6. PARAMETER DESCRIPTION

Setting Alarms (Device is normal but errors have been made in the setting of parameters.)

Display unit/BRAIN

Alarm countermeasure

Alarm description

Countermeasure

(￿60) content

message on display unit

 

 

 

 

Span flow velocity setting is 11 m/s or

Check whether parameters

50:Span > 10m/s

Check parameter C40, C41,

more

C40, C41, and C42 are correct.

 

In case that multiple range or

and C42

 

 

 

forward and reverse flow

 

 

 

 

 

 

measurement functions is used,

51:Span < 0.1m/s

Check parameter C40, C41,

Span flow velocity setting is 0.05 m/s or

check whether parameters F30

and C42

less

through F36 are correct.

 

52:TTL>10000p/s

Check parameter D10 and

Totalization rate is 11000 pps or more

 

D11

 

Check whether parameters

 

 

53:TTL<0.0001p/s

Check parameter D10 and

Totalization rate is 0.00005 pps or less

D10 and D11 are correct.

D11

 

 

 

 

 

54:4-20 Lmt Err

Check parameter J11and J12

The condition [4-20 low limit (J11) <

Check whether parameters J11

 

 

4-20 high limit (J12)] is not satisfied.

and J12 are correct.

55:Multi Rng Err

 

The condition [No. 1 range < No. 2

Check whether parameters

Check parameter F30 to F36

range < No. 3 range < No. 4 range] is

F30 through F36 are correct.

 

 

not satisfied for multiple ranges.

 

 

 

 

 

 

 

 

 

The condition [High Alarm (G11) –

 

56:H/L HH/LL Set

 

Low Alarm (G10) > H/L Alarm Hys

Check whether parameters

Check parameter G10 to G14

(G14)] or the condition [High High

G10 through G14 are correct.

 

 

Alarm (G13) – Low Low Alarm (G12)

 

 

 

 

 

> H/L Alarm Hys (G14)] is not satisfied.

 

 

Check parameter C40, C45,

Mass units have been selected for Base

Check whether parameters

57:Dens Set Err

Flow Unit (C40) but density is set to 0.

and C46

C40, C45, and C46 are correct.

 

 

 

 

 

 

 

 

Pulse rate is 11000 pps or more with

 

60:PLS > 10000p/s

Check parameter E10, E11,

50% duty selection.

 

and E12

Pulse rate is 10000 pps or more with

 

 

 

 

 

0.05 ms selection.

 

61:PLS > 5000p/s

Check parameter E10, E11,

Pulse rate is 5000 pps or more with 0.1

 

and E12

ms selection.

 

 

 

 

 

 

 

62:PLS > 1000p/s

Check parameter E10, E11,

Pulse rate is 1000 pps or more with 0.5

 

and E12

ms selection.

 

 

 

63:PLS > 500p/s

Check parameter E10, E11,

Pulse rate is 500 pps or more with 1 ms

 

and E12

selection.

Check whether parameters

 

64:PLS > 25p/s

Check parameter E10, E11,

Pulse rate is 25 pps or more with 20 ms

E10, E11, and E12 are correct.

and E12

selection.

 

 

 

 

 

 

 

65:PLS > 15p/s

Check parameter E10, E11,

Pulse rate is 15 pps or more with 33 ms

 

and E12

selection.

 

 

 

 

 

 

 

66:PLS > 10p/s

Check parameter E10, E11,

Pulse rate is 10 pps or more with 50 ms

 

and E12

selection.

 

 

 

 

 

 

 

67:PLS > 5p/s

Check parameter E10, E11,

Pulse rate is 5 pps or more with 100 ms

 

and E12

selection.

 

 

 

 

 

 

 

70:PLS<0.0001p/s

Check parameter E10, E11,

Pulse rate is 0.00005 pps or less.

 

and E12

 

 

 

 

 

 

 

 

71:Meas Mod Set

Check parameter C20

Measure Mode (C20) is set to Enhanced

Check whether parameter C20

DF without selecting an optional

is correct.

 

 

code HF1 or HF2.

 

 

 

72: Size Set Err

Check parameter C32

A value of 3000.1 mm or more is set for

Check whether parameter C32

Nominal Size (C32).

is correct.

 

 

73: Adh Set Err

Check parameter K11 to K14

The condition in Adhesion detection

Check whether parameters K11,

 

level, Level:1<Level:2<Level:3<Level:4

K12, K13 and K14 are correct.

 

 

is not satisfied.

 

 

 

 

T0653.EPS

Setting Alarms (Device and measurements are normal but a warning is issued.)

Display unit/BRAIN

Alarm countermeasure

Alarm description

Countermeasure

(￿60) content

message on display unit

 

 

80:Adhesion Wng

Slight adhesion to

Clean and check the electrodes.

electrodes.

Refer to parameter K13.

 

 

 

 

 

Carry out adjustment as follows:

 

 

Results of automatic zero

Check if the flowtube is filled with

82:Auto Zero Wng

fluid.

adjustment are higher than

Check if the flow velocity is

 

 

 

the rated values.

 

 

completely zero.

 

 

 

 

 

 

Check the condition of grounding.

 

 

 

 

 

 

 

Confirm whether the flow rate is in

83:Fix Cur Wng

The current value is fixed.

excess of the upper limit (108%) or

below the lower limit (-8%), or

 

 

 

whether upon entry to the Test Mode

 

 

 

or not.

84:Disp Over Wng

Overflow in the display digits during

Check whether parameter C43 is

(only for display unit)

instantaneous flow rate display.

correct.

 

 

 

 

 

90:Disp SW Wng

 

Display unit switches are not

Investigate whether the display unit

cover is fitted or whether the cover’s

(only for display unit)

operating.

glass surface is dirty.

 

 

 

 

 

 

 

 

 

 

T0654.EPS

6-39

IM 01E20C01-01E

Image 67
Contents Users ManualAXFA11G Contents Operation VIA Hart Communicator Revision Record Introduction Safety and Modification PrecautionsWarranty Using the Magnetic Flowmeter SafelyGuarantee will not apply in the following cases Combination Remote Flowtubes Storage Precautions Handling PrecautionsChecking Model and Specifications AccessoriesSurface Mounting Wall Mounting InstallationInstallation Location MountingUnit mm approx. inch Panel cutout 203 Mounting fixture 172 Inch Pipe MountingPanel Mounting Inch pipe Mounting fixture Washer Clamp screwCables WiringWiring Precautions Dedicated Signal Cable AxfcOn the converter side Wiring PortsExcitation Cable/Power Cable/Output Cable Unit mm approx. inch Crimp TerminalPlastic gland When there are no particular optional specificationsWiring using waterproof glands Conduit WiringTerminal Configuration Wiring ConnectionsRemoving Cover DC Power Connection Precautions for Wiring of Power Supply CablesProtective grounding terminals Wiring the Remote Flowtube with the AXFA11 ConverterAxfc Dedicated signal cable Excitation cable GroundingTo 20 mA DC Current Output Connecting to External Instru- mentsStatus Input Pulse Output13 Status Output Connection Status Output / Alarm OutputSetting switch operations Operating Panel Configuration and FunctionsData display area None ContentDisplay items Reversed-character indicationDisplay Mode → Setting Mode Display Unit Setting MethodsSetting Mode To Parameter Search Mode Setting Mode Parameter Setting ProcedureSetting Example for Selection-Type Data Flow rate units Parameter Sub-item Sub-item Current unit settingSub-item Parameter Search Mode B21 Base Flow UnitB23 Flow Span Setting Example for Numeric-Type Data Flow rate spanSetting Mode Condition Major Item Parameter Search Mode Setting modeIncorrect setting Setting Example for Alphanumeric-Type Data Tag numberSetting Mode Condition Sub-item Parameter Search Mode Major itemSub-item Major item Search Mode Selection CModified value Parameter Description Item a Menu a Display items Item B Menu B Easy Setup itemsName Data range Default value Units Description Display unitIndicated item Name Data range Position Default value Units Display unit Item C Menu C Basic Setting itemsName Item D Menu D Total Setting itemsDefault value Units Display unit Name Data range Position Default value Display unitItem F Menu F Status Functions Setting items Item E Menu E Pulse Setting itemsItem G Menu G Alarm Setting items ALM-SIG Over ALM-SETTINGALM HYS ALM OUT ACTItem J Menu J Auxiliary Function Setting items Item H Menu H Display Setting itemsUnits Display unit Item K Menu K Diagnostic Function Setting itemsItem P Menu P Parameter Protection items Item N Menu N Loop Test Setting itemsData Range Menu B Easy Setup itemsUnit/P 10-9 FU u Unit/P 10-6 FU Pulse/s Example When 1000 m3/h is set for B23/C42 Flow SpanDescription B41/H11 Display Select2 and B42/H12 Display Menu C Basic Setting itemsMeter Factor Settings Is set forB22/C41 Base Time Unit Menu D Total Setting itemsC47 User Span Select Selection of the use C49 Flow User SpanIs set forB22/C41 Base Time TOTAL, and A32 DIF Total when BrainF13 Flow UnitSetting for E20 Pulse Active Mode Menu E Pulse Setting itemsE20 Pulse Active Mode G33 Alm-HH/LL to Yes Menu F Status Functions Setting itemsSetting Function Description Switch Lower and D24 Ttl Switch UpperVal Upper No Function No input function1 Status Output for Automatic Multiple Ranges Switching Multiple ranges settingMultiple Ranges Setting Automatic multiple ranges switching Status Output for Automatic Multiple Ranges SwitchingF10 SO1 Function F11 SO2 Function Parameter setting sequenceFor automatic multiple ranges switching F14 SO1/2 Active ModeF15 2 Multiple Ranges Switching via External Status InputG10 Low Alarm Low alarm setting Menu G Alarm Setting itemsG11 High Alarm High alarm setting Output Example High High Alarm100 High Alarm 100 High Alarm Instantaneous flow rate Low Low Alarm Time t25% 100 85% 75% High Alarm35% Low Alarm 100ClosedOn OpenOff Setting FunctionExample G22 4-20mA Burn Out Display of the currentG45 Alm Record3 Alarm record3 G41 Alm Record1 Alarm record1Alarm Items G43 Alm Record2 Alarm record2Setting range 0 to 10% Hysteresis Menu J Auxiliary Function Setting itemsExample 2 Slurry noise Function orDetermining rate limit value and dead time Example 1 Step inputPower Supply Code 100 to 240 V AC or 100 to 120 V DC J31 Power FrequencyMenu K Diagnostic Function Setting items K10 Adhesion Check Setting of adhesionSO1 Function or F11 SO2 Function, then Adhesion Diagnostic FunctionMenu M Automatic Zero Adjustment Function Setting items Menu N Loop Test Setting itemsWrite Protect function N11 Test Output Value Setting for test outputDefault setting Enable Menu P Parameter Protection itemsAfter password setting Default settingDisplay and output condition for process alarms G20 Alm Out Act Mode Setting of the activeOut Act Mode Output Unit Display and output condition for setting alarm occurrencesAlarm description Selection Totali Pulse Display AlarmAlarms & Warning Messages K12, K13 and K14 are correct Key Layout and Display Operation VIA Brain Terminal BT200BT200 Basic Operations Key DescriptionsFunction Keys Command FunctionKey-in Sequence Data update and upload/ download function of BT200 AXFA11 Operation Using a BT2001 BT200 Connection 3 BT200 Screens & Flow Rate Data Display Parameter Setting Using a BT200Moved to B Easy Setup 1 BT200 Setting of Selection-Type Data Flow rate unitsB21 Flow UNIT, press B23 Flow SPAN, press 2 BT200 Setting of Numeric-Type Data Flow rate spanTo C Basic Setup 3 BT200 Setting of Alphanumeric-Type Data Tag numberMoved to C Basic Setup Operation VIA Hart Communicator Communication Line Requirements Specifications for Communication LineConditions of Communication Line Interconnectwion between AXFA11 and Hart CommunicatorKeys and Functions Basic Operation of the Hart Communicator ModelCalling Up Menu Addresses DisplayKey Operation Data Renewing Entering, Setting and Sending DataParameter Configuration ParametersSetting Parameters Checking for ProblemsExample of Setting Selection-Type Data Base Flow Unit PV Span Example of Setting Numeric-Type Data PV SpanDisplay Example of Setting Alphanumeric-Type Data Tag NumberSelect MF Set Example 1 of Other Settings Meter FactorSelect Power Frequency Trim Example 3 of Other Settings Trim Analog OutputReturning fld dev to original output appears Select the Scaled D/A trim Returning fld dev to original output appears Example 5 of Other Settings Multidrop Mode Example 4 of Other Settings Burst ModeSetting of Burst Mode Exiting Burst ModePolling Example 6 of Other Settings Write Protection Exiting Multidrop ModeOpen Wrt Protect Menu in the Hot key Menu Setting PasswordEnable Wrt 10min Changing PasswordOther Functions and Operations Joker password and Software SealOn next Menu TreeE12 J24 J30 P22 P23 Actual Operation Pre-operation Zero AdjustmentSelection D Major item Sub- item selection D Zero Adjustment Using Display Unit SwitchesZero Adjustment via External Status Input Ext Auto Zero Fuse Replacement MaintenanceComponents Replacement Display Unit ReplacementOutput Setting Pins for Burnout Setting of SwitchesAmplifier Replacement Setting of Burnout Switch4 Switch Configuration Setting of Write Protect SwitchNo Indication TroubleshootingStart Unstable ZeroDisagreement Between Indication and Actual Flow 65 10 Cf 10,000 OutlineFunctions How to Set ParametersStandard Performance Size 2200 mm 88 in. to 2600 mm 104 Maximum Power ConsumptionSafety Requirement Standards Power Supply Power supply codeAccessories Terminal Wiring Optional Specifications for AXFA11 Remote ConverterSurface mounting Remote Converter AXFA11GGeneral tolerance in the dimensional outline drawing Remote Converter AXFA11G /RK AM11 Replacement BracketAXFC-0-L Dedicated Signal Cable AXFC-4-LEdition Date Revised item Revision Record