10.2 Operation and Performance (Examples)

Example 1

 

Weighing each article, the scale makes the accumulation according to f8 3 .

 

Step

1

If accumulation has previously been used press the

 

switch to display

.

MODE

Step

2

Press the

 

 

 

 

 

switch to reset the accumulation data.

 

ZERO

 

 

Step

3

Return to the kg mode using the

 

switch.

 

MODE

 

 

 

Press the

 

 

 

 

 

switch with nothing on

the pan.

 

 

 

 

ZERO

 

 

Step

4

Put an article on the pan. Wait for the stability mark to be displayed and the

 

 

 

value to be blinking. Remove the article and press the

 

switch.

 

 

 

ZERO

 

Step

5

Weigh additional articles using step 4.

 

Step

6

Press the

 

 

switch to display the number of articles and total mass with

 

MODE

 

 

 

 

 

displayed.

 

Example 2

 

This example accumulates the articles that were removed from the pan. The

 

function parameter is set to f8 4 .

 

 

Step

1

Enter into the kg mode using the

 

 

 

 

switch.

 

 

MODE

 

 

Put all articles on the pan and press the

 

switch.

 

 

TARE

Step

2

Press the

 

 

 

switch to display

.

 

MODE

 

Step

3

Press the

 

 

switch to reset the accumulation data.

ZERO

 

 

Retun to kg mode with the

 

 

 

 

 

switch.

 

 

MODE

 

 

Step

4

Remove an article from the pan. Wait for the stability mark to be displayed and

 

 

the value to blink. Press the

 

 

switch.

 

 

 

 

TARE

 

 

 

Step

5

Weigh additional articles using step 4.

 

 

Step 6

Press the

 

 

switch to display the number of articles and total mass with

MODE

 

 

 

displayed.

 

 

HV-G Instruction Manual Page 27

Image 28
Contents HV-G Series Contents Page Compliance Compliance with FCC rulesOutline and Features Unpacking Accessories and Options list Type Products AccessoriesOrder code or option name Accessories Options ListPrecautions for Operating the Scale Precautions for Installing the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names Display and Symbols OFFPage Switches Page Turning the Scale On/Off & Basic Weighing Basic OperationMemo Type L with Batteries Tare and Net Display Semi-Automatic Tare Input by WeighingWeighing Range Weighing RangeOperation and Performance Products Weighing RangeMode Switch Changing Unit & Mode Counting Mode Storing a Unit MassCounting the number of articles Percentage Mode Storing a 100% MassReading Percentage Accumulation Function Operation and SwitchesFunction table Meaning & Purpose Preparation Setting Parameters Operation and Performance Examples ExampleComparator Function Comparator SignComparison Condition Function table Meaning and PurposeAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Full/Dribble batch sub-function ComparatorHold Parameter and display an item of the full/dribble Switch to display the HI Simple Batch Function Function Table Meaning and purpose Simple batch functionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Complete Calibration Procedure Gravity Acceleration CorrectionCalibration of the Zero Point Procedure for Setting Parameters Function Table15.2 Parameter List Display Meaning and purpose Function or full/dribble batch functionPrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Data Format FormatFormat Stream Mode Preparation and Performance ExamplesCancel of tare Command ModeTare Preset tare Upper limit valueFinal value Preparation Setting Parameters Options Pin connections Pin No Signal name Direction Description17.1 RS-232C / Relay output / Buzzer OP-03 Pin No Signal name Direction Description 17.2 RS-422 / RS-485 / Relay output OP-04Installing OP-04 Communication Format Zero Lower limit valueReset of Accumulation data and number Accumulation data outputAccumulation count output Final valueInternal Printer for Type V OP-06 SpecificationConsumables Disassembly InstallationPerformance Check Specifications HV-G seriesDimensions Page Maintenance Check Points Before Calling MaintenanceRepair This situation Confirm these items