A&D HV-G Series instruction manual Counting Mode, Storing a Unit Mass

Page 22

8 Counting Mode

Counting Mode

In counting mode the total weight of articles is converted to a count and assumes all articles have the same mass value.

It is necessary to store a unit mass to count articles.

8.1 Storing a Unit Mass

Step

1

Press the

 

MODE

switch to display the unit PCS.

Step

2

Press the

 

switch to enter the mode to store a

SET

 

 

unit mass.

Step

3

Press the ^ switch to select the number of sam-

 

 

ples. The greater the quantity of samples, the

 

 

greater the accuracy of the count.

 

 

5 pieces, 10 pieces, 20 pieces, 50 pieces, 100

 

 

pieces

Step

4

Put the container onto the pan.

 

 

When the stability mark is displayed press the

 

 

 

switch.

 

 

TARE

Step 5 Put in the number of samples selected at step 3. Press the ENTER switch to store the weight after the stability mark is displayed. The count is then displayed.

Caution

When the sample is too light and it is not possible to calculate a unit mass, the scale displays lo ut and returns to step 3. It is necessary to increase the number of samples.

Pressing the ENTER switch after the lo ut .is displayed, the next weighing mode is displayed.

When the unit mass is too light to store, the scale displays lo ut .

Step 6 Remove all items from the pan.

HV-G Instruction Manual Page 21

Image 22
Contents HV-G Series Contents Page Compliance Compliance with FCC rulesOutline and Features Unpacking Accessories and Options list Type Products AccessoriesOrder code or option name Accessories Options ListPrecautions for Operating the Scale Precautions for Installing the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names Display and Symbols OFFPage Switches Page Turning the Scale On/Off & Basic Weighing Basic OperationMemo Type L with Batteries Tare and Net Display Semi-Automatic Tare Input by WeighingOperation and Performance Weighing RangeWeighing Range Products Weighing RangeMode Switch Changing Unit & Mode Counting Mode Storing a Unit MassCounting the number of articles Percentage Mode Storing a 100% MassReading Percentage Accumulation Function Operation and SwitchesFunction table Meaning & Purpose Preparation Setting Parameters Operation and Performance Examples ExampleComparison Condition Comparator FunctionComparator Sign Function table Meaning and PurposeAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Full/Dribble batch sub-function ComparatorHold Parameter and display an item of the full/dribble Switch to display the HI Simple Batch Function Function Table Meaning and purpose Simple batch functionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Complete Calibration Procedure Gravity Acceleration CorrectionCalibration of the Zero Point Procedure for Setting Parameters Function Table15.2 Parameter List Display Meaning and purpose Function or full/dribble batch functionPrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Data Format FormatFormat Stream Mode Preparation and Performance ExamplesCancel of tare Command ModeTare Preset tare Upper limit valueFinal value Preparation Setting Parameters Options Pin connections Pin No Signal name Direction Description17.1 RS-232C / Relay output / Buzzer OP-03 Pin No Signal name Direction Description 17.2 RS-422 / RS-485 / Relay output OP-04Installing OP-04 Communication Format Zero Lower limit valueAccumulation count output Reset of Accumulation data and numberAccumulation data output Final valueInternal Printer for Type V OP-06 SpecificationConsumables Disassembly InstallationPerformance Check Specifications HV-G seriesDimensions Page Repair MaintenanceCheck Points Before Calling Maintenance This situation Confirm these items