(1)When there are no particular optional specifications
The wiring port is sealed with a cap (not
(2) Wiring using waterproof glands
IMPORTANT
To prevent water or condensation from entering the converter housing, waterproof glands are recommended. Do not
Washer
Gasket
Waterproof gland
Cable
F0404.EPS
Figure 4.3.1 Waterproof Gland (Optional Code EG)
For working on the electric wire tubes or the flexible tubes (PF1/2), remove the waterproof gland and attach them directly to the wiring port.
Washer
PF1/2
Gasket
Tightening gland
Cable
When working on conduit pipes or flexible pipes (PF1/2 only)
F0405.EPS
Figure 4.3.2 Waterproof Gland with Union Joint
(Optional Code EU)
4. WIRING
Plastic gland
F0406.EPS
Figure 4.3.3 Plastic Gland (Optional Code EP)
PF1/2
PF3/4
Gasket
Washer
Extension plug (x2)
Conversion plug (x5)
Cable
When working on electric wire tube or flexible tube (PF3/4)
*When connecting PF1/2, remove the conversion plug and connect directly to wiring port.
F0407.EPS
Figure 4.3.4 PF3/4 Waterproof Gland (Optional Code EW)
(3) Conduit Wiring
When wiring the conduits, pass the conduit through the wiring connection port, and utilize the waterproof gland to prevent water from flowing in. Place the conduit pipe on an angle as shown in Figure 4.3.5. Install a drain valve at the low end of the vertical pipe, and open the valve regularly.
Drain valve
F0408.EPS
Figure 4.3.5 Conduit Wiring