) |
| Error |
|
|
|
| |
|
|
|
|
|
|
|
|
T0612.EPS