A&D HV-G Series instruction manual Format

Page 51

Format 2

There are four headers for the type of data and weighing condition.

Stable weighing data

S

T

Unstable weighing data

U S

Stable counting data

Q T

Out of range (over)

O L

The weighing data consists of 7 characters including decimal point and polarity. The data consists of 7 characters in the counting mode.

The polarity is always output.

There are three units. The unit is 2 characters.

In case of “out of range”, numbers become all 9s.

The terminator is always output as CR LF.

Caution

When the data is longer than 7 characters in the counting or percentage mode, the overflow is ignored.

Data in the unit of kg

Counting mode

Percentage mode

Out of range

Definition of symbols CR (carriage return) 0D[H]

Space 20[H]

C R

LF (line feed) 0A[H]

L

F

HV-G Instruction Manual Page 50

Image 51
Contents HV-G Series Contents Page Compliance with FCC rules ComplianceOutline and Features Unpacking Type Products Accessories Accessories and Options listOptions List Order code or option name AccessoriesPrecautions for Installing the Scale Precautions for Operating the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names OFF Display and SymbolsPage Switches Page Basic Operation Turning the Scale On/Off & Basic WeighingMemo Type L with Batteries Semi-Automatic Tare Input by Weighing Tare and Net DisplayProducts Weighing Range Weighing RangeWeighing Range Operation and PerformanceMode Switch Changing Unit & Mode Storing a Unit Mass Counting ModeCounting the number of articles Storing a 100% Mass Percentage ModeReading Percentage Operation and Switches Accumulation FunctionFunction table Meaning & Purpose Preparation Setting Parameters Example Operation and Performance ExamplesFunction table Meaning and Purpose Comparator FunctionComparator Sign Comparison ConditionAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Comparator Full/Dribble batch sub-functionHold Parameter and display an item of the full/dribble Switch to display the HI Function Table Meaning and purpose Simple batch function Simple Batch FunctionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Gravity Acceleration Correction Complete Calibration ProcedureCalibration of the Zero Point Function Table Procedure for Setting Parameters15.2 Parameter List Function or full/dribble batch function Display Meaning and purposePrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Format Data FormatFormat Preparation and Performance Examples Stream ModeCommand Mode Cancel of tareTare Upper limit value Preset tareFinal value Preparation Setting Parameters Pin connections Pin No Signal name Direction Description Options17.1 RS-232C / Relay output / Buzzer OP-03 17.2 RS-422 / RS-485 / Relay output OP-04 Pin No Signal name Direction DescriptionInstalling OP-04 Communication Format Lower limit value ZeroFinal value Reset of Accumulation data and numberAccumulation data output Accumulation count outputSpecification Internal Printer for Type V OP-06Consumables Installation DisassemblyPerformance Check HV-G series SpecificationsDimensions Page This situation Confirm these items MaintenanceCheck Points Before Calling Maintenance Repair