APC AXFA11G Name Data range, Default value Units Description Display unit, Indicated item

Page 31

 

 

 

 

 

 

 

6. PARAMETER DESCRIPTION

 

 

 

 

 

 

 

 

 

 

Name

 

Data range

 

Position

Default value

 

 

Item

 

 

 

Units

Description

 

Display unit

R/W

Display unit

of decimal

 

(*): Indicated item

 

 

 

 

(BRAIN)

 

/BRAIN

 

point

 

 

 

 

 

 

 

 

 

 

 

 

B22

Base Time Unit

W

/d

 

 

/s (*)

Selects time units for the flow

 

 

(TIME UNIT)

 

/h

 

 

 

rate span. Linked with C41.

 

 

 

 

/min

 

 

 

 

 

 

 

 

/s

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B23

Flow Span

W

0.0001 to 32000

B21

0 to 4

1 m/s (*)

Sets flow rate span (with units

 

 

(FLOW SPAN)

 

 

/B22

 

 

from B21 and B22). Linked with

 

 

 

 

 

(C40/C41)

 

 

C42.

 

B24

Flow Decimal Pnt

W

Auto

 

 

Auto (*)

Selects decimal point position

 

 

(FLOW DECIMAL)

 

0

 

 

 

for the display unit's

 

 

 

 

1

 

 

 

instantaneous flow rate. Linked

 

 

 

 

2

 

 

 

with C43.

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B30

Total Unit

W

n Unit/P

 

 

Pulse/s (*)

Selects the flow rate unit per one

 

 

(TOTAL UNIT)

 

u Unit/P

 

 

 

pulse as used for totalization

 

 

 

 

m Unit/P

 

 

 

display. Linked with D10.

 

 

 

 

Unit/P

 

 

 

 

 

 

 

 

k Unit/P

 

 

 

 

 

 

 

 

M Unit/P

 

 

 

 

 

 

 

 

Pulse/s

 

 

 

 

 

B31

Total Scale

W

0 to 32000

B30

0 to 4

0 (*)

Sets the flow rate per one pulse

 

 

(TOTAL SCALE)

 

 

(D10)

 

 

for the totalization display.

 

 

 

 

 

 

 

 

Linked with D11.

 

B32

Pulse Unit

W

n Unit/P

 

 

Pulse/s (*)

Selects the flow rate unit per one

 

 

(PULSE UNIT)

 

u Unit/P

 

 

 

pulse as used for pulse output.

 

 

 

 

m Unit/P

 

 

 

Linked with E10.

 

 

 

 

Unit/P

 

 

 

 

 

 

 

 

k Unit/P

 

 

 

 

 

 

 

 

M Unit/P

 

 

 

 

 

 

 

 

Pulse/s

 

 

 

 

 

B33

Pulse Scale

W

0 to 32000

B32

0 to 4

0 (*)

Sets the flow rate per one pulse

 

 

(PULSE SCALE)

 

 

(E10)

 

 

as used for pulse output.

 

 

 

 

 

 

 

 

Linked with E11.

 

B40

Display Select1

W

Flow Rate(%)

 

 

Flow Rate

Selects content of the first line

 

 

(DISP SELECT1)

 

Flow Rate

 

 

 

for Display Mode. Linked with

 

 

 

 

Flow Rate(mA)

 

 

 

H10.

 

 

 

 

Forward Total

 

 

 

 

 

 

 

 

Reverse Total

 

 

 

 

 

 

 

 

Dif Total

 

 

 

 

 

B41

Display Select2

W

Off

 

 

Off

Selects content of the second line

 

 

(DISP SELECT2)

 

Flow Rate(%)

 

 

 

for Display Mode. Linked with

 

 

 

 

Flow Rate

 

 

 

H11.

 

 

 

 

Flow Rate(mA)

 

 

 

 

 

 

 

 

Flow Rate(Bar)

 

 

 

 

 

 

 

 

Forward Total

 

 

 

 

 

 

 

 

Reverse Total

 

 

 

 

 

 

 

 

Dif Total

 

 

 

 

 

 

 

 

Tag No

 

 

 

 

 

 

 

 

Adhesion Check

 

 

 

 

 

 

 

 

Communication

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B42

Display Select3

W

ame as B41

 

 

Off

Selects content of the third line

 

 

(DISP SELECT3)

 

(Display Select2)

 

 

 

for Display Mode. Linked with

 

 

 

 

 

 

 

 

H12.

 

B50

Auto Zero Exe

W

No Execution

 

 

No Execution

Selects whether or not automatic

 

 

(AUTOZERO EXE)

 

Execution

 

 

 

zero adjustment is carried out.

 

 

 

 

 

 

 

 

Linked with M10.

 

 

 

 

 

 

 

 

 

 

B60

R

Good

 

 

 

See “6.5 Alarm Functions”.

 

 

(SELF CHECK)

 

Error

 

 

 

 

 

 

 

 

 

 

 

 

 

 

T0603-2.EPS

6-3

IM 01E20C01-01E

Image 31
Contents Users ManualAXFA11G Contents Operation VIA Hart Communicator Revision Record Introduction Safety and Modification PrecautionsWarranty Using the Magnetic Flowmeter SafelyGuarantee will not apply in the following cases Combination Remote Flowtubes Storage Precautions Handling PrecautionsChecking Model and Specifications AccessoriesSurface Mounting Wall Mounting InstallationInstallation Location MountingUnit mm approx. inch Panel cutout 203 Mounting fixture 172 Inch Pipe MountingPanel Mounting Inch pipe Mounting fixture Washer Clamp screwCables WiringWiring Precautions Dedicated Signal Cable AxfcOn the converter side Wiring PortsExcitation Cable/Power Cable/Output Cable Unit mm approx. inch Crimp TerminalPlastic gland When there are no particular optional specificationsWiring using waterproof glands Conduit WiringTerminal Configuration Wiring ConnectionsRemoving Cover DC Power Connection Precautions for Wiring of Power Supply CablesProtective grounding terminals Wiring the Remote Flowtube with the AXFA11 ConverterAxfc Dedicated signal cable Excitation cable GroundingTo 20 mA DC Current Output Connecting to External Instru- mentsStatus Input Pulse Output13 Status Output Connection Status Output / Alarm OutputSetting switch operations Operating Panel Configuration and FunctionsData display area None ContentDisplay items Reversed-character indicationDisplay Mode → Setting Mode Display Unit Setting MethodsSetting Mode To Parameter Search Mode Setting Mode Parameter Setting ProcedureSetting Example for Selection-Type Data Flow rate units Parameter Sub-item Sub-item Current unit settingSub-item Parameter Search Mode B21 Base Flow UnitB23 Flow Span Setting Example for Numeric-Type Data Flow rate spanSetting Mode Condition Major Item Parameter Search Mode Setting modeIncorrect setting Setting Example for Alphanumeric-Type Data Tag numberSetting Mode Condition Sub-item Parameter Search Mode Major itemSub-item Major item Search Mode Selection CModified value Parameter Description Item a Menu a Display items Item B Menu B Easy Setup itemsName Data range Default value Units Description Display unitIndicated item Name Data range Position Default value Units Display unit Item C Menu C Basic Setting itemsName Item D Menu D Total Setting itemsDefault value Units Display unit Name Data range Position Default value Display unitItem F Menu F Status Functions Setting items Item E Menu E Pulse Setting itemsItem G Menu G Alarm Setting items ALM-SIG Over ALM-SETTINGALM HYS ALM OUT ACTItem J Menu J Auxiliary Function Setting items Item H Menu H Display Setting itemsUnits Display unit Item K Menu K Diagnostic Function Setting itemsItem P Menu P Parameter Protection items Item N Menu N Loop Test Setting itemsData Range Menu B Easy Setup itemsUnit/P 10-9 FU u Unit/P 10-6 FU Pulse/s Example When 1000 m3/h is set for B23/C42 Flow SpanDescription B41/H11 Display Select2 and B42/H12 Display Menu C Basic Setting itemsMeter Factor Settings Is set forB22/C41 Base Time Unit Menu D Total Setting itemsC47 User Span Select Selection of the use C49 Flow User SpanIs set forB22/C41 Base Time TOTAL, and A32 DIF Total when BrainF13 Flow UnitSetting for E20 Pulse Active Mode Menu E Pulse Setting itemsE20 Pulse Active Mode G33 Alm-HH/LL to Yes Menu F Status Functions Setting itemsSetting Function Description Switch Lower and D24 Ttl Switch UpperVal Upper No Function No input function1 Status Output for Automatic Multiple Ranges Switching Multiple ranges settingMultiple Ranges Setting Automatic multiple ranges switching Status Output for Automatic Multiple Ranges SwitchingF10 SO1 Function F11 SO2 Function Parameter setting sequenceFor automatic multiple ranges switching F14 SO1/2 Active ModeF15 2 Multiple Ranges Switching via External Status InputG10 Low Alarm Low alarm setting Menu G Alarm Setting itemsG11 High Alarm High alarm setting Output Example High High Alarm100 High Alarm 100 High Alarm Instantaneous flow rate Low Low Alarm Time t25% 100 85% 75% High Alarm35% Low Alarm 100ClosedOn OpenOff Setting FunctionExample G22 4-20mA Burn Out Display of the currentG45 Alm Record3 Alarm record3 G41 Alm Record1 Alarm record1Alarm Items G43 Alm Record2 Alarm record2Setting range 0 to 10% Hysteresis Menu J Auxiliary Function Setting itemsExample 2 Slurry noise Function orDetermining rate limit value and dead time Example 1 Step inputPower Supply Code 100 to 240 V AC or 100 to 120 V DC J31 Power FrequencyMenu K Diagnostic Function Setting items K10 Adhesion Check Setting of adhesionSO1 Function or F11 SO2 Function, then Adhesion Diagnostic FunctionMenu M Automatic Zero Adjustment Function Setting items Menu N Loop Test Setting itemsWrite Protect function N11 Test Output Value Setting for test outputDefault setting Enable Menu P Parameter Protection itemsAfter password setting Default settingDisplay and output condition for process alarms G20 Alm Out Act Mode Setting of the activeOut Act Mode Output Unit Display and output condition for setting alarm occurrencesAlarm description Selection Totali Pulse Display AlarmAlarms & Warning Messages K12, K13 and K14 are correct Key Layout and Display Operation VIA Brain Terminal BT200BT200 Basic Operations Key DescriptionsFunction Keys Command FunctionKey-in Sequence Data update and upload/ download function of BT200 AXFA11 Operation Using a BT2001 BT200 Connection 3 BT200 Screens & Flow Rate Data Display Parameter Setting Using a BT200Moved to B Easy Setup 1 BT200 Setting of Selection-Type Data Flow rate unitsB21 Flow UNIT, press B23 Flow SPAN, press 2 BT200 Setting of Numeric-Type Data Flow rate spanTo C Basic Setup 3 BT200 Setting of Alphanumeric-Type Data Tag numberMoved to C Basic Setup Operation VIA Hart Communicator Communication Line Requirements Specifications for Communication LineConditions of Communication Line Interconnectwion between AXFA11 and Hart CommunicatorKeys and Functions Basic Operation of the Hart Communicator ModelCalling Up Menu Addresses DisplayKey Operation Data Renewing Entering, Setting and Sending DataParameter Configuration ParametersSetting Parameters Checking for ProblemsExample of Setting Selection-Type Data Base Flow Unit PV Span Example of Setting Numeric-Type Data PV SpanDisplay Example of Setting Alphanumeric-Type Data Tag NumberSelect MF Set Example 1 of Other Settings Meter FactorSelect Power Frequency Trim Example 3 of Other Settings Trim Analog OutputReturning fld dev to original output appears Select the Scaled D/A trim Returning fld dev to original output appears Example 5 of Other Settings Multidrop Mode Example 4 of Other Settings Burst ModeSetting of Burst Mode Exiting Burst ModePolling Example 6 of Other Settings Write Protection Exiting Multidrop ModeOpen Wrt Protect Menu in the Hot key Menu Setting PasswordEnable Wrt 10min Changing PasswordOther Functions and Operations Joker password and Software SealOn next Menu TreeE12 J24 J30 P22 P23 Actual Operation Pre-operation Zero AdjustmentSelection D Major item Sub- item selection D Zero Adjustment Using Display Unit SwitchesZero Adjustment via External Status Input Ext Auto Zero Fuse Replacement MaintenanceComponents Replacement Display Unit ReplacementOutput Setting Pins for Burnout Setting of SwitchesAmplifier Replacement Setting of Burnout Switch4 Switch Configuration Setting of Write Protect SwitchNo Indication TroubleshootingStart Unstable ZeroDisagreement Between Indication and Actual Flow 65 10 Cf 10,000 OutlineFunctions How to Set ParametersStandard Performance Size 2200 mm 88 in. to 2600 mm 104 Maximum Power ConsumptionSafety Requirement Standards Power Supply Power supply codeAccessories Terminal Wiring Optional Specifications for AXFA11 Remote ConverterSurface mounting Remote Converter AXFA11GGeneral tolerance in the dimensional outline drawing Remote Converter AXFA11G /RK AM11 Replacement BracketAXFC-0-L Dedicated Signal Cable AXFC-4-LEdition Date Revised item Revision Record