APC AXFA11G user manual Item K Menu K Diagnostic Function Setting items, Units Display unit

Page 38

 

 

 

 

 

 

 

6. PARAMETER DESCRIPTION

 

 

 

 

 

 

 

 

 

 

Name

 

Data range

 

Position

Default value

 

 

 

 

 

 

Units

 

 

Item

Display unit

R/W

Display unit

of decimal

Description

 

(*): Indicated item

 

 

 

 

(BRAIN)

 

/BRAIN

 

point

 

 

 

 

 

 

 

 

 

 

 

 

J30

Power Synch

W

No

 

 

Yes

Selects whether or not the internal

 

 

(POWER SYNCH)

 

Yes

 

 

 

frequency is to be synchronized

 

 

 

 

 

 

 

 

with the power supply frequency.

 

 

 

 

 

 

 

 

 

 

J31

Power Frequency

R/W

47.00 to 63.00

Hz

2

50.00

Displays the power-supply

 

 

(POWER FREQ)

 

 

 

 

 

frequency (for Power Synch =

 

 

 

 

 

 

 

 

“Yes”), or sets the power-supply

 

 

 

 

 

 

 

 

frequency (for Power Synch=“No”).

 

 

 

 

 

 

 

 

 

 

J40

Memo 1

W

ASCII 16 characters

 

 

 

Memo field

 

 

(MEMO 1)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

J41

Memo 2

W

ASCII 16 characters

 

 

 

Memo field

 

 

(MEMO 2)

 

 

 

 

 

 

 

J42

Memo 3

W

ASCII 16 characters

 

 

 

Memo field

 

 

(MEMO 3)

 

 

 

 

 

 

 

J50

Software Rev No

R

 

 

 

Software revision number

 

 

(SOFTWARE REV)

 

 

 

 

 

 

 

J60

R

Good

 

 

 

See “6.5 Alarm Functions”.

 

 

(SELF CHECK)

 

Error

 

 

 

 

 

T0610-2.EPS

(10) Item K (Menu K): Diagnostic Function Setting items

Menu K contains items that are relevant to the diagnosis of insulation adhesion to the electrode.

 

Name

 

Data range

 

Position

Default value

 

 

 

 

 

Units

 

Item

Display unit

R/W

Display unit

of decimal

Description

(*): Indicated item

 

(BRAIN)

 

/BRAIN

 

point

 

 

K00

Diagnosis

 

 

 

 

 

 

 

(DIAGNOSIS)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

K10

Adhesion Check

W

No

 

 

No

Selects whether or not to perform diagnosis

 

(ADHESION CHK)

 

Yes

 

 

 

of adhesion to the electrode.

K11

Adhesion Level1

W

0.00 to 100.00

M ohm

2

0.10

Sets the resistance value for adhesion

 

(ADH LEVEL1)

 

 

 

 

 

Level 1 to the electorode.

 

 

 

 

 

 

 

 

K12

Adhesion Level2

W

0.00 to 100.00

M ohm

2

0.50

Sets the resistance value for adhesion

 

(ADH LEVEL2)

 

 

 

 

 

Level 2 to the electorode.

K13

Adhesion Level3

W

0.00 to 100.00

M ohm

2

1.00

Sets the resistance value for adhesion

 

(ADH LEVEL3)

 

 

 

 

 

Level 3 to the electorode.

K14

Adhesion Level4

W

0.00 to 100.00

M ohm

2

3.00

Sets the resistance value for adhesion

 

(ADH LEVEL4)

 

 

 

 

 

Level 4 to the electorode.

 

 

 

 

 

 

 

 

K15

Adh Measure Value

R

M ohm

2

 

Displays the resistance value for

 

(ADH MEAS VAL)

 

 

 

 

 

adhesion to the electrode.

K60

R

Good

 

 

 

See “6.5 Alarm Functions”.

 

(SELF CHECK)

 

Error

 

 

 

 

T0611.EPS

(11) Item M (Menu M): Automatic Zero-Adjustment Function Setting items

Menu M contains items that are relevant to automatic zero adjustment.

 

Name

 

Data range

 

Position

Default value

 

 

 

 

 

Units

 

Item

Display unit

R/W

Display unit

of decimal

Description

(*): Indicated item

 

(BRAIN)

 

/BRAIN

 

point

 

 

M00

Adjustment

 

 

 

 

 

 

 

(ADJUSTMENT)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M10

Auto Zero Exe

W

No Execution

 

 

No Execution

Selects whether or not automatic

 

(AUTOZERO EXE)

 

Execution

 

 

 

zero adjustment is carried out.

 

 

 

 

 

 

 

Linked with B50.

M11

Magflow Zero

R/W

-99.999 to 99.999

 

3

0.000

Displays the result of the automatic

 

(MAGFLOW ZERO)

 

 

 

 

 

zero adjustment, or sets the zero point.

 

 

 

 

 

 

 

 

M60

R

Good

 

 

 

See “6.5 Alarm Functions”.

 

(SELF CHECK)

 

Error

 

 

 

 

 

 

 

 

 

 

 

 

T0612.EPS

6-10

IM 01E20C01-01E

Image 38
Contents Users ManualAXFA11G Contents Operation VIA Hart Communicator Revision Record Safety and Modification Precautions IntroductionGuarantee will not apply in the following cases Using the Magnetic Flowmeter SafelyWarranty Combination Remote Flowtubes Accessories Handling PrecautionsChecking Model and Specifications Storage PrecautionsMounting InstallationInstallation Location Surface Mounting Wall MountingInch pipe Mounting fixture Washer Clamp screw Inch Pipe MountingPanel Mounting Unit mm approx. inch Panel cutout 203 Mounting fixture 172Dedicated Signal Cable Axfc WiringWiring Precautions CablesUnit mm approx. inch Crimp Terminal Wiring PortsExcitation Cable/Power Cable/Output Cable On the converter sideConduit Wiring When there are no particular optional specificationsWiring using waterproof glands Plastic glandRemoving Cover Wiring ConnectionsTerminal Configuration Precautions for Wiring of Power Supply Cables DC Power ConnectionGrounding Wiring the Remote Flowtube with the AXFA11 ConverterAxfc Dedicated signal cable Excitation cable Protective grounding terminalsConnecting to External Instru- ments To 20 mA DC Current OutputPulse Output Status InputStatus Output / Alarm Output 13 Status Output ConnectionData display area Operating Panel Configuration and FunctionsSetting switch operations Reversed-character indication ContentDisplay items NoneDisplay Unit Setting Methods Display Mode → Setting ModeSetting Mode To Parameter Search Mode Setting Example for Selection-Type Data Flow rate units Parameter Setting ProcedureSetting Mode B21 Base Flow Unit Sub-item Current unit settingSub-item Parameter Search Mode Parameter Sub-itemMajor Item Parameter Search Mode Setting mode Setting Example for Numeric-Type Data Flow rate spanSetting Mode Condition B23 Flow SpanSub-item Parameter Search Mode Major item Setting Example for Alphanumeric-Type Data Tag numberSetting Mode Condition Incorrect settingModified value Search Mode Selection CSub-item Major item Parameter Description Item B Menu B Easy Setup items Item a Menu a Display itemsIndicated item Default value Units Description Display unitName Data range Item C Menu C Basic Setting items Name Data range Position Default value Units Display unitName Data range Position Default value Display unit Item D Menu D Total Setting itemsDefault value Units Display unit NameItem E Menu E Pulse Setting items Item F Menu F Status Functions Setting itemsItem G Menu G Alarm Setting items ALM OUT ACT ALM-SETTINGALM HYS ALM-SIG OverItem H Menu H Display Setting items Item J Menu J Auxiliary Function Setting itemsItem K Menu K Diagnostic Function Setting items Units Display unitItem N Menu N Loop Test Setting items Item P Menu P Parameter Protection itemsMenu B Easy Setup items Data RangeExample When 1000 m3/h is set for B23/C42 Flow Span Unit/P 10-9 FU u Unit/P 10-6 FU Pulse/sDescription Menu C Basic Setting items B41/H11 Display Select2 and B42/H12 DisplayMeter Factor Settings C49 Flow User Span Menu D Total Setting itemsC47 User Span Select Selection of the use Is set forB22/C41 Base Time UnitFlow Unit TOTAL, and A32 DIF Total when BrainF13 Is set forB22/C41 Base TimeE20 Pulse Active Mode Menu E Pulse Setting itemsSetting for E20 Pulse Active Mode Switch Lower and D24 Ttl Switch Upper Menu F Status Functions Setting itemsSetting Function Description G33 Alm-HH/LL to YesNo Function No input function Val UpperStatus Output for Automatic Multiple Ranges Switching Multiple ranges settingMultiple Ranges Setting Automatic multiple ranges switching 1 Status Output for Automatic Multiple Ranges SwitchingF14 SO1/2 Active Mode Parameter setting sequenceFor automatic multiple ranges switching F10 SO1 Function F11 SO2 Function2 Multiple Ranges Switching via External Status Input F15G11 High Alarm High alarm setting Menu G Alarm Setting itemsG10 Low Alarm Low alarm setting 100 High Alarm Instantaneous flow rate Low Low Alarm Time t High High Alarm100 High Alarm Output Example100 100 85% 75% High Alarm35% Low Alarm 25%G22 4-20mA Burn Out Display of the current Setting FunctionExample ClosedOn OpenOffG43 Alm Record2 Alarm record2 G41 Alm Record1 Alarm record1Alarm Items G45 Alm Record3 Alarm record3Menu J Auxiliary Function Setting items Setting range 0 to 10% HysteresisExample 1 Step input Function orDetermining rate limit value and dead time Example 2 Slurry noiseK10 Adhesion Check Setting of adhesion J31 Power FrequencyMenu K Diagnostic Function Setting items Power Supply Code 100 to 240 V AC or 100 to 120 V DCMenu N Loop Test Setting items Adhesion Diagnostic FunctionMenu M Automatic Zero Adjustment Function Setting items SO1 Function or F11 SO2 Function, thenMenu P Parameter Protection items N11 Test Output Value Setting for test outputDefault setting Enable Write Protect functionDefault setting After password settingOut Act Mode G20 Alm Out Act Mode Setting of the activeDisplay and output condition for process alarms Totali Pulse Display Alarm Display and output condition for setting alarm occurrencesAlarm description Selection Output UnitAlarms & Warning Messages K12, K13 and K14 are correct Key Descriptions Operation VIA Brain Terminal BT200BT200 Basic Operations Key Layout and DisplayKey-in Sequence Command FunctionFunction Keys 1 BT200 Connection AXFA11 Operation Using a BT200Data update and upload/ download function of BT200 Parameter Setting Using a BT200 3 BT200 Screens & Flow Rate Data DisplayB21 Flow UNIT, press 1 BT200 Setting of Selection-Type Data Flow rate unitsMoved to B Easy Setup 2 BT200 Setting of Numeric-Type Data Flow rate span B23 Flow SPAN, pressMoved to C Basic Setup 3 BT200 Setting of Alphanumeric-Type Data Tag numberTo C Basic Setup Operation VIA Hart Communicator Interconnectwion between AXFA11 and Hart Communicator Specifications for Communication LineConditions of Communication Line Communication Line RequirementsBasic Operation of the Hart Communicator Model Keys and FunctionsKey Operation DisplayCalling Up Menu Addresses Parameters Entering, Setting and Sending DataParameter Configuration Data RenewingExample of Setting Selection-Type Data Base Flow Unit Checking for ProblemsSetting Parameters Example of Setting Numeric-Type Data PV Span PV SpanExample of Setting Alphanumeric-Type Data Tag Number DisplayExample 1 of Other Settings Meter Factor Select MF SetSelect Power Frequency Example 3 of Other Settings Trim Analog Output TrimReturning fld dev to original output appears Select the Scaled D/A trim Returning fld dev to original output appears Exiting Burst Mode Example 4 of Other Settings Burst ModeSetting of Burst Mode Example 5 of Other Settings Multidrop ModePolling Exiting Multidrop Mode Example 6 of Other Settings Write ProtectionSetting Password Open Wrt Protect Menu in the Hot key MenuChanging Password Enable Wrt 10minJoker password and Software Seal Other Functions and OperationsMenu Tree On nextE12 J24 J30 P22 P23 Pre-operation Zero Adjustment Actual OperationZero Adjustment Using Display Unit Switches Selection D Major item Sub- item selection DZero Adjustment via External Status Input Ext Auto Zero Display Unit Replacement MaintenanceComponents Replacement Fuse ReplacementSetting of Burnout Switch Setting of SwitchesAmplifier Replacement Output Setting Pins for BurnoutSetting of Write Protect Switch 4 Switch ConfigurationTroubleshooting No IndicationUnstable Zero StartDisagreement Between Indication and Actual Flow Outline 65 10 Cf 10,000How to Set Parameters FunctionsStandard Performance Power Supply Power supply code Maximum Power ConsumptionSafety Requirement Standards Size 2200 mm 88 in. to 2600 mm 104Accessories Optional Specifications for AXFA11 Remote Converter Terminal WiringRemote Converter AXFA11G Surface mountingRemote Converter AXFA11G /RK AM11 Replacement Bracket General tolerance in the dimensional outline drawingDedicated Signal Cable AXFC-4-L AXFC-0-LRevision Record Edition Date Revised item