6AXFA11 :
Fld dev output 4.000 mA equal to reference meter?
1 Yes
2 No
ABORT ENTER
7AXFA11 :
ABORT OK
8AXFA11 :
HELP DEL ABORT ENTER
9AXFA11 :
Fld dev output 20.000 mA equal to reference meter?
1 Yes
2 No
ABORT ENTER
10AXFA11 :
OK
11AXFA11 :
Adjustment
1 Auto Zero Exe
2 Magflow Zero
3D/A trim
4 Scaled D/A trim
HELP SAVE HOME
F4
(ENTER)
F4
(OK)
‘19.050’
F4
(ENTER)
F4
(ENTER)
F4
(OK)
F3
(HOME)
8. OPERATION VIA HART COMMUNICATOR
Ammeter reading: 4.000
Because the reading on the ammeter is 4.000 mA, select “Yes” and press ENTER (F4).
If the reading is not 4.000 mA, select “No”. Repeat step 5 until the ammeter reads 4.000 mA.
Press OK (F4), and the AXFA11 outputs an output signal of 100%.
Ammeter reading: 19.050
Carry out the same procedures as those described in 5 .
Ammeter reading: 20.000
“Returning fld dev to original output” appears.
Press OK (F4).
Press HOME (F3), and return to the “Online” menu.
IM |