8. OPERATION VIA HART COMMUNICATOR
8.3.4.4Example 1 of Other Settings: Meter Factor
The “Meter Factor” is engraved on the data plate of the combined flowtube.
The meter factor is required to compute the correct electromotive force proportional to the fluid velocity, and is determined by actual
Open the “MF” setting display:
1. Device Setup 4. Detailed Setup 1. Basic Setup 3. MF Set
1
AXFA11 :
Basic Setup
1Tag
2PV Damping
3MF Set
4Select Flow Tube
5 Nominal Size Unit
SAVE HOME
or
YZ/
3
Select “ MF Set ”.
2AXFA11 :
Standard DF Enhanced DF
ABORT ENTER
3AXFA11 :
1.0000
1.1111
DEL ABORT ENTER
4AXFA11 :
1.0000
1.2222
DEL ABORT ENTER
5AXFA11 :
Basic Setup
1Tag
2PV Damping
3MF Set
4Select Flow Tube
5 Nominal Size Unit
SAVE HOME
STU or F4 1
(ENTER)
F4
(ENTER)
F4
(ENTER)
F3
(HOME)
Select “ Standard DF ”.
Enter “1.1111” with alphanumeric keys. Press ENTER (F4).
Enter “1.2222” with alphanumeric keys. Press ENTER (F4).
Press HOME (F3), and return to the “Online” menu.
F0812.EPS