A&D HV-G Series instruction manual Parameter List

Page 45

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15.2

 

Parameter List

 

Item

Display

 

Meaning and purpose

 

 

 

 

 

 

 

The type L scale is turned off after 5 minutes of no operation,

 

 

 

 

 

 

 

when displaying zero.

 

Automatic

f1 0 #

Not used

 

turning off

f1 1

 

Used

 

 

 

 

 

 

 

 

Selects the method of changing the weighing range

 

Range

f2 0 #

Automatic range change.

 

 

 

 

 

f2

1

 

Manual range change using the

 

 

 

switch.

 

 

 

 

 

 

RANGE

 

 

 

 

 

 

 

 

Transmission rate of the serial

interface

(RS-232C/ 422/485).

 

 

 

 

 

f4

0

#

2400bps

 

Baud rate

f4 1

 

4800bps

 

 

 

 

 

f4

2

 

9600bps

 

 

 

 

 

 

 

 

Mode selection for the serial interface (RS-232C/ 422/485).

 

 

 

 

 

f5

0

#

Stream mode. (Refer to “16.2 Stream Mode”)

 

 

 

 

 

f5

1

 

Command mode. (Refer to “16.3 Command Mode”)

 

 

 

 

 

f5

2

 

A data is output, when the

 

 

 

switch is pressed.

 

 

 

 

 

 

PRINT

 

 

 

 

 

 

f5

3

 

Auto-print +

 

 

 

 

 

 

 

 

 

 

 

 

 

When the display becomes a positive stable value above

 

 

 

 

 

 

 

 

near-zero, the scale outputs the data automatically. Next

 

Output mode

 

 

 

output can be performed after the display becomes near-

 

 

 

 

 

 

 

 

zero or a negative value.

 

 

 

 

 

f5 4`

Auto-print +/-

 

 

 

 

 

 

 

 

When the display becomes a stable value outside near-zero,

 

 

 

 

 

 

 

 

the scale outputs the data automatically. Next output can be

 

 

 

 

 

 

 

 

performed after the display becomes near-zero.

 

 

 

 

 

f5 5

 

At each finish of full/dribble batch function, the data is output.

 

 

 

 

 

 

 

 

 

 

 

 

f6 0 #

The scale always compares the current display value. Pressing

 

 

 

 

 

 

 

 

the

 

switch turns the comparator on and off

 

 

 

 

 

 

 

 

SET

 

 

 

 

 

f6 1

 

The

scale always compares the display value.

 

Comparator

f6 2

 

The scale always compares the display value if not near-zero.

 

 

 

 

 

 

 

 

Pressing the

 

switch turns the comparator on and off

 

 

 

 

 

 

 

 

SET

 

 

 

 

 

f6 3

 

The scale always compares the display value when not

 

 

 

 

 

 

 

 

near-zero.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Notes

The “#” indicates factory settings.

The “near-zero” is within ±4 digits from zero point in the unit of kg.

HV-G Instruction Manual Page 44

Image 45
Contents HV-G Series Contents Page Compliance with FCC rules ComplianceOutline and Features Unpacking Type Products Accessories Accessories and Options listOptions List Order code or option name AccessoriesPrecautions for Installing the Scale Precautions for Operating the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names OFF Display and SymbolsPage Switches Page Basic Operation Turning the Scale On/Off & Basic WeighingMemo Type L with Batteries Semi-Automatic Tare Input by Weighing Tare and Net DisplayWeighing Range Weighing RangeOperation and Performance Products Weighing RangeMode Switch Changing Unit & Mode Storing a Unit Mass Counting ModeCounting the number of articles Storing a 100% Mass Percentage ModeReading Percentage Operation and Switches Accumulation FunctionFunction table Meaning & Purpose Preparation Setting Parameters Example Operation and Performance ExamplesComparator Sign Comparator FunctionComparison Condition Function table Meaning and PurposeAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Comparator Full/Dribble batch sub-functionHold Parameter and display an item of the full/dribble Switch to display the HI Function Table Meaning and purpose Simple batch function Simple Batch FunctionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Gravity Acceleration Correction Complete Calibration ProcedureCalibration of the Zero Point Function Table Procedure for Setting Parameters15.2 Parameter List Function or full/dribble batch function Display Meaning and purposePrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Format Data FormatFormat Preparation and Performance Examples Stream ModeCommand Mode Cancel of tareTare Upper limit value Preset tareFinal value Preparation Setting Parameters Pin connections Pin No Signal name Direction Description Options17.1 RS-232C / Relay output / Buzzer OP-03 17.2 RS-422 / RS-485 / Relay output OP-04 Pin No Signal name Direction DescriptionInstalling OP-04 Communication Format Lower limit value ZeroAccumulation data output Reset of Accumulation data and numberAccumulation count output Final valueSpecification Internal Printer for Type V OP-06Consumables Installation DisassemblyPerformance Check HV-G series SpecificationsDimensions Page Check Points Before Calling Maintenance MaintenanceRepair This situation Confirm these items