A&D HV-G Series instruction manual Operation and Performance Examples, Case

Page 32
Upper limit value (HI) Lower limit value (LO)

11.2 Operation and Performance (Examples)

Example 1

This example is set as follows:

Function table f6 3 (If the current display value is not near-zero, the scale compares the display value with the

upper limit value and the lower limit value immediately.)

7.000kg

6.500kg

Case

The comparison starts at turning the scale on.

When the current value is less than 6.500kg, LO is displayed.

When the current value is between 6.500kg and 7.000kg, OK is displayed. When the current value is greater than 7.000kg, HI is displayed.

Example 2

This example is set as follows:

 

 

 

 

Function table

 

(Pressing the

 

 

switch, after the stability

f6 4

SET

 

 

mark is displayed, the scale compares the

 

 

current display value with the upper limit value

 

 

and the lower limit value immediately.)

Upper limit value (HI)

2.000kg

Lower limit value (LO)

-1.000kg

Case

Pressing the SET stability mark.

switch, the comparison is performed after displaying the

When the current value is less than -1.000kg, LO is displayed.

When the current value is between -1.000kg and 2.000kg, OK is displayed.

When the current value is greater than 2.000kg, HI is displayed.

HV-G Instruction Manual Page 31

Image 32
Contents HV-G Series Contents Page Compliance Compliance with FCC rulesOutline and Features Unpacking Accessories and Options list Type Products AccessoriesOrder code or option name Accessories Options ListPrecautions for Storing the Scale Precautions for Installing the ScalePrecautions for Operating the Scale Installing the Scale Installing the Batteries for Type L Names Display and Symbols OFFPage Switches Page Memo Basic OperationTurning the Scale On/Off & Basic Weighing Type L with Batteries Tare and Net Display Semi-Automatic Tare Input by WeighingWeighing Range Weighing RangeOperation and Performance Products Weighing RangeMode Switch Changing Unit & Mode Counting Mode Storing a Unit MassCounting the number of articles Percentage Mode Storing a 100% MassReading Percentage Function table Meaning & Purpose Operation and SwitchesAccumulation Function Preparation Setting Parameters Operation and Performance Examples Example Comparator Function Comparator Sign Comparison Condition Function table Meaning and PurposeAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Hold ComparatorFull/Dribble batch sub-function Parameter and display an item of the full/dribble Switch to display the HI Simple Batch Function Function Table Meaning and purpose Simple batch functionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Complete Calibration Procedure Gravity Acceleration CorrectionCalibration of the Zero Point Procedure for Setting Parameters Function Table15.2 Parameter List Display Meaning and purpose Function or full/dribble batch functionPrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Data Format FormatFormat Stream Mode Preparation and Performance ExamplesTare Command ModeCancel of tare Preset tare Upper limit valueFinal value Preparation Setting Parameters 17.1 RS-232C / Relay output / Buzzer OP-03 Pin connections Pin No Signal name Direction DescriptionOptions Pin No Signal name Direction Description 17.2 RS-422 / RS-485 / Relay output OP-04Installing OP-04 Communication Format Zero Lower limit valueReset of Accumulation data and number Accumulation data outputAccumulation count output Final valueConsumables SpecificationInternal Printer for Type V OP-06 Disassembly InstallationPerformance Check Specifications HV-G seriesDimensions Page Maintenance Check Points Before Calling MaintenanceRepair This situation Confirm these items